Detection of drugs in urine by ambient direct ionization mass spectrometry

被引:3
|
作者
Xiong Shiling [1 ]
Hong Huanhuan [1 ,2 ]
Wen Luhong [1 ,2 ]
Hu Shundi [1 ,2 ]
Chen Anqi [1 ,2 ]
Xiong Wei [1 ,2 ]
Chen La [1 ,2 ]
机构
[1] Ningbo Univ, Res Inst Adv Technol, Ningbo 315211, Peoples R China
[2] China Innovat Instrument Co Ltd, Ningbo 315100, Peoples R China
关键词
ambient direct ionization; portable mass spectrometer; drug; urine; rapid detection; RAPID DETECTION; SULFONAMIDES;
D O I
10.3724/SP.J.1123.2022.01013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
According to the Report of Drug Situation in China (2020), the growth rate of the number of drug abusers in China has decreased, but the number of drug abusers is still large. An efficient screening method is necessary for controlling drug abuse. As an important type of biological sample, urine is widely used for the rapid screening of drug addicts. However, because of the complex composition, low content, and strong interference from the body' s metabolism, the detection of drugs in urine remains a challenge. Traditional rapid screening techniques such as immunocolloidal gold analysis have a high false positive rate and insufficient quantitative capability. In addition, laboratory mass spectrometry methods require complicated time. consuming sample pre. processing and strict environmental conditions, and hence, are un. suitable for on. site rapid analysis. In recent years, various direct ionization mass spectrometry techniques such as direct analysis in real time (DART), desorption electrospray ionization (DESI), and dielectric barrier discharge ionization (DBDI) have advanced rapidly. These tech. niques have been applied to public safety, food safety, environmental detection, etc. In con. trast to traditional ionization mass spectrometry methods, these direct ionization techniques allow for the in situ analysis of samples with simple or no pretreatment; moreover, they have the advantages of high analytical efficiency and sensitivity. In particular, pulsed electrospray ionization has the characteristics of less sample demand, compact, lightweight equipment, and no carrier gas. This paper presents a rapid method based on pulsed electrospray ionization mass spectrometry for the detection of urine samples. A rapid detection platform comprising a probe electrospray ionization source, a portable linear ion trap mass spectrometer (MS), and their coupling interface is adopted. The probe electrospray ion source includes a conducting metal wire, plastic handle, and silica glass capillary, whose tip has an inner diameter of 50 mu m. The guide rail at the coupling interface is used to align the probe with the sample inlet of the porta. ble mass spectrometer and maintain a distance of 10 mm between the probe tip and the sample inlet of the MS. The spray voltage of the probe electrospray ion source and the temperature of the MS inlet capillary are optimized at 1.8 kV and 205 degrees C, respectively. In addition, rapid and efficient pretreatment techniques for urine samples have been developed. Buffer salts used for pH regulation and liquid.liquid extraction based on ethyl acetate were adopted for the pretreat. ment process. The linearity of the detection ability and the linear ranges of various drug.spiked solutions were also investigated. The results showed that the correlation coefficients for the quantitative detection of methamphetamine, ketamine, methylenedioxymethamphetamine (MDMA), and cocaine were greater than 0.99 at concentrations ranging from 1 to 100 ng / mL. Moreover, the limits of detection (LODs) for the five conventional drug. spiked urine were 0.5-30 ng / mL. The spiked recoveries ranged from 56.1% to 103.7%, with relative standard devia. tions (RSDs) of 9.0% - 27.8%, implying that the combination of the instruments and the pre. treatment method can lead to good accuracy. To validate the performance of the rapid detection method, 40 positive and 110 negative urine samples were tested and analyzed. The overall accu. racy was over 99%, and the five conventional drugs in urine samples could be detected within 20 s. The research findings of this work could promote the development of rapid detection tech. nology, accelerate the popularization and application of ambient direct ionization mass spec. trometry, and improve the services of on. site law enforcement.
引用
收藏
页码:677 / 683
页数:7
相关论文
共 19 条
  • [11] Development of a dielectric barrier discharge ion source for ambient mass spectrometry
    Na, Na
    Zhao, Mengxia
    Zhang, Sichun
    Yang, Chengdui
    Zhang, Xinrong
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (10) : 1859 - 1862
  • [12] [钱振华 Qian Zhenhua], 2021, [质谱学报, Journal of Chinese Mass Spectrometry Society], V42, P84
  • [13] [尚宇瀚 Shang Yuhan], 2021, [质谱学报, Journal of Chinese Mass Spectrometry Society], V42, P439
  • [14] Gas Flow in the Capillary of the Atmosphere-to-Vacuum Interface of Mass Spectrometers
    Skoblin, Michael
    Chudinov, Alexey
    Soulimenkov, Ilia
    Brusov, Vladimir
    Kozlovskiy, Viacheslav
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2017, 28 (10) : 2132 - 2142
  • [15] Monodisperse electrosprays of low electric conductivity liquids in the cone-jet mode
    Tang, KQ
    Gomez, A
    [J]. JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1996, 184 (02) : 500 - 511
  • [16] Wang L, 2015, CHINESE J ANAL CHEM, V43, P33
  • [17] Wei Z W., 2016, THESIS TSINGHUA U
  • [18] Pulsed Direct Current Electrospray: Enabling Systematic Analysis of Small Volume Sample by Boosting Sample Economy
    Wei, Zhenwei
    Xiong, Xingchuang
    Guo, Chengan
    Si, Xingyu
    Zhao, Yaoyao
    He, Muyi
    Yang, Chengdui
    Xu, Wei
    Tang, Fei
    Fang, Xiang
    Zhang, Sichun
    Zhang, Xinrong
    [J]. ANALYTICAL CHEMISTRY, 2015, 87 (22) : 11242 - 11248
  • [19] Rapid detection of sulfonamides by heat assisted desorption.dielectric barrier discharge ionization mass spectrometry
    Ye Qian
    Zhu Qiumeng
    Zhou Feng
    Wu Jianping
    Yan Feng
    Zhao Peng
    Wen Luhong
    [J]. CHINESE JOURNAL OF CHROMATOGRAPHY, 2020, 38 (07) : 868 - 874