Statistical analysis of IMRT dosimetry quality assurance measurements for local delivery guideline

被引:15
|
作者
Chung, Jin Beom [3 ]
Kim, Jae Sung [3 ]
Ha, Sung Whan [1 ,2 ]
Ye, Sung-Joon [1 ,2 ,4 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Radiat Oncol, Seoul 110744, South Korea
[2] Seoul Natl Univ, Coll Med, Inst Radiat Med, Seoul 110744, South Korea
[3] Seoul Natl Univ, Bundang Hosp Seongnam, Dept Radiat Oncol, Gyeonggi Do 463707, South Korea
[4] Seoul Natl Univ, Dept Intelligent Convergence Syst, Grad Sch Convergence Sci & Technol, Seoul 151742, South Korea
来源
RADIATION ONCOLOGY | 2011年 / 6卷
关键词
INTENSITY-MODULATED RADIOTHERAPY; RADIATION-THERAPY; MONTE-CARLO; CHAMBER; FIELDS;
D O I
10.1186/1748-717X-6-27
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To establish our institutional guideline for IMRT delivery, we statistically evaluated the results of dosimetry quality assurance (DQA) measurements and derived local confidence limits using the concept confidence limit of vertical bar mean vertical bar + 1.96 sigma. Materials and methods: From June 2006 to March 2009, 206 patients with head and neck cancer, prostate cancer, liver cancer, or brain tumor were treated using LINAC-based IMRT technique. In order to determine site specific DQA tolerances at a later stage, a hybrid plan with the same fluence maps as in the treatment plan was generated on CT images of a cylindrical phantom of acryl. Points of measurement using a 0.125 cm(3) ion-chamber were typically located in the region of high and uniform doses. The planar dose distributions perpendicular to the central axis were measured by using a diode array in solid water with all fields delivered, and assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop the local confidence and tolerance limits. The dose differences and gamma pass rates for the different treatment sites were also evaluated in terms of total monitor uints (MU), MU/cGy, and the number of PTV's pieces. Results: The mean values and standard deviations of ion-chamber dosimetry differences between calculated and measured doses were -1.6 +/- 1.2% for H&N cancer, -0.4 +/- 1.2% for prostate and abdominal cancer, and -0.6 +/- 1.5% for brain tumor. Most of measured doses (92.2%) agreed with the calculated doses within a tolerance limit of +/- 3% recommended in the literature. However, we found some systematic under-dosage for all treatment sites. The percentage of points passing the gamma criteria, averaged over all treatment sites was 97.3 +/- 3.7%. The gamma pass rate and the agreement of ion-chamber dosimetry generally decreased with increasing the number of PTV's pieces, the degree of modulation (MU/cGy), and the total MU beyond 700. Our local confidence limits were comparable to those of AAPM TG 119 and ESTRO guidelines that were provided as a practical baseline for center-to-center commissioning comparison. Thus, our institutional confidence and action limits for IMRT delivery were set into the same levels of those guidelines. Discussion and Conclusions: The systematic under-dosage were corrected by tuning up the MLC-related factors (dosimetric gap and transmission) in treatment planning system (TPS) and further by incorporating the tongue-and groove effect into TPS. Institutions that have performed IMRT DQA measurements over a certain period of time need to analyze their accrued DQA data. We confirmed the overall integrity of our IMRT system and established the IMRT delivery guideline during this procedure. Dosimetric corrections for the treatment plans outside of the action level can be suggested only with such rigorous DQA and statistical analysis.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] An analysis of tolerance levels in IMRT quality assurance procedures
    Basran, Parminder S.
    Woo, Milton K.
    MEDICAL PHYSICS, 2008, 35 (06) : 2300 - 2307
  • [42] Factors That Affect the Accuracy of IMRT Quality Assurance Measurements and Their Clinical Significance
    Cao, Y.
    Riegel, A.
    Kapur, A.
    Gill, G.
    Mallalieu, L. Brewster
    Jamshidi, A.
    Potters, L.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [43] Does IMRT Treatment Plan Complexity or Mismatched Dosimetry Data Contribute to Dose Delivery Errors Detected Using an IMRT H&N Quality Assurance Phantom?
    Tonigan, J.
    Kry, S.
    Dong, L.
    Purdie, T.
    White, R.
    Ibbott, G.
    Followill, D.
    MEDICAL PHYSICS, 2011, 38 (06) : 3804 - +
  • [44] Quality assurance for treatment planning dose delivery by 3DRTP and IMRT
    Boyer, AL
    Mok, E
    Luxton, G
    Findley, D
    Chen, Y
    Pawlicki, T
    Ma, CMC
    Xing, L
    Xia, P
    GENERAL PRACTICE OF RADIATION ONCOLOGY PHYSICS IN THE 21ST CENTURY, 2000, (26): : 187 - 230
  • [45] Optimal matching of 3D film dosimetry with calculated dose for IMRT quality assurance
    Park, S
    Shin, D
    Lee, S
    Kim, J
    Kim, D
    Kim, T
    Cho, K
    MEDICAL PHYSICS, 2005, 32 (06) : 1986 - 1986
  • [46] IMPLEMENTATION OF PORTAL DOSIMETRY FOR DAILY QUALITY ASSURANCE, IMRT VERIFICATION AND DEFINITION OF ACTION LEVELS.
    Menhel, J.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S203 - S203
  • [47] Time-resolved dosimetry using a pinpoint ionization chamber as quality assurance for IMRT and VMAT
    Louwe, Robert J. W.
    Wendling, Markus
    Monshouwer, Rene
    Satherley, Thomas
    Day, Rebecca A.
    Greig, Lynne
    MEDICAL PHYSICS, 2015, 42 (04) : 1625 - 1639
  • [48] Quality Assurance for IMRT in Belgium: Verification of the Integral Delivered Dose with Alanine-ESR Dosimetry
    Schaeken, B.
    Van den Weyngaert, D.
    Janssens, H.
    Verellen, D.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [49] A Complexity-Based Correction Factor for a Commercial 4-Dimensional Ionization Chamber Array System on IMRT Dosimetry Quality Assurance Measurements
    Kong, C. W.
    Geng, H.
    Lam, W. W.
    Lo, S. G.
    Cheung, C. W.
    Chiu, T. L.
    Yu, S. K.
    Cheung, K. Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E609 - E610
  • [50] Analysis of patient specific dosimetry quality assurance measurements in intensity modulated radiotherapy: A multi centre study
    Kumar, Rajesh
    Sharma, S. D.
    Deshpande, S.
    Sresty, N. V. N. Madhusudhana
    Bhatt, C. P.
    Amols, Howard I.
    Chourasiya, G.
    Mayya, Y. S.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2014, 10 (03) : 611 - 617