Elimination of metal interference in the determination of fluoride ion by non-suppressor type ion chromatography

被引:2
|
作者
Tokunaga, S
Uchiumi, A
Haron, MJ
Tatsumoto, H
机构
[1] UNIV PERTANIAN MALAYSIA,DEPT CHEM,SERDANG 43400,MALAYSIA
[2] CHIBA UNIV,FAC ENGN,CHIBA 263,JAPAN
关键词
ion chromatography; fluoride analysis; metal interference; alkali pretreatment;
D O I
10.1080/03067319608028133
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple method has been developed to eliminate metal interferences in the ion-chromatographic determination of fluoride ion in aqueous solutions. Negative interferences become appreciable on 0.2-mM F- at 0.01 to 0.02 mM of Al(III), Ce(IV), La(III), Y(III), ce(III) and Pb(II); and at 0.25 mM of Ca(II). The interference from Cd(II), Co(II), Fe(II) and Ni(II) is insignificant at less than or equal to 5.0-mM metal for 5.0-mM F-. By alkalifying F- solutions at pH 12.3, the metal interference can be eliminated up to the concentrations of 0.25-mM AI(III), La(III) and Y(III); 1.0-mM Ce(IV); 2.5-mM Ce(In); and 5.0-mM Pb(II), Cd(II), Co(II), Fe(II) and NI(II). The Ca(II) interference cannot be eliminated. The pretreatment enables determination of 0.2- to 5.0-mM F- in presence of 0.25-mM AI(III) or La(III) with coefficient-of-variations of 1.99 to 6.20%.
引用
收藏
页码:191 / 205
页数:15
相关论文
共 50 条
  • [21] DETERMINATION OF FLUORIDE IN SEMICONDUCTOR PROCESS CHEMICALS BY ION CHROMATOGRAPHY WITH ION-SELECTIVE ELECTRODES
    TALASEK, RT
    JOURNAL OF CHROMATOGRAPHY, 1989, 465 (01): : 1 - 10
  • [22] A new method for determination of fluoride ion in commodity tea by ion-exclusion chromatography
    Zhou, Lili
    Li, Huaiping
    Ye, Shanrong
    Tan, Heping
    CYTA-JOURNAL OF FOOD, 2018, 16 (01) : 637 - 641
  • [23] Determination of fluoride in soil water extract through ion chromatography
    Saha, JK
    Kundu, S
    COMMUNICATIONS IN SOIL SCIENCE AND PLANT ANALYSIS, 2003, 34 (1-2) : 181 - 188
  • [24] Simultaneous determination of fluoride and monofluorophosphate in toothpastes by supressed ion chromatography
    Michalski, Rajmund
    Mathews, Barbara
    CENTRAL EUROPEAN JOURNAL OF CHEMISTRY, 2006, 4 (04): : 798 - 807
  • [25] ION CHROMATOGRAPHY WITH AN ION-EXCHANGE MEMBRANE SUPPRESSOR
    HANAOKA, Y
    MURAYAMA, T
    MURAMOTO, S
    MATSUURA, T
    NANBA, A
    JOURNAL OF CHROMATOGRAPHY, 1982, 239 (APR): : 537 - 548
  • [26] Determination of carbonate ion by ion chromatography
    Hashimoto, Y
    Miyanaga, A
    Ishikawa, O
    BUNSEKI KAGAKU, 1996, 45 (05) : 431 - 434
  • [27] AN ACCURATE FLUORIDE CALIBRATION IN THE PRESENCE OF STRONG INTERFERENCE FROM CARBONATE IN ION CHROMATOGRAPHY
    CHAKRABORTY, DK
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1994, 17 (08): : 607 - 608
  • [28] Accurate fluoride calibration in the presence of strong interference from carbonate in ion chromatography
    Chakraborty, D.K.
    HRC Journal of High Resolution Chromatography, 1994, 17 (08):
  • [29] AUTOMATED FLUORIDE-ION DETERMINATION - DETERMINATION OF URINE FLUORIDE-ION LEVELS
    COWELL, DC
    ANNALS OF CLINICAL BIOCHEMISTRY, 1977, 14 (SEP) : 269 - 274
  • [30] AUTOMATED FLUORIDE-ION DETERMINATION - DETERMINATION OF SERUM FLUORIDE-ION LEVELS
    COWELL, DC
    ANNALS OF CLINICAL BIOCHEMISTRY, 1977, 14 (SEP) : 275 - 278