Direct determination of selenium in serum by electrothermal atomic absorption spectrometry using automated ultrasonic slurry sampling

被引:21
|
作者
Chen, WK
Yen, CC [1 ]
Wei, BL
Hu, CC
Yu, JJ
Chung, C
Kuo, SC
机构
[1] Chung Shang Med & Dent Coll, Inst Toxicol, Taichung, Taiwan
[2] Ta Jen Pharmaceut Coll, Dept Environm Engn & Hlth, Pingtung, Taiwan
[3] Feng Chia Univ, Dept Environm Engn & Sci, Taichung 40724, Taiwan
[4] Natl Tsing Hua Univ, Inst Nucl Sci, Hsinchu, Taiwan
[5] China Med Coll, Inst Pharmaceut Chem, Taichung, Taiwan
关键词
selenium; serum; ultrasonic slurry sampling; ETAAS;
D O I
10.1016/S0584-8547(97)00103-1
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Selenium concentration in body fluids is a good index to establish human selenium status. This work discusses the determination of selenium in serum by ETAAS using longitudinal Zeeman-effect background correction and combining the use of automated slurry sampling. The standard reference materials bovine serum (NIST, SRM 1598) and second-generation biological freeze-dried human serum are analyzed to verify the accuracy and precision of this technique. The direct method proposed in this study is used for the determination of selenium in human serum collected from healthy people of 19-25 years. The average accuracy values of certified reference serum samples and the recovery values of spiked samples indicate this method to be an efficient and rapid technique for determining selenium in biological samples. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 50 条
  • [31] DIRECT DETERMINATION OF IMPURITIES IN POWDERED SILICON-CARBIDE BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING THE SLURRY SAMPLING TECHNIQUE
    DOCEKAL, B
    KRIVAN, V
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (03) : 521 - 528
  • [32] Direct determination of lead in bones using slurry sampling and high-resolution continuum source electrothermal atomic absorption spectrometry
    Husakova, Lenka
    Sidova, Tereza
    Ibrahimova, Lucie
    Svizelova, Monika
    Mikysek, Tomas
    ANALYTICAL METHODS, 2019, 11 (09) : 1254 - 1263
  • [33] Green direct ultrasonic slurry sampling and electrothermal atomic absorption spectrometry method for metal analysis of solid plastics samples
    Chovancova, Katarina
    Michalides, Nicolas Milan
    Koslabova, Martina
    Koncalova, Martina
    Halko, Radoslav
    ANALYTICAL METHODS, 2024, 16 (07) : 990 - 1002
  • [34] Determination of trace impurities in titanium dioxide by slurry sampling electrothermal atomic absorption spectrometry
    Dong, HM
    Krivan, V
    Welz, B
    Schlemmer, G
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (12) : 1747 - 1762
  • [35] Speciation analysis of chromium in water samples using nanometer zirconium dioxide and direct ultrasonic slurry sampling with electrothermal atomic absorption spectrometry
    Prochazkova, Simona
    Kriegerova, Katarina
    Bohacova, Iveta
    Halko, Radoslav
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2019, 99 (02) : 157 - 171
  • [36] Direct determination of selenium in a wild fruit juice by electrothermal atomic absorption spectrometry
    Liu, YM
    Gong, BL
    Li, ZH
    Xu, YL
    Lin, TZ
    TALANTA, 1996, 43 (07) : 985 - 989
  • [37] Determination of aluminium and manganese in human scalp hair by electrothermal atomic absorption spectrometry using slurry sampling
    Bermejo-Barrera, P
    Moreda-Pineiro, A
    Moreda-Pineiro, J
    Bermejo-Barrera, A
    TALANTA, 1998, 45 (06) : 1147 - 1154
  • [38] DETERMINATION OF ARSENIC IN MUSSELS BY SLURRY SAMPLING AND ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY (ETAAS)
    BERMEJOBARRERA, P
    LORENZOALONSO, MJ
    ABOALSOMOZA, M
    BERMEJOBARRERA, A
    MIKROCHIMICA ACTA, 1994, 117 (1-2) : 49 - 64
  • [39] Determination of mercury in fluorescent lamp cullet by slurry sampling electrothermal atomic absorption spectrometry
    Dobrowolski, R
    Mierzwa, J
    ANALYST, 1996, 121 (07) : 897 - 900
  • [40] Determination of impurities in boron nitride powder by slurry sampling electrothermal atomic absorption spectrometry
    Krivan, V
    Rommelt, T
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (02) : 137 - 141