Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions

被引:91
|
作者
Kropff, J. [1 ]
Bruttomesso, D. [2 ]
Doll, W. [3 ]
Farret, A. [4 ,5 ]
Galasso, S. [2 ]
Luijf, Y. M. [1 ]
Mader, J. K. [3 ]
Place, J. [4 ,5 ]
Boscari, F. [2 ]
Pieber, T. R. [3 ]
Renard, E. [4 ,5 ]
DeVries, J. H. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Internal Med, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Padua, Dept Clin & Expt Med, Padua, Italy
[3] Med Univ Graz, Dept Internal Med, Graz, Austria
[4] Montpellier Univ Hosp, Dept Endocrinol, Diabet, Nutr, Montpellier, France
[5] Montpellier Univ Hosp, INSERM Clin Invest Ctr, Montpellier, France
来源
DIABETES OBESITY & METABOLISM | 2015年 / 17卷 / 04期
关键词
type; 1; diabetes; CSII; QUALITY;
D O I
10.1111/dom.12378
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: To assess the accuracy and reliability of the two most widely used continuous glucose monitoring (CGM) systems. Methods: We studied the Dexcom (R) G4 Platinum (DG4P; Dexcom, San Diego, CA, USA) and Medtronic Paradigm Veo Enlite system (ENL; Medtronic, Northridge, CA, USA) CGM systems, in 24 patients with type 1 diabetes. The CGM systems were tested during 6-day home use and a nested 6-h clinical research centre (CRC) visit. During the CRC visit, frequent venous blood glucose samples were used as reference while patients received a meal with an increased insulin bolus to induce an aggravated postprandial glucose nadir. At home, patients performed at least six reference capillary blood measurements per day. A Wilcoxon signed-rank test was performed using all data points >= 15 min apart. Results: The overall mean absolute relative difference (MARD) value [standard deviation (s.d.)] measured at the CRC was 13.6 (11.0)% for the DG4P and 16.6 (13.5)% for the ENL [p<0.0002, confidence interval of difference (CI Delta) 1.7-4.3%, n=530]. The overall MARD assessed at home was 12.2 (12.0)% for the DG4P and 19.9 (20.5)% for the ENL (p<0.0001, CI Delta=5.8-8.7%, n=839). During the CRC visit, the MARD in the hypoglycaemic range [<= 3.9 mmol/l (70mg/dl)], was 17.6 (12.2)% for the DG4P and 24.6 (18.8)% for the ENL (p=0.005, CI Delta 3.1-10.7%, n=117). Both sensors showed higher MARD values during hypoglycaemia than during euglycaemia [3.9-10 mmol/l (70-180 mg/dl)]: for the DG4P 17.6 versus 13.0% and for the ENL 24.6 versus 14.2%. Conclusions: During circumstances of intended use, including both a CRC and home phase, the ENL was noticeably less accurate than the DG4P sensor. Both sensors showed lower accuracy in the hypoglycaemic range. The DG4P was less affected by this negative effect of hypoglycaemia on sensor accuracy than was the ENL.
引用
收藏
页码:343 / 349
页数:7
相关论文
共 50 条
  • [21] Accuracy and Large Inaccuracy of Two Continuous Glucose Monitoring (CGM) Systems
    Leelarathna, Lalantha
    Nodale, Marianna
    Allen, Janet M.
    Elleri, Daniela
    Kumareswaran, Kavita
    Haidar, Ahmad
    Caldwell, Karen
    Wilinska, Mal-Gorzata E.
    Acerini, Carlo L.
    Evans, Mark L.
    Murphy, Helen R.
    Dunger, David B.
    Hovorka, Roman
    DIABETES, 2012, 61 : A1 - A2
  • [22] Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
    Schrangl, Patrick
    Reiterer, Florian
    Heinemann, Lutz
    Freckmann, Guido
    del Re, Luigi
    BIOSENSORS-BASEL, 2018, 8 (02):
  • [23] Continuous Glucose Monitoring and Recreational Scuba Diving in Type 1 Diabetes: Head-to-Head Comparison Between Free Style Libre 3 and Dexcom G7 Performance
    Gamarra, Elena
    Careddu, Giovanni
    Fazi, Andrea
    Turra, Valentina
    Morelli, Ambra
    Camponovo, Chiara
    Trimboli, Pierpaolo
    DIABETES TECHNOLOGY & THERAPEUTICS, 2024, 26 (11) : 829 - 841
  • [24] Discrete choice experiment versus swing-weighting: A head-to-head comparison of diabetic patient preferences for glucose-monitoring devices
    Whichello, Chiara
    Smith, Ian W.
