Effects of off-resonance spins on the performance of the modulated gradient spin echo sequence

被引:4
|
作者
Sersa, Igor [1 ]
Bajd, Franci
Mohoric, Ales
机构
[1] Jozef Stefan Inst, Jamova 39, Ljubljana 1000, Slovenia
关键词
Diffusion spectrum; Modulated gradients; CPMG sequence; Coherence pathways; Signal attenuation; AUTOCORRELATION SPECTRA; DIFFUSION MEASUREMENTS; POROUS-MEDIA; NMR; CPMG; PROPAGATOR; PULSES; MOTION; WATER; TIME;
D O I
10.1016/j.jmr.2016.07.001
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Translational molecular dynamics in various materials can also be studied by diffusion spectra. These can be measured by a constant gradient variant of the modulated gradient spin echo (MGSE) sequence which is composed of a CPMG RF pulse train superimposed to a constant magnetic field gradient. The application of the RF train makes the effective gradient oscillating thus enabling measurements of diffusion spectra in a wide range of frequencies. However, seemingly straightforward implementation of the MGSE sequence proved to be complicated and can give overestimated results for diffusion if not interpreted correctly. In this study, unrestricted diffusion in water and other characteristic materials was analyzed by the MGSE sequence in the frequency range 50-3000 Hz using a 6 T/m diffusion probe. First, it was shown that the MGSE echo train acquired from the entire sample decays faster than the train acquired only from a narrow band at zero frequency of the sample. Then, it was shown that the decay rate is dependent on the band's off-resonance characterized by the ratio Delta omega(0)/omega(1) and that with higher off resonances the decay is faster. The faster decay therefore corresponds to a higher diffusion coefficient if the diffusion is calculated using standard Stejskal-Tanner formula. The result can be explained by complex coherence pathways contributing to the MGSE echo signals when vertical bar Delta omega(0)vertical bar/omega(1) > 0. In a magnetic field gradient, all the pathways are more diffusion attenuated than the direct coherence pathway and therefore decay faster, which leads to an overestimation of the diffusion coefficient. A solution to this problem was found in an efficient off-resonance signal reduction by using only zero frequency filtered MGSE echo train signals. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:77 / 86
页数:10
相关论文
共 50 条
  • [41] Exact correction of sharply varying off-resonance effects in spiral MRI
    Makhijani, Mahender K.
    Nayak, Krishna S.
    2006 3RD IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1-3, 2006, : 730 - +
  • [42] Low rank off-resonance correction for double half-echo k-space acquisitions
    Bydder, Mark
    Ali, Fadil
    Saucedo, Andres
    Ghodrati, Vahid
    Samsonov, Alexei
    Akhtari, Massoud
    Wang, Chencai
    Hagiwara, Akifumi
    Yao, Jingwen
    Ellingson, Ben
    MAGNETIC RESONANCE IMAGING, 2022, 94 : 43 - 47
  • [43] Off-resonance long-period fiber gratings and spin-optics response
    Fridman, Moti
    2018 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2018,
  • [44] Improved 3D Phase Contrast MRI With Off-Resonance Corrected Dual Echo VIPR
    Johnson, Kevin M.
    Lum, Darren P.
    Turski, Patrick A.
    Block, Walter F.
    Mistretta, Charles A.
    Wieben, Oliver
    MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (06) : 1329 - 1336
  • [45] The importance of k-space trajectory on off-resonance artifact in segmented echo-planar imaging
    Bender, Jacob A.
    Ahmad, Rizwan
    Simonetti, Orlando P.
    CONCEPTS IN MAGNETIC RESONANCE PART A, 2013, 42 (02) : 23 - 31
  • [46] Improving the off-resonance energy harvesting performance using dynamic magnetic preloading
    Feng Qian
    Shengxi Zhou
    Lei Zuo
    Acta Mechanica Sinica, 2020, 36 : 624 - 634
  • [47] OFF-RESONANCE ROTATING-FRAME SPIN-LATTICE RELAXATION OF QUADRUPOLAR (SPIN-1) NUCLEI
    RYDZEWSKI, JM
    SCHLEICH, T
    JOURNAL OF MAGNETIC RESONANCE SERIES B, 1994, 105 (02): : 129 - 136
  • [48] Improving the off-resonance energy harvesting performance using dynamic magnetic preloading
    Qian, Feng
    Zhou, Shengxi
    Zuo, Lei
    ACTA MECHANICA SINICA, 2020, 36 (03) : 624 - 634
  • [49] OFF-RESONANCE PERFORMANCE OF ADIABATIC EXCITATION PULSES AND THEIR USE FOR SELECTIVE PEAK SUPPRESSION
    BANSAL, N
    JOURNAL OF MAGNETIC RESONANCE SERIES B, 1993, 102 (01): : 73 - 76
  • [50] Exploiting dynamic interaction of magnets to enhance off-resonance energy harvesting performance
    Qian, Feng
    Zhou, Shengxi
    Zuo, Lei
    ACTIVE AND PASSIVE SMART STRUCTURES AND INTEGRATED SYSTEMS XII, 2018, 10595