Comparison of the standards for air kerma of the KRISS and the BIPM for 60Co gamma radiation

被引:4
|
作者
Kessler, C. [1 ]
Yi, Chul-Young [2 ]
Chung, Jae-Pil [2 ]
Roger, P. [1 ]
Allisy-Roberts, P. J. [1 ]
机构
[1] Bur Int Poids & Mesures, F-92312 Sevres, France
[2] Korea Res Inst Stand & Sci, Taejon 305340, South Korea
关键词
CHAMBER;
D O I
10.1088/0026-1394/47/1A/06021
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A direct comparison of the standards for air kerma of the Korea Research Institute of Standards and Science (KRISS), Korea, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the Co-60 radiation beam of the BIPM in 2010. The result, expressed as a ratio of the KRISS and the BIPM standards for air kerma, is 0.9995 with a combined standard uncertainty of 1.6 x 10(-3). The degrees of equivalence between the KRISS and the other participants in the key comparison, designated as BIPM. RI(I)-K1, have been calculated and the results are presented in the form of a matrix. A graphical presentation is also given.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the ARPANSA, Australia and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Oliver, C.
    Bailey, T.
    Harty, P.
    Butler, D.
    METROLOGIA, 2021, 58 (1A)
  • [42] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the BFKH, Hungary, and the BIPM in 60Co gamma radiation
    Kessler C.
    Burns D.
    Finta V.
    Metrologia, 2023, 60 (1 A)
  • [43] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the NIM, China and the BIPM in 60Co gamma radiation
    Kessler C.
    Burns D.
    Jin J.
    Zhang X.
    Wang Z.P.
    Liu F.B.
    Huang J.
    Qi Y.P.
    Yang X.Y.
    Wang K.
    Metrologia, 2022, 59 (1 A)
  • [44] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the GUM, Poland and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Knyziak, A.
    Szymko, M.
    Derlacinski, M.
    METROLOGIA, 2021, 58 (1A)
  • [45] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the PTB, Germany and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Kapsch, R. -P.
    Krauss, A.
    METROLOGIA, 2016, 53
  • [46] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the PTB, Germany and the BIPM in 60Co gamma radiation
    Kessler C.
    Burns D.T.
    Kapsch R.-P.
    Krauss A.
    1600, IOP Publishing Ltd (53):
  • [47] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Voros, S.
    Hofstetter-Boillat, B.
    METROLOGIA, 2015, 52
  • [48] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the BEV, Austria and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Steurer, A.
    Tiefenboeck, W.
    Merza, V.
    METROLOGIA, 2021, 58 (1A)
  • [49] Key comparison BIPM.RI(I)-K5 of the air kerma standards of the VSL and the BIPM in 137Cs gamma radiation
    Kessler, C.
    Burns, D. T.
    Jansen, B. J.
    De Pooter, J. A.
    METROLOGIA, 2016, 53
  • [50] Change of air kerma standards of the BEV for 137Cs and 60Co γ-rays
    Witzani, J
    Gabris, F
    Leitner, A
    METROLOGIA, 2004, 41 (01) : L1 - L2