Effect of Implant Materials on Bone Remodelling Around Cemented Acetabular Cup

被引:4
|
作者
Kumar, Ajay [1 ]
Ghosh, Rajesh [1 ]
Kumar, Rajeev [1 ]
机构
[1] Indian Inst Technol Mandi, Sch Engn, Kamand 175005, Himachal Prades, India
关键词
Acetabular cup; Implant material; Bone remodelling; TOTAL HIP-REPLACEMENT; STRESS DISTRIBUTIONS; THICKNESS; FATIGUE; MANTLE; FAILURE; COMPONENT; REGION;
D O I
10.1007/978-981-16-0550-5_3
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Bone resorption around cemented acetabular cup caused failure in the long tenure. This study aims to investigate the effect of different implant materials on bone remodelling around the cemented acetabular cup. Finite element (FE) models of intact and implanted pelvic models were established using CT-scan data sets. Three combinations (acetabular cup-femoral head) of implant materials, UHMWPE-CoCrMo, CoCrMo-CoCrMo, and alumina-alumina ceramic, were considered to predict the effect of implant materials on bone remodelling around cemented acetabular cup. Less bone resorption was observed for UHMWPE acetabular cup. Bone remodelling was nearly similar for the ceramic and metallic cup. Considering the bone remodelling scenario, the results of the current study stated that ceramic could be a good alternative to metallic implant in the cemented acetabular cup since ceramic has better wear resistance property than metal.
引用
收藏
页码:17 / 26
页数:10
相关论文
共 50 条
  • [1] Bone remodelling around a cemented polyethylene cup - A longitudinal densitometry study
    Shetty, NR
    Hamer, AJ
    Kerry, RM
    Stockley, I
    Eastell, R
    Wilkinson, JM
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (04): : 455 - 459
  • [2] Bone remodelling around cementless composite acetabular components: The effects of implant geometry and implant-bone interfacial conditions
    Ghosh, Rajesh
    Gupta, Sanjay
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 32 : 257 - 269
  • [3] Biomechanical performance of the cemented acetabular cup with combined effects of bone quality, implant material combinations and bodyweight
    Kumar, Ajay
    Mondal, Subrata
    Ghosh, Rajesh
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (09) : 1309 - 1327
  • [4] Impaction bone grafting and a cemented cup after acetabular fracture
    Bronsema, Eelco
    te Stroet, Martijn A. J.
    Zengerink, Maartje
    van Kampen, Albert
    Schreurs, B. Willem
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (12) : 2441 - 2446
  • [5] Impaction bone grafting and a cemented cup after acetabular fracture
    Eelco Bronsema
    Martijn A. J. te Stroet
    Maartje Zengerink
    Albert van Kampen
    B. Willem Schreurs
    International Orthopaedics, 2014, 38 : 2441 - 2446
  • [6] Acetabular revision with metal mesh, impaction bone grafting, and a cemented cup
    Buttaro, Martin A.
    Comba, Fernando
    Pusso, Rodolfo
    Piccaluga, Francisco
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (10) : 2482 - 2490
  • [7] Bone remodelling around uncemented metallic and ceramic acetabular components
    Ghosh, Rajesh
    Mukherjee, Kaushik
    Gupta, Sanjay
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (H5) : 490 - 502
  • [8] Early periprosthetic bone remodelling around cemented and uncemented custom-made femoral components and their uncemented acetabular cups
    Mulier, M.
    Jaecques, S. V. N.
    Raaijmaakers, M.
    Nijs, J.
    Van der Perre, G.
    Jonkers, I.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (07) : 941 - 948
  • [9] Early periprosthetic bone remodelling around cemented and uncemented custom-made femoral components and their uncemented acetabular cups
    M. Mulier
    S. V. N. Jaecques
    M. Raaijmaakers
    J. Nijs
    G. Van der Perre
    I. Jonkers
    Archives of Orthopaedic and Trauma Surgery, 2011, 131 : 941 - 948
  • [10] Mechanobiochemical bone remodelling around an uncemented acetabular component: influence of bone orthotropy
    Ceby Mullakkara Saviour
    Basil Mathai
    Sanjay Gupta
    Medical & Biological Engineering & Computing, 2024, 62 : 1717 - 1732