Biomechanical effects of interbody cage height on adjacent segments in patients with lumbar degeneration: a 3D finite element study

被引:6
|
作者
Lu, Xiao [1 ]
Li, Dachuan [1 ]
Wang, Hongli [1 ]
Xia, Xinlei [1 ]
Ma, Xiaosheng [1 ]
Lv, Feizhou [1 ]
Zou, Fei [1 ]
Jiang, Jianyuan [1 ]
机构
[1] Fudan Univ, Huashan Hosp, Dept Orthoped, 12 Middle Wulumuqi Rd, Shanghai 200040, Peoples R China
基金
中国国家自然科学基金;
关键词
TLIF; Interbody cage; Intervertebral height; Degenerative lumbar spine; Biomechanics; INTERVERTEBRAL DISC DEGENERATION; ANTERIOR CERVICAL DISKECTOMY; LOW-BACK-PAIN; MECHANICAL-BEHAVIOR; FUSION; DISEASE; SPINE; DISTRACTION; COMPRESSION; LEVEL;
D O I
10.1186/s13018-022-03220-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective To investigate the biomechanical effects of interbody cage height on adjacent segments in patients with lumbar degeneration undergoing transforaminal lumbar interbody fusion (TLIF) surgery, so as to provide references for selection of interbody cage. Methods The finite element model of normal lower lumbar spine (L3-S1) was built and validated, then constructed three different degenerative segments in L3-L4, and the cages with different height (8, 10, 12, 14 mm) were implanted into L4-L5 disc. All the twelve models were loaded with pure moment of 7.5 N m to produce flexion, extension, lateral bending and axial rotation motions on lumbar spine, and the effects of cage height on range of motion (RoM) and intervertebral pressure in lumbar spine were investigated. Results The RoM of adjacent segments and the maximum stress of intervertebral discs increased with the increase in cage height, but this trend was not obvious in mild and moderate degeneration groups. After implantation of four different height cages (8, 10, 12, 14 mm), the RoM of L3/L4 segment reached the maximum during extension. The RoM of mild degeneration group was 2.07 degrees, 2.45 degrees, 2.48 degrees, 2.54 degrees, that of moderate degeneration group was 1.79 degrees, 1.97 degrees, 2.05 degrees, 2.05 degrees, and that of severe degeneration group was 1.43 degrees, 1.66 degrees, 1.74 degrees, 1.74 degrees. The stress of L3-L4 intervertebral disc reached the maximum during flexion. The maximum stress of L3-L4 intervertebral disc was 20.16 MPa, 20.28 MPa, 20.31 MPa and 20.33 MPa in the mild group, 20.58 MPa, 20.66 MPa, 20.71 MPa and 20.75 MPa in the moderate group, and 21.27 MPa, 21.40 MPa, 21.50 MPa and 21.60 MPa in the severe group. Conclusion For patients with mild-to-moderate lumbar degenerative disease who need to undergo TLIF surgery, it is recommended that the height of fusion cage should not exceed the original intervertebral space height by 2 mm, while for patients with severe degeneration, a fusion cage close to the original intervertebral height should be selected as far as possible, and the intervertebral space should not be overstretched.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Biomechanical evaluation of modified and traditional cortical bone trajectory technique on adjacent segment degeneration in transforaminal lumbar interbody fusion-finite element analysis
    Tuoheti, Abudusalamu
    Xiao, Yang
    Wang, Yixi
    Maimaiti, Abulikemu
    Zhang, Rui
    Kahaer, Alafate
    Tuoheti, Abuduaini
    Wu, Xianghui
    Rexiti, Paerhati
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
  • [32] Does oblique lumbar interbody fusion promote adjacent degeneration in degenerative disc disease: A finite element analysis
    Du, Cheng-Fei
    Cai, Xin-Yi
    Gui, Wu
    Sun, Meng-Si
    Liu, Zi-Xuan
    Liu, Chun-Jie
    Zhang, Chun-Qiu
    Huang, Yun-Peng
    COMPUTERS IN BIOLOGY AND MEDICINE, 2021, 128
  • [33] Does oblique lumbar interbody fusion promote adjacent degeneration in degenerative disc disease: A finite element analysis
    Du C.-F.
    Cai X.-Y.
    Gui W.
    Sun M.-S.
    Liu Z.-X.
    Liu C.-J.
    Zhang C.-Q.
    Huang Y.-P.
    Computers in Biology and Medicine, 2021, 128
  • [34] Biomechanical effects of an oblique lumbar interbody fusion combined with posterior augmentation: a finite element analysis
    Shengjia Huang
    Shaoxiong Min
    Suwei Wang
    Anmin Jin
    BMC Musculoskeletal Disorders, 23
  • [35] Finite element biomechanical analysis of 3D printed intervertebral fusion cage in osteoporotic population
    Wu, Jincheng
    Miao, Jun
    Chen, Guangdong
    Xu, Hanpeng
    Wen, Wangqiang
    Xu, Haoxiang
    Liu, Lizhu
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
  • [36] Biomechanical effects of an oblique lumbar interbody fusion combined with posterior augmentation: a finite element analysis
    Huang, Shengjia
    Min, Shaoxiong
    Wang, Suwei
    Jin, Anmin
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [37] Finite element biomechanical analysis of 3D printed intervertebral fusion cage in osteoporotic population
    Jincheng Wu
    Jun Miao
    Guangdong Chen
    Hanpeng Xu
    Wangqiang Wen
    Haoxiang Xu
    Lizhu Liu
    BMC Musculoskeletal Disorders, 25
  • [38] Impact of cage position on biomechanical performance of stand-alone lateral lumbar interbody fusion: a finite element analysis
    Chong Nan
    Zhanbei Ma
    Yuxiu Liu
    Liang Ma
    Jiaqi Li
    Wei Zhang
    BMC Musculoskeletal Disorders, 23
  • [39] Impact of cage position on biomechanical performance of stand-alone lateral lumbar interbody fusion: a finite element analysis
    Nan, Chong
    Ma, Zhanbei
    Liu, Yuxiu
    Ma, Liang
    Li, Jiaqi
    Zhang, Wei
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [40] Biomechanical analysis of 3D printed porous extremely-low modulus Ti-24Nb-4Zr-8Sn lumbar interbody fusion cage-A finite element study
    Shuai, Huang
    Xuhui, Jin
    Lei, Li
    Shujun, Li
    Cheng, Li
    MATERIALS TECHNOLOGY, 2024, 39 (01)