Endovascular Stent-Graft Repair of Late Pseudoaneurysms After Surgery for Aortic Coarctation

被引:11
|
作者
Zipfel, Burkhart [1 ]
Ewert, Peter
Buz, Semih
Abd El Al, Alaa
Hammerschmidt, Robert
Hetzer, Roland
机构
[1] Deutsch Herzzentrum Berlin, Dept Cardiothorac & Vasc Surg, D-13353 Berlin, Germany
来源
ANNALS OF THORACIC SURGERY | 2011年 / 91卷 / 01期
关键词
SURGICAL REPAIR; AORTOBRONCHIAL FISTULA; THORACIC AORTA; ANEURYSMS; AORTOPLASTY; MANAGEMENT;
D O I
10.1016/j.athoracsur.2010.08.053
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. We analyzed the potential of endovascular stent grafts to treat late aortic pseudoaneurysms after coarctation repair. Methods. Eight patients (7 male; age 28 to 58, mean 43 years) presented with aortic pseudoaneurysms after primary repair performed at the age of 5 to 27 (mean 12) years; 2 patients had rupture with hemorrhagic shock. The mean interval between the procedures was 31 (19 to 42) years; one patient had 3 previous operations. This subset represents 2.2% of our overall experience in thoracic endovascular repair (n = 368). Thoracic endografts were implanted using the transfemoral technique. Custom-made reverse tapered stent grafts were used in 4 cases. The left subclavian artery (LSA) was covered in 5 patients. Protective transposition of the left subclavian artery was performed in 4 patients. Results. Hospital mortality was 12.5%; 1 patient died from secondary rupture after emergency repair. Primary complete exclusion of the aneurysm was achieved in 6 patients. Secondary exclusion after implantation of a second stent graft was successful in the second rupture patient. No endoleak was present at discharge. All discharged patients are alive after 8 to 63 (mean 36) months. Follow-up computed tomography or transesophageal echocardiography revealed no secondary endoleaks or late expansion. The pseudoaneurysms had shrunk completely in 3 patients, were reduced in size in 2 patients, and remained unchanged in 2 patients. No late secondary interventions were noted. Conclusions. Single piece, reversed, tapered stent grafts adapt better to the special anatomy of the hypoplastic aortic arch. Midterm results are excellent; complete shrinkage in 50% is remarkable. However, the long-term behavior of these implants in young patients requires further evaluation and surveillance. (Ann Thorac Surg 2011;91:85-91) (C) 2011 by The Society of Thoracic Surgeons
引用
收藏
页码:85 / 91
页数:7
相关论文
共 50 条
  • [21] Stent-graft detachment from aortic wall after stent-graft repair of acute aortic dissection
    Kato, N
    Shimono, T
    Hirano, T
    Ishida, M
    Takeda, K
    JOURNAL OF VASCULAR SURGERY, 2003, 38 (05) : 1130 - 1131
  • [22] Factors and forces influencing stent-graft migration after endovascular aortic aneurysm repair
    Mohan, IV
    Harris, PL
    van Marrewijk, CJ
    Laheij, RJ
    How, TV
    JOURNAL OF ENDOVASCULAR THERAPY, 2002, 9 (06) : 748 - 755
  • [23] A Stent-Graft Fracture After a 14-Year Thoracic Endovascular Aortic Repair
    Wu, Wenhui
    Pu, Junzhou
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 43 (09) : 1413 - 1415
  • [24] 3D imaging of the stent-graft after endovascular repair of aortic diseases
    Czermak, BV
    Perkmann, R
    Rieger, M
    Maas, M
    Mallouhi, A
    Jaschke, WR
    RADIOLOGY, 2001, 221 : 262 - 263
  • [25] A Stent-Graft Fracture After a 14-Year Thoracic Endovascular Aortic Repair
    Wenhui Wu
    Junzhou Pu
    CardioVascular and Interventional Radiology, 2020, 43 : 1413 - 1415
  • [26] Thoracic EndoVascular Aortic/Aneurysm Stent-Graft Repair (TEVAR) primer
    Janko, Matthew
    Louis, Clauden
    JOURNAL OF CARDIAC SURGERY, 2022, 37 (07) : 2212 - 2216
  • [27] Endovascular stent graft repair for infected thoracic aortic pseudoaneurysms - a durable option?
    Ting, Albert C. W.
    Cheng, Stephen W. K.
    Ho, Pei
    Poon, Jensen T. C.
    JOURNAL OF VASCULAR SURGERY, 2006, 44 (04) : 701 - 705
  • [28] Results of Endovascular Abdominal Aortic Aneurysm Repair with the Zenith stent-graft
    Bos, W. T. G. J.
    Tielliu, I. F. J.
    Zeebregts, C. J.
    Prins, T. R.
    van den Dungen, J. J. A. M.
    Verhoeven, E. L. G.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2008, 36 (06) : 653 - 660
  • [29] Development of a branched stent-graft for endovascular repair of aortic arch aneurysms
    Chuter, TAM
    Buck, DG
    Schneider, DB
    Reilly, LM
    Messina, LM
    JOURNAL OF ENDOVASCULAR THERAPY, 2003, 10 (05) : 940 - 945
  • [30] Endovascular Repair of Ascending Aortic Rupture: Effectiveness of a Fenestrated Stent-Graft
    Uchida, Keiji
    Imoto, Kiyotaka
    Yanagi, Hiromasa
    Machida, Daisuke
    Okiyama, Makoto
    Yasuda, Shota
    Sugiura, Tadahisa
    Kawaguchi, Satoshi
    Yokoi, Yoshihiko
    Shigematsu, Hiroshi
    Masuda, Munetaka
    JOURNAL OF ENDOVASCULAR THERAPY, 2010, 17 (03) : 395 - 398