We report a case of intra-articular fracture of a bioabsorbable fixation device from the femoral tunnel in an anterior cruciate ligament reconstruction using a bone-tendon-bone graft. Thirteen months after successful reconstruction surgery, the patient experienced episodes of locking and medial joint pain. There was no history of trauma and no symptoms of instability or swelling. On revision arthroscopy, a fractured tip of a bioabsorbable RIGIDfix cross pin (Mitek, Westwood, MA) was identified in the medial compartment of the knee. There was a broad area of chondral erosion affecting the medial femoral condyle and a small defect to the medial tibial plateau where the loose body had been lodged. The bone-tendon-bone graft was intact without disruption. After arthroscopy, the patient was symptom free for 3 weeks but then developed further symptoms of locking. Magnetic resonance imaging showed another loose body within the knee. A repeat arthroscopy was performed 6 weeks after the earlier procedure and another piece of the polylactic acid RIGIDfix cross pin was removed, this time from the lateral gutter. This case raises concern about the potential for breakage and resultant loose body formation that may occur after bioabsorbable cross-pin fixation and, particularly, the associated chondral damage that can occur if early intervention is not conducted.
引用
收藏
页码:348 / 350
页数:3
相关论文
共 50 条
  • [1] Biodegradable screw presents as a loose intra-articular body after anterior cruciate ligament reconstruction
    MacDonald, P
    Arneja, S
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2003, 19 (06):
  • [3] The outcome of intra-articular debris, following anterior cruciate ligament reconstruction
    Basso, O
    Johnson, DP
    Jewell, F
    Wakeley, CJ
    KNEE, 2001, 8 (03): : 235 - 237
  • [4] The fate of intra-articular debris following arthroscopic anterior cruciate ligament reconstruction
    Wnorowski, DC
    ARTHROSCOPY, 1997, 13 (05): : 620 - 626
  • [5] Intra-articular analgesia after anterior cruciate ligament reconstruction
    Musil, D.
    Sadovsky, P.
    Stehlik, J.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2007, 74 (03) : 182 - 188
  • [6] The effect of intra-articular methadone on postoperative pain following anterior cruciate ligament reconstruction
    Stewart, DJ
    Lambert, EW
    Stack, KM
    Pellegrini, J
    Unger, DV
    Hood, RJ
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (01): : 140 - 144
  • [7] The effect of intra-articular methadone on postoperative pain following anterior cruciate ligament reconstruction
    Muratli, HH
    Muratli, SN
    Bicimoglu, A
    Gogus, N
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08): : 1883 - 1883
  • [8] Intra-articular protrusion of malpositioned transfix implant Following anterior cruciate ligament reconstruction
    Misra, R
    Strover, A
    El-Shazly, M
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (02): : 226 - 226
  • [9] RESULTS OF INTRA-ARTICULAR ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION USING PATELLAR LIGAMENT
    BARTLETT, RJ
    CROWE, R
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1984, 66 (05): : 788 - 789
  • [10] INTRA-ARTICULAR ILIOTIBIAL BAND RECONSTRUCTION FOR ANTERIOR CRUCIATE LIGAMENT INSUFFICIENCY
    YOST, JG
    CHEKOFSKY, K
    SCHOSCHEIM, P
    NOLAN, P
    SLOVIN, H
    SCOTT, WN
    AMERICAN JOURNAL OF SPORTS MEDICINE, 1981, 9 (04): : 220 - 224