Development of a lactate biosensor based on conducting copolymer bound lactate oxidase

被引:91
|
作者
Suman, S
Singhal, R
Sharma, AL
Malthotra, BD
Pundir, CS [1 ]
机构
[1] Maharshi Dayanand Univ, Dept Biosci, Biochem Res Lab, Rohtak 124001, Haryana, India
[2] Natl Phys Lab, Biomol Elect & Conducting Polymer Res Grp, New Delhi 110012, India
关键词
lactate; lactate oxidase; immobilization; polyaniline-co-fluoroaniline; lactate biosensor; serum;
D O I
10.1016/j.snb.2004.12.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric enzyme electrode was developed for determination of lactate in serum. To prepare this electrode, commercial lactate oxidase from Pediococcus species has been immobilized through glutaraldehyde coupling onto polyaniline-co-fluoroaniline film deposited on an Indium tin oxide (ITO) coated glass plate. This plate acted as working electrode when combined with Pt electrode as counter electrode to the electrometer for the development of a biosensor. The method is based on generation of electrons from H2O2 which is formed from lactic acid by immobilized lactate oxidase. The concentration of lactic acid is directly proportional to the current measured. The enzyme electrode showed optimum response when operated at 42 degrees C in 0.05 M, sodium phosphate buffer pH 6.5 for 50s, The biosensor showed a good performance with a linear response range from 0.1 to 5.5 mM/l. The minimum detection limit of the method is 0.1 mM/l and sensitivity of the sensor is 1.18 mu A/mM/l lactate. This electrode was employed for determination of lactate in serum. The serum values in healthy and diseased persons were in the range 0.51-2.9 and 5.0-15.0 mM. respectively. The analytical recovery of added lactic acid was 71%. Within batch and between batch CV were < 4 and < 5%. respectively. Among the various serum substances tested only 8-hydroxyquinoline, urea, ammonium molybdate and uric acid caused 64, 38, 34 and 31% inhibition, respectively. The electrode Was used 150 times over 26 days without any considerable loss of activity, when stored at 4 degrees C. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:768 / 772
页数:5
相关论文
共 50 条
  • [21] Engineering of Lactate Oxidase and Dehydrogenase for Development of Direct Electron Transfer Type Lactate Sensor and Application for Multiplexed Biosensor System
    la Belle J.T.
    Denki Kagaku, 2024, 92 (02) : 103 - 108
  • [22] Development of a potentiometric biosensor based on nanostructured surface for lactate determination
    Lupu, A.
    Valsesia, A.
    Bretagnol, F.
    Colpo, P.
    Rossi, F.
    SENSORS AND ACTUATORS B-CHEMICAL, 2007, 127 (02) : 606 - 612
  • [23] Development of Potentiometric Lactate Biosensor Based on Composite pH Sensor
    Uygun, Hilmiye Deniz Ertugrul
    Tinkilic, Nihat
    Attar, Azade
    Isildak, Ibrahim
    JOURNAL OF NEW MATERIALS FOR ELECTROCHEMICAL SYSTEMS, 2016, 19 (03) : 151 - 156
  • [24] Lactate oxidase and glucose oxidase biosensor with ultra-fine Au particles
    Meng, Xian-Wei
    Gao, Feng-Tang
    Ren, Xiang-Ling
    Wei, Zheng
    Tang, Fang-Qiong
    Guocheng Gongcheng Xuebao/The Chinese Journal of Process Engineering, 2002, 2 (04):
  • [25] Dataset on fabrication of an improved L-lactate biosensor based on lactate oxidase/cMWCNT/CuNPs/PANI modified PG electrode
    Dagar, Kusum
    Pundir, C. S.
    DATA IN BRIEF, 2018, 17 : 1163 - 1167
  • [26] ELECTROPOLYMERIZED O-PHENYLENEDIAMINE FILM AS MEANS OF IMMOBILISING LACTATE OXIDASE FOR A L-LACTATE BIOSENSOR
    DEMPSEY, E
    WANG, J
    SMYTH, MR
    TALANTA, 1993, 40 (03) : 445 - 451
  • [27] Sol-Gel Immobilization of Lactate Oxidase from Organic Solvent: Toward the Advanced Lactate Biosensor
    Yashina, Eugenia I.
    Borisova, Anastasiya V.
    Karyakina, Elena E.
    Shchegolikhina, Olga I.
    Vagin, Mikhail Yu.
    Sakharov, Dmitry A.
    Tonevitsky, Alexandr G.
    Karyakin, Arkady A.
    ANALYTICAL CHEMISTRY, 2010, 82 (05) : 1601 - 1604
  • [28] Development of a disposable biosensor for lactate monitoring in saliva
    Petropoulos, Konstantinos
    Piermarini, Silvia
    Bernardini, Sergio
    Palleschi, Giuseppe
    Moscone, Danila
    SENSORS AND ACTUATORS B-CHEMICAL, 2016, 237 : 8 - 15
  • [29] Current Status of Fluorescent Lactate Biosensor Development
    Nasu, Yusuke
    BUNSEKI KAGAKU, 2023, 72 (12) : 493 - 496
  • [30] Prussian Blue- and lactate oxidase-based amperometric biosensor for lactic acid
    Garjonyte, R
    Yigzaw, Y
    Meskys, R
    Malinauskas, A
    Gorton, L
    SENSORS AND ACTUATORS B-CHEMICAL, 2001, 79 (01) : 33 - 38