Cerebral oxygenation measured by near-infrared spectroscopy: Comparison with jugular bulb oximetry

被引:128
|
作者
Daubeney, PEF [1 ]
Pilkington, SN [1 ]
Janke, E [1 ]
Charlton, GA [1 ]
Smith, DC [1 ]
Webber, SA [1 ]
机构
[1] SOUTHAMPTON GEN HOSP,WESSEX CARDIOTHORAC CTR,SOUTHAMPTON SO9 4XY,HANTS,ENGLAND
来源
ANNALS OF THORACIC SURGERY | 1996年 / 61卷 / 03期
关键词
D O I
10.1016/0003-4975(95)01186-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Near-infrared spectroscopy is a potential tool for measuring adequacy of cerebral oxygenation during cardiac operations. The cerebral microcirculation is predominantly venous (by volume) and therefore regional cerebral oxygenation measured by near-infrared spectroscopy should reflect jugular bulb venous saturations. Methods. We compared simultaneous regional cerebral oxygenation and jugular bulb venous saturation measurements in 40 children (median age, 4.5 years; range, 2 weeks to 14.5 years) in the cardiac catheter laboratory (n = 29) and during cardiac operations (n = 11). Results. For all patients combined the correlation between regional cerebral oxygenation and jugular bulb venous saturation was 0.69 (p < 0.0001) and was similar for the two groups. For individual children undergoing cardiac operations excellent correlations were obtained (r = 0.78 to 0.96; median, 0.91). However, at low values of jugular bulb venous saturation, regional cerebral oxygenation tended to run high, whereas the converse was true for high values of jugular bulb venous saturation. Conclusions. These findings suggest that near-infrared spectroscopy may be a useful tool for assessing intravascular cerebral oxygenation during pediatric cardiac operations. Prospective studies of neurologic outcome will be required to establish the value of this technique for assessing the adequacy of cerebral protection.
引用
收藏
页码:930 / 934
页数:5
相关论文
共 50 条
  • [1] Cerebral oxygenation measured by near-infrared spectroscopy: Comparison with jugular bulb oximetry - Invited commentary
    Gunter, JB
    ANNALS OF THORACIC SURGERY, 1996, 61 (03): : 934 - 934
  • [2] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Nagdyman, N
    Fleck, T
    Schubert, S
    Ewert, P
    Peters, B
    Lange, PE
    Abdul-Khaliq, H
    INTENSIVE CARE MEDICINE, 2005, 31 (06) : 846 - 850
  • [3] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Fleck, TPK
    Nagdyman, N
    Schubert, S
    Peters, B
    Reisenkampff, E
    Ewert, P
    Abdul-Khaliq, H
    Lange, PE
    PEDIATRIC RESEARCH, 2004, 56 (03) : 478 - 478
  • [4] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Nicole Nagdyman
    Thilo Fleck
    Stephan Schubert
    Peter Ewert
    Björn Peters
    Peter E. Lange
    Hashim Abdul-Khaliq
    Intensive Care Medicine, 2005, 31 : 846 - 850
  • [5] Cerebral oxygenation measured by near-infrared spectroscopy
    Wollert, HG
    Eckel, L
    ANNALS OF THORACIC SURGERY, 1997, 63 (01): : 292 - 292
  • [6] Cerebral oxygenation monitoring by near-infrared spectroscopy is not clinically useful in patients with severe closed-head injury: A comparison with jugular venous bulb oximetry
    Lewis, SB
    Myburgh, JA
    Thornton, EL
    Reilly, PL
    CRITICAL CARE MEDICINE, 1996, 24 (08) : 1334 - 1338
  • [7] MONITORING OF THE CEREBRAL OXYGENATION WITH NEAR-INFRARED SPECTROSCOPY IN COMPARISON TO JUGULAR BULB OXYGEN-SATURATION DURING CAROTID ENDARTERECTOMY
    SCHINDLER, E
    ZICKMANN, B
    MULLER, M
    BOLDT, J
    KROLL, J
    HEMPELMANN, G
    VASA-JOURNAL OF VASCULAR DISEASES, 1995, 24 (02): : 168 - 175
  • [8] A comparison of near-infrared spectroscopy and jugular bulb oximetry in comatose patients resuscitated from a cardiac arrest
    Buunk, G
    van der Hoeven, JG
    Meinders, AE
    ANAESTHESIA, 1998, 53 (01) : 13 - 19
  • [9] Comparison of regional transcranial oximetry with near infrared spectroscopy (NIRS) and jugular venous bulb oxygen saturation for the monitoring of cerebral oxygenation in infants and children
    Abdul-Khaliq, H
    Troitzsch, D
    Berger, F
    Lange, PE
    BIOMEDIZINISCHE TECHNIK, 2000, 45 (11): : 328 - 332
  • [10] 86 Comparison Between Cerebral Tissue Oxygenation Index Measured by Near- Infrared Spectroscopy and Venous Jugular Bulb Saturation in Children
    T P K Fleck
    N Nagdyman
    S Schubert
    B Peters
    E Riesenkampff
    P Ewert
    H Abdul-Khaliq
    P E Lange Deutsches
    Pediatric Research, 2004, 56 (3) : 478 - 478