Surface dose measurement for helical tomotherapy

被引:8
|
作者
Snir, Jonatan A. [1 ]
Mosalaei, Homeira [1 ]
Jordan, Kevin [1 ,2 ,3 ]
Yartsev, Slav [1 ,2 ,3 ]
机构
[1] London Hlth Sci Ctr, London Reg Canc Program, London, ON N6A 4L6, Canada
[2] Univ Western Ontario, Dept Oncol, London, ON N6A 3K7, Canada
[3] Univ Western Ontario, Dept Med Biophys, London, ON N6A 3K7, Canada
关键词
helical tomotherapy; OSL dosimeter; MOSFET dosimeter; skin dose; MOSFET; HEAD;
D O I
10.1118/1.3592021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the surface dose measurements made by different dosimeters for the helical tomotherapy (HT) plan in the case of the target close to the surface. Methods: Surface dose measurements in different points for the HT plan to deliver 2 Gy to the planning target volume (PTV) at 5 mm below the surface of the cylindrical phantom were performed by radiochromic films, single use metal oxide semiconductor field-effect transistor (MOSFET) dosimeters, silicon IVD QED diode, and optically stimulated luminescence (OSL) dosimeters. Results: The measured doses by all dosimeters were within 12 +/- 8% difference of each other. Conclusions: Radiochromic films, EBT, and EBT(2), provide high spatial resolution, although it is difficult to get accurate measurements of dose. Both the OSL and QED measured similar dose to that of the MOSFET detectors. The QED dosimeter is promising as a reusable on-line wireless dosimeter, while the OSL dosimeters are easier to use, require minimum setup time and are very precise. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3592021]
引用
收藏
页码:3104 / 3107
页数:4
相关论文
共 50 条
  • [31] Measurement and correction of leaf open times in helical tomotherapy
    Sevillano, David
    Minguez, Cristina
    Sanchez, Alicia
    Sanchez-Reyes, Alberto
    MEDICAL PHYSICS, 2012, 39 (11) : 6972 - 6980
  • [32] Helical tomotherapy
    Mackie, TR
    RADIOTHERAPY AND ONCOLOGY, 2006, 78 : S12 - S13
  • [33] Evaluation of Superficial Dose for Helical Tomotherapy with GAFCHROMIC® EBT Film
    Lee, H.
    Chong, N.
    Tien, H.
    Lin, C.
    Yen, H.
    Shueng, P.
    Wu, L.
    Hsieh, C.
    Chen, C.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [34] Dose reconstruction for cord compressions retreatments using helical tomotherapy
    Mahan, SL
    Ramsey, C
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S640 - S641
  • [35] Radiation dose for normal organs by helical tomotherapy for lung cancer
    Tseng, Hsien-Chun
    Liu, Wen-Shan
    Tsai, Hsiao-Han
    Chu, Hsin-Yi
    Lin, Jye-Bin
    Chen, Chien-Yi
    APPLIED RADIATION AND ISOTOPES, 2015, 102 : 35 - 41
  • [36] Dose calibration of nonconventional treatment systems applied to helical tomotherapy
    Jeraj, R
    Mackie, TR
    Balog, J
    Olivera, G
    MEDICAL PHYSICS, 2005, 32 (02) : 570 - 577
  • [37] The impact of linac output variations on dose distributions in helical tomotherapy
    Flynn, R. T.
    Kissick, M. W.
    Mehta, M. P.
    Olivera, G. H.
    Jeraj, R.
    Mackie, T. R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (02): : 417 - 430
  • [38] A Clarkson-Based Independent Dose Verification for the Helical Tomotherapy
    Nagata, H.
    Hongo, H.
    Kawai, D.
    Takahashi, R.
    Hashimoto, H.
    Tachibana, H.
    MEDICAL PHYSICS, 2016, 43 (06) : 3524 - 3524
  • [39] Dose painting with IMPT, helical tomotherapy and IMXT: A dosimetric comparison
    Thorwarth, Daniela
    Soukup, Martin
    Atber, Markus
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 30 - 34
  • [40] Dose Summation for Multi-Phase Helical Tomotherapy Plans
    Turian, J.
    Chen, S.
    Coleman, J.
    Khelashvili, G.
    Chu, J. C. H.
    MEDICAL PHYSICS, 2008, 35 (06) : 2838 - +