    Veldwijk, Jorien
    de Wit, G. Ardine
    van Molken, Maureen P. M. H. Rutten
    de Bekker-Grob, Esther
    PLOS ONE, 2023, 18 (07):
  • [25] Prospective head-to-head comparison of accuracy of two sequencing platforms for screening for fetal aneuploidy by cell-free DNA: the PEGASUS study
    Rousseau, Francois
    Langlois, Sylvie
    Johnson, Jo-Ann
    Gekas, Jean
    Bujold, Emmanuel
    Audibert, Francois
    Walker, Mark
    Giroux, Sylvie
    Caron, Andre
    Clement, Valerie
    Blais, Jonatan
    MacLeod, Tina
    Moore, Richard
    Gauthier, Julie
    Jouan, Loubna
    Laporte, Alexandre
    Diallo, Ousmane
    Parker, Jeremy
    Swanson, Lucas
    Zhao, Yongjun
    Labelle, Yves
    Giguere, Yves
    Forest, Jean-Claude
    Little, Julian
    Karsan, Aly
    Rouleau, Guy
    EUROPEAN JOURNAL OF HUMAN GENETICS, 2019, 27 (11) : 1701 - 1715
  • [26] Prospective head-to-head comparison of accuracy of two sequencing platforms for screening for fetal aneuploidy by cell-free DNA: the PEGASUS study
    François Rousseau
    Sylvie Langlois
    Jo-Ann Johnson
    Jean Gekas
    Emmanuel Bujold
    François Audibert
    Mark Walker
    Sylvie Giroux
    André Caron
    Valérie Clément
    Jonatan Blais
    Tina MacLeod
    Richard Moore
    Julie Gauthier
    Loubna Jouan
    Alexandre Laporte
    Ousmane Diallo
    Jeremy Parker
    Lucas Swanson
    Yongjun Zhao
    Yves Labelle
    Yves Giguère
    Jean-Claude Forest
    Julian Little
    Aly Karsan
    Guy Rouleau
    European Journal of Human Genetics, 2019, 27 : 1701 - 1715
  • [27] Head-to-head hydrodynamic performance comparison of six foil setups with four and two foil configurations on trimarans in different operational conditions
    Ghadimi, Aliakbar
    Ghassemi, Hassan
    Ghadimi, Parviz
    PHYSICS OF FLUIDS, 2025, 37 (04)
  • [28] Head-to-head comparison of two engineered cardiac grafts for myocardial repair: From scaffold characterization to pre-clinical testing
    Perea-Gil, Isaac
    Galvez-Monton, Carolina
    Prat-Vidal, Cristina
    Jorba, Ignasi
    Segu-Verges, Cristina
    Roura, Santiago
    Soler-Botija, Carolina
    Iborra-Egea, Oriol
    Revuelta-Lopez, Elena
    Fernandez, Marco A.
    Farre, Ramon
    Navajas, Daniel
    Bayes-Genis, Antoni
    SCIENTIFIC REPORTS, 2018, 8
  • [29] Head-to-head comparison of two biological acellular scaffolds for myocardial repair: a pre-clinical myocardial infarction swine model
    Galvez-Monton, C. Carol
    Perea-Gil, I.
    Soler-Botija, C.
    Prat-Vidal, C.
    Diaz-Guemes, I.
    Iborra-Egea, O.
    Crisostomo, V.
    Sanchez-Margallo, F. M.
    Roura, S.
    Bayes-Genis, A.
    EUROPEAN JOURNAL OF HEART FAILURE, 2017, 19 : 170 - 170
  • [30] Head-to-head comparison of two engineered cardiac grafts for myocardial repair: From scaffold characterization to pre-clinical testing
    Isaac Perea-Gil
    Carolina Gálvez-Montón
    Cristina Prat-Vidal
    Ignasi Jorba
    Cristina Segú-Vergés
    Santiago Roura
    Carolina Soler-Botija
    Oriol Iborra-Egea
    Elena Revuelta-López
    Marco A. Fernández
    Ramon Farré
    Daniel Navajas
    Antoni Bayes-Genis
    Scientific Reports, 8