Nonconvex prior image constrained compressed sensing (NCPICCS): Theory and simulations on perfusion CT

被引:66
|
作者
Ramirez-Giraldo, J. C. [1 ,2 ]
Trzasko, J. [3 ]
Leng, S. [1 ]
Yu, L. [1 ]
Manduca, A. [2 ,3 ]
McCollough, C. H. [1 ]
机构
[1] Mayo Clin, CT Clin Innovat Ctr, Dept Radiol, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN 55905 USA
[3] Mayo Clin, Ctr Adv Imaging Res, Rochester, MN 55905 USA
关键词
image reconstruction; dynamic CT; perfusion CT; compressed sensing; radiation dose reduction; BEAM COMPUTED-TOMOGRAPHY; HIGH TEMPORAL RESOLUTION; DUAL-ENERGY CT; RADIATION-EXPOSURE; RECONSTRUCTION; PICCS; MRI; FEASIBILITY; ALGORITHMS;
D O I
10.1118/1.3560878
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present and evaluate a new image reconstruction method for dynamic CT based on a nonconvex prior image constrained compressed sensing (NCPICCS) algorithm. The authors systematically compared the undersampling potential, functional information recovery, and solution convergence speed of four compressed sensing (CS) based image reconstruction methods using perfusion CT data: Standard l(1)-based CS, nonconvex CS (NCCS), and l(1)-based and nonconvex CS, including an additional constraint based on a prior image (PICCS and NCPICCS, respectively). Methods: The Shepp-Logan phantom was modified such that its uppermost ellipses changed attenuation through time, simulating both an arterial input function (AIF) and a homogeneous tissue perfusion region. Data were simulated with and without Poisson noise added to the projection data and subsequently reconstructed with all four CS-based methods at four levels of undersampling: 20, 12, 6, and 4 projections. Root mean squared (RMS) error of reconstructed images and recovered time attenuation curves (TACs) were assessed as well as convergence speed. The performance of both PICCS and NCPICCS methods were also evaluated using a kidney perfusion animal experiment data set. Results: All four CS-based methods were able to reconstruct the phantoms with 20 projections, with similar results on the RMS error of the recovered TACs. NCCS allowed accurate reconstructions with as few as 12 projections, PICCS with as few as six projections, and NCPICCS with as few as four projections. These results were consistent for noise-free and noisy data. NCPICCS required the fewest iterations to converge across all simulation conditions, followed by PICCS, NCCS, and then CS. On animal data, at the lowest level of undersampling tested (16 projections), the image quality of NCPICCS was better than PICCS with fewer streaking artifacts, while the TAC accuracy on the selected region of interest was comparable. Conclusions: The authors have presented a novel method for image reconstruction using highly undersampled dynamic CT data. The NCPICCS method takes advantage of the information provided by a prior image, as in PICCS, but employs a more general nonconvex sparsity measure [such as the l(p)-norm (0 <= p <= 1)] rather than the conventional convex l(1)-norm. Despite the lack of guarantees of a globally optimal solution, the proposed nonconvex extension of PICCS consistently allowed for image reconstruction from fewer samples than the analogous l(1)-based PICCS method. Both nonconvex sparsity measures as well as prior image information (when available) significantly reduced the number of iterations required for convergence, potentially providing computational advantages for practical implementation of CS-based image reconstruction techniques. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3560878]
引用
收藏
页码:2157 / 2167
页数:11
相关论文
共 50 条
  • [1] Low Dose Myocardial CT Perfusion Measurements Using Prior Image Constrained Compressed Sensing (PICCS)
    Nett, B.
    Tang, J.
    Speidel, M.
    Chen, G-H
    MEDICAL PHYSICS, 2008, 35 (06) : 2870 - +
  • [2] Prior Image Constrained Compressed Sensing (PICCS)
    Chen, Guang-Hong
    Tang, Jie
    Leng, Shuai
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2008: THE NINTH CONFERENCE ON BIOMEDICAL THERMOACOUSTICS, OPTOACOUSTICS, AND ACOUSTIC-OPTICS, 2008, 6856
  • [3] Perfusion measurements by micro-CT using prior image constrained compressed sensing (PICCS): initial phantom results
    Nett, Brian E.
    Brauweiler, Robert
    Kalender, Willi
    Rowley, Howard
    Chen, Guang-Hong
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (08): : 2333 - 2350
  • [4] A Performance Comparison between Prior Image Constrained Compressed Sensing and ASiR in Pediatric CT
    Tang, J.
    Chen, G-H
    MEDICAL PHYSICS, 2011, 38 (06)
  • [5] Discriminative Prior - Prior Image Constrained Compressed Sensing Reconstruction for Low-Dose CT Imaging
    Yang Chen
    Jin Liu
    Lizhe Xie
    Yining Hu
    Huazhong Shu
    Limin Luo
    Libo Zhang
    Zhiguo Gui
    Gouenou Coatrieux
    Scientific Reports, 7
  • [6] Discriminative Prior - Prior Image Constrained Compressed Sensing Reconstruction for Low-Dose CT Imaging
    Chen, Yang
    Liu, Jin
    Xie, Lizhe
    Hu, Yining
    Shu, Huazhong
    Luo, Limin
    Zhang, Libo
    Gui, Zhiguo
    Coatrieux, Gouenou
    SCIENTIFIC REPORTS, 2017, 7
  • [7] Sub-MSv Cerebral CT Perfusion Imaging Using Prior Image Constrained Compressed Sensing: A Prospective Clinical Trial
    Li, K.
    Yang, P.
    Cathey, M.
    Tang, J.
    Rowley, H.
    Turski, P.
    Prabhakaran, V.
    Strother, C.
    Chen, G. -H.
    MEDICAL PHYSICS, 2016, 43 (06) : 3862 - 3862
  • [8] A new approach to solving the prior image constrained compressed sensing (PICCS) with applications in CT image reconstruction
    Tang, Yuchao
    Long, Chunxiang
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [9] Prior image constrained compressed sensing: a quantitative performance evaluation
    Lauzier, Pascal Theriault
    Tang, Jie
    Chen, Guang-Hong
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [10] GPU implementation of Prior Image Constrained Compressed Sensing (PICCS)
    Nett, Brian E.
    Tang, Jie
    Chen, Guang-Hong
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622