Absolute measurement of photon flux using ionization chamber in (6∼20) keV synchrotron radiation X-rays

被引:0
|
作者
Guo, Siming [1 ]
Wang, Peiwei [1 ]
Wang, Ji [1 ]
Li, Fan [2 ]
Wu, Jinjie [1 ]
Zhang, Jian [1 ]
机构
[1] Natl Inst Metrol, Div Ionizing Radiat Metrol, Beijing, Peoples R China
[2] Inst High Energy Phys, Beijing, Peoples R China
基金
国家重点研发计划;
关键词
DETECTOR CALIBRATION; RECOMBINATION; RADIOMETRY;
D O I
10.1002/xrs.3152
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Absolute measurement of photon flux under high-fluence rate X-rays was required in photon diagnostics for X-ray lasers. A free-air ionization chamber used for the reproducibility of air-kerma in high-fluence rate X-ray was designed, and the monoenergetic X-rays flux of (6-20) keV were absolutely measured on Beijing Synchrotron Radiation Facility. The energy spectrum of the monoenergetic X-rays was measured using a silicon drift detector, and the proportion of higher harmonics was determined. The saturation curve, air attenuation correction factor, and composite loss correction factor of the free-air ionization chamber were obtained experimentally. The air-kerma rate for the (6 similar to 20) keV monoenergetic X-ray was measured accurately, and the relative combined standard uncertainty of the air-kerma is 0.61%.
引用
收藏
页码:572 / 579
页数:8
相关论文
共 50 条
  • [1] Absolute measurement of radiant power for synchrotron radiation monochromatized X-rays
    Fan Li
    Yi-Dong Zhao
    Pei-Wei Wang
    Lei Zheng
    Si-Ming Guo
    Ji Wang
    Kun Tang
    Xiao-Juan Zhao
    Jian-Min Li
    Radiation Detection Technology and Methods, 2020, 4 : 97 - 105
  • [2] Absolute measurement of radiant power for synchrotron radiation monochromatized X-rays
    Li, Fan
    Zhao, Yi-Dong
    Wang, Pei-Wei
    Zheng, Lei
    Guo, Si-Ming
    Wang, Ji
    Tang, Kun
    Zhao, Xiao-Juan
    Li, Jian-Min
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2020, 4 (01) : 97 - 105
  • [3] Measurement of stress using synchrotron x-rays
    Weidner, Donald J.
    Li, Li
    JOURNAL OF PHYSICS-CONDENSED MATTER, 2006, 18 (25) : S1061 - S1067
  • [4] MEASUREMENTS OF ABSOLUTE SENSITIVITY OF SEMICONDUCTOR-DETECTORS FOR 7-20 KEV X-RAYS USING SYNCHROTRON RADIATION OF VEPP-3 STORAGE RING
    DOLBNYA, IP
    MEZENTSEV, NA
    PINDYURIN, VF
    ROMANENKO, KP
    SUBBOTIN, AN
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 685 - 688
  • [5] Improved free-air ionization chamber for the measurement of x-rays
    Chen, WL
    Su, SH
    Su, LL
    Hwang, WS
    METROLOGIA, 1999, 36 (01) : 19 - 24
  • [6] Measurement of supervoltage x-rays with the free-air ionization chamber
    Taylor, LS
    Singer, G
    Charlton, AL
    AMERICAN JOURNAL OF ROENTGENOLOGY AND RADIUM THERAPY, 1939, 41 : 256 - 275
  • [7] Measurement of supervoltage X-rays with the free-air ionization chamber
    Taylor, LS
    Singer, G
    Charlton, AL
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1938, 21 (01): : 19 - 44
  • [8] Construction of an ionization chamber for the measurement of dose of low energy x-rays
    Alcantara Perez, Y. B.
    Ramirez Jimenez, F. J.
    Montano Zetina, L. M.
    MEDICAL PHYSICS, 2008, 1032 : 131 - +
  • [9] Two-photon correlations in X-rays from a synchrotron radiation source
    Kunimune, Y
    Yoda, Y
    Izumi, K
    Yabashi, M
    Zhang, XW
    Harami, T
    Ando, M
    Kikuta, S
    JOURNAL OF SYNCHROTRON RADIATION, 1997, 4 : 199 - 203
  • [10] Examining catalysts using X-rays from synchrotron radiation
    Evans, John
    Chemistry and Industry (London), 1995, (04):