Evaluation of an in-clinic dry chemistry analyzer for canine, equine, and feline plasma samples

被引:11
|
作者
Boes, Katie M. [1 ]
Sink, Carolyn A. [2 ]
Camus, Melinda S. [4 ]
Werre, Stephen R. [3 ]
机构
[1] Virginia Polytech Inst & State Univ, Virginia Maryland Coll Vet Med, Dept Biomed Sci & Pathobiol, Blacksburg, VA 24061 USA
[2] Virginia Polytech Inst & State Univ, Virginia Maryland Coll Vet Med, Vet Teaching Hosp, Blacksburg, VA 24061 USA
[3] Virginia Polytech Inst & State Univ, Virginia Maryland Coll Vet Med, Dept Populat Hlth Sci, Blacksburg, VA 24061 USA
[4] Univ Georgia, Coll Vet Med, Dept Pathol, Athens, GA 30602 USA
关键词
Catalyst One; clinical chemistry tests; point-of-care systems; quality control; GUIDELINES;
D O I
10.1177/1040638718809407
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Method validation studies characterize the performance of new laboratory methods relative to established methods using quality guidelines in order to define the new method's performance characteristics and to identify differences that could influence data interpretation. We investigated the performance of an in-clinic dry chemistry analyzer (Catalyst One, IDEXX) for measuring 19 routine plasma biochemistry analytes in dogs, cats, and horses. We analyzed 2 levels of quality control material (QCM) in duplicate twice daily for 5 d to determine the coefficient of variation (CV), percent bias, observed total error (TEobs), and sigma metric (sigma) for each analyte at each level of QCM. We analyzed 82 canine, equine, and feline plasma samples with the in-clinic dry chemistry analyzer and a reference wet chemistry analyzer, and results were compared using correlation coefficients, Deming regression, and Bland-Altman analyses. CVs were <5% for 16 analytes and > 5% for 3 analytes. TEobs was less than allowable total error (TEa) for 9 analytes, and exceeded TEa for 10 analytes. Sigma metrics were >4 at both levels of QCM for 5 analytes, and at one level of QCM for 5 analytes; sigma metrics were <3 or could not be calculated at the remaining analyte concentrations. All analytes, except glucose, showed various magnitudes of bias compared to the wet chemistry analyzer. Based on these results, we recommend statistical (5 analytes) and non-statistical (14 analytes) QC measures and analyzer-specific reference intervals.
引用
收藏
页码:902 / 910
页数:9
相关论文
共 50 条
  • [21] In-clinic diagnosis of canine anaplasmosis based on haematological abnormalities and evaluation of a stained blood smear
    Kirtz, G.
    Leidinger, E.
    TIERAERZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2015, 43 (02): : 101 - 106
  • [22] EVALUATION OF A NEW DRY-CHEMISTRY ANALYZER, THE CLINISTAT, FOR BLOOD HEMOGLOBIN AND SERUM OR PLASMA CONSTITUENTS
    GLICK, M
    GLICK, S
    RYDER, KW
    CLINICAL CHEMISTRY, 1989, 35 (06) : 1108 - 1108
  • [23] Serologic diagnosis of equine borreliosis and anaplasmosis:: Evaluation of an in-clinic ELISA SNAP®4Dx®)
    Chandrashekar, R.
    Daniluk, D.
    Esty, C.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2008, 22 (03) : 820 - 820
  • [24] Evaluation of an in-clinic Serum Amyloid A (SAA) assay and assessment of the effects of storage on SAA samples
    Anna Hillström
    Harold Tvedten
    Inger Lilliehöök
    Acta Veterinaria Scandinavica, 52
  • [25] Evaluation of an in-clinic Serum Amyloid A (SAA) assay and assessment of the effects of storage on SAA samples
    Hillstrom, Anna
    Tvedten, Harold
    Lilliehook, Inger
    ACTA VETERINARIA SCANDINAVICA, 2010, 52
  • [26] SERUM LEVELS OF RADIOIMMUNOASSAY AND OTHER ROUTINE CHEMISTRY PROCEDURES FOR CANINE, FELINE, BOVINE, AND EQUINE SPECIES
    BAJEMA, L
    CLINICAL CHEMISTRY, 1980, 26 (07) : 1045 - 1045
  • [27] Comparative evaluation of field samples using 2 in-clinic assays for heartworm antigen detection in dogs
    Burton, K. W.
    Liu, J.
    Drexel, J. P.
    Pulaski, C.
    Heartsill, S.
    Dupuy, C.
    VETERINARY PARASITOLOGY, 2020, 283
  • [28] EFFECTS OF BILIRUBINEMIA, HEMOLYSIS, AND LIPEMIA ON CLINICAL-CHEMISTRY ANALYTES IN BOVINE, CANINE, EQUINE, AND FELINE SERA
    JACOBS, RM
    LUMSDEN, JH
    GRIFT, E
    CANADIAN VETERINARY JOURNAL-REVUE VETERINAIRE CANADIENNE, 1992, 33 (09): : 605 - 608
  • [29] Comparative clinical study of canine and feline total blood