Multicenter trial of the Jostent balloon-expandable stent-graft in renal and iliac artery lesions

被引:0
|
作者
Gaxotte, V
Laurens, B
Haulon, S
Lions, C
Mounier-Véhier, C
Beregi, JP [1 ]
机构
[1] CHRU Lille, Hop Cardiol, Serv Radiol Vasc, Dept Vasc Radiol, F-59037 Lille, France
[2] CHRU Lille, Hop Cardiol, Dept Vasc Surg, F-59037 Lille, France
[3] CHRU Lille, Hop Cardiol, Dept Cardiol, F-59037 Lille, France
关键词
renal artery; common iliac artery; stenosis; in-stent stenosis; dissection; arterial rupture; aneurysm; Jostent stent-graft; restenosis;
D O I
10.1583/1545-1550(2003)010<0361:MTOTJB>2.0.CO;2
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To report the results of a multicenter feasibility study of the Jostent balloon-expandable stent-graft in the treatment of renal and iliac artery lesions. Methods: Twenty-three patients (17 men; mean age 62 years, range 38-80) with lesions in the renal (n=12) or iliac arteries (n=12) were enrolled in 6 centers over a 1-year period. Preprocedural computed tomography (CT) and angiography were performed in all patients. The Jostent device was implanted in the 24 arteries to treat 11 in-stent stenoses, 2 arterial ruptures, 2 aneurysms, 2 dissections, 2 ulcerated stenoses, and 5 chronic occlusions. Follow-up included color duplex ultrasound examination on the day after the procedure and at 6 months; patients with renal artery stent-grafts were also evaluated with CT angiography. Results: Twenty-seven stent-grafts were deployed successfully in the 24 (100%) arteries. Seven (30%) patients required adjunctive procedures to address 1 acute in-stent thrombosis, 2 dissections, and 4 in-stent residual stenoses. At 6-month follow-up, 2 (8.3%) restenoses occurred in the renal arteries; these were treated successfully using balloon angioplasty. Conclusions: These data suggest that a balloon-expandable stent-graft may be safe and useful in patients with selected peripheral indications.
引用
收藏
页码:361 / 365
页数:5
相关论文
共 50 条
  • [31] Iatrogenic Iliac Artery Rupture: Emergency Management by Longer Stent-Graft on a Shorter Balloon
    Vijay Trehan
    Arima Nigam
    S. Ramakrishnan
    CardioVascular and Interventional Radiology, 2007, 30 : 108 - 110
  • [32] Balloon-expandable cobalt chromium stent versus self-expandable nitinol stent for the Atherosclerotic Iliac Arterial Disease (SENS-ILIAC Trial) Trial: a randomized controlled trial
    Choi, Woong Gil
    Rha, Seung-Woon
    Choi, Byoung Geol
    Park, Soohyung
    Kim, Ji Bak
    Kang, Dong Oh
    Choi, Cheol Ung
    Seo, Yong Sung
    Cho, Yoon Hyung
    Park, Sang Ho
    Lee, Seung Jin
    Ko, Young Guk
    Her, Ae-Young
    Kim, Sang Min
    Kim, Ki Chang
    Cho, Jang Hyun
    Kang, Won Yu
    Kim, Joo Han
    Kim, Min Woong
    Kim, Do Hoi
    Bae, Jang Ho
    Ahn, Ji Hoon
    Jo, Sang Cheol
    Seo, Jae Bin
    Jung, Woo Young
    Park, Sang Min
    HEART AND VESSELS, 2024, 39 (12) : 1060 - 1067
  • [33] Iatrogenic iliac artery rupture: Emergency management by longer stent-graft on a shorter balloon
    Trehan, Vijay
    Nigam, Arima
    Ramakrishnan, S.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (01) : 108 - 110
  • [34] Excellent Midterm Results for the Viabahn Balloon-Expandable Endoprosthesis in Iliac Lesions
    Piazza, Michele
    JOURNAL OF ENDOVASCULAR THERAPY, 2020, 27 (05) : 737 - 738
  • [35] Emergency management of iatrogenic arterial injuries with a low-profile balloon-expandable stent-graft Preliminary results
    Ruffino, Maria Antonella
    Fronda, Marco
    Varello, Sara
    Discalzi, Andrea
    Mancini, Andrea
    Muratore, Pierluigi
    Rossato, Denis
    Bergamasco, Laura
    Righi, Dorico
    Fonio, Paolo
    MEDICINE, 2020, 99 (15) : E19655
  • [36] PERSISTENT ILIAC ENDARTERITIS WITH PSEUDOANEURYSM FORMATION FOLLOWING BALLOON-EXPANDABLE STENT PLACEMENT
    LIU, P
    DRAVID, V
    FREIMAN, D
    ZEGEL, H
    WEINBERG, D
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1995, 18 (01) : 39 - 42
  • [37] Experimental trial of balloon-expandable, metallic Palmaz stent in the trachea
    Fraga, JC
    Filler, RM
    Forte, V
    Bahoric, A
    Smith, C
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1997, 123 (05) : 522 - 528
  • [38] Balloon-expandable stents treat renal artery stenosis
    Al-Awqati, Q
    KIDNEY INTERNATIONAL, 2006, 69 (01) : 21 - 21
  • [39] Type Ic Endoleak after LifeStream Balloon-Expandable Stent Graft and Zenith Iliac Branch Device Placement
    Kim, Hyeon Ju
    Yun, Woo-Sung
    Kim, Hyung-Kee
    VASCULAR SPECIALIST INTERNATIONAL, 2023, 39
  • [40] Fixing Bridge Technique Combining Distal Covered Self-Expandable With Proximal Balloon-Expandable Stent-Graft in Case of Branched Endografting
    Di Stefano, Lucia
    Palermo, Dario
    Zezza, Alessandra
    Angiletta, Domenico
    Pulli, Raffaele
    JOURNAL OF VASCULAR SURGERY, 2022, 75 (06) : E162 - E163