Triple Leads Programmed to Perform as Longitudinal Guarded Cathodes in Spinal Cord Stimulation: A Modeling Study

被引:17
|
作者
Sankarasubramanian, Vishwanath [1 ]
Buitenweg, Jan R. [1 ]
Holsheimer, Jan [1 ]
Veltink, Peter [1 ]
机构
[1] Univ Twente, Inst Biomed Technol & Tech Med, MIRA, NL-7500 AE Enschede, Netherlands
来源
NEUROMODULATION | 2011年 / 14卷 / 05期
关键词
Constant anodal current ratio; guarded cathodes; low-thoracic; paresthesia; spinal cord stimulation; variable anodal current ratio; LOW-BACK-PAIN; PERCUTANEOUS ELECTRODES; MYELINATED NERVE; COLUMN FIBERS; DORSAL; SYSTEM; SUPERIORITY; EXCITATION; MANAGEMENT; GEOMETRY;
D O I
10.1111/j.1525-1403.2011.00383.x
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective: In spinal cord stimulation, neurosurgeons increasingly tend to implant dual leads. Dual leads (longitudinal bipole/tripole) provide medio-lateral control over the recruited dorsal column (DC) area by steering the injected cathodal currents. However, the DC recruited area is suboptimal when dual aligned leads straddling the midline programmed as longitudinal guarded cathodes (+-+) are used instead of a single lead placed over the spinal cord midline with the same configuration. As a potential improvement, an additional third lead between the two aligned leads is modeled to maximize the medio-lateral extent of the DCs at the low-thoracic vertebral region (T10-T12). Methods and Materials: The University of Twente Spinal Cord Stimulation software (UT-SCS) is used in this modeling study. Longitudinal guarded cathodes were modeled on the low-thoracic vertebral region (T10-T12) using percutaneous triple lead configurations. The central lead was modeled over the spinal cord midline and the two lateral leads were modeled at several transverse distances to the midline lead. Medio-lateral field steering was performed with the midline lead and the second lead on each side to achieve constant anodal current ratios and variable anodal current ratios. Results: Reducing the transverse lead separation resulted in increasing the depths and widths of the recruited DC area. The triple lead configuration with the least transverse separation had the largest DC recruited area and usage range. The maximum DC recruited area (in terms of both depth and width) was always found to be larger under variable anodal current ratio than constant anodal current ratio conditions. Conclusions: Triple leads programmed to perform as longitudinal guarded cathodes provide more postoperative flexibility than single and dual leads in covering a larger width of the low-thoracic DCs. The transverse separation between the leads is a major determinant of the area and distribution of paresthesia.
引用
收藏
页码:401 / 410
页数:10
相关论文
共 50 条
  • [31] A longitudinal study of depression in survivors of spinal cord injury
    Saunders, L. L.
    Krause, J. S.
    Focht, K. L.
    SPINAL CORD, 2012, 50 (01) : 72 - 77
  • [32] A longitudinal study of depression in survivors of spinal cord injury
    L L Saunders
    J S Krause
    K L Focht
    Spinal Cord, 2012, 50 : 72 - 77
  • [33] Percutaneous adjustment method for transversely migrated spinal cord stimulation leads: a technical report
    Sang Yoon Jeon
    Jae Young Ji
    Sie Hyeon Yoo
    Jin Young Chon
    Sung Hoon Jung
    Ho Sik Moon
    Journal of Anesthesia, 2015, 29 : 953 - 956
  • [34] Percutaneous adjustment method for transversely migrated spinal cord stimulation leads: a technical report
    Jeon, Sang Yoon
    Ji, Jae Young
    Yoo, Sie Hyeon
    Chon, Jin Young
    Jung, Sung Hoon
    Moon, Ho Sik
    JOURNAL OF ANESTHESIA, 2015, 29 (06) : 953 - 956
  • [35] Electrophysiologic Monitoring for Placement of Laminectomy Leads for Spinal Cord Stimulation Under General Anesthesia
    Air, Ellen L.
    Toczyl, Greg R.
    Mandybur, George T.
    NEUROMODULATION, 2012, 15 (06): : 573 - 580
  • [36] Spinal cord stimulation with percutaneous leads after loss of coverage with implanted surgical lead
    Yakovlev, Alexander E.
    Resch, Beth E.
    NEUROMODULATION, 2010, 13 (02): : 117 - 120
  • [37] Incidence and Avoidance of Neurologic Complications with Paddle Type Spinal Cord Stimulation Leads COMMENTS
    Raso, Lou
    Deer, Tim
    NEUROMODULATION, 2011, 14 (05): : 422 - 422
  • [38] Source localization in the spinal cord: A modeling study
    Moffitt, MA
    Grill, WM
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 1977 - 1978
  • [39] Spinal cord stimulation: Cost-benefit study
    Budd, K
    NEUROMODULATION, 2002, 5 (02): : 75 - 78
  • [40] Analysis of spinal cord stimulation and design of epidural electrodes by computer modeling
    Holsheimer, J
    Struijk, JJ
    Wesselink, WA
    NEUROMODULATION, 1998, 1 (01): : 14 - 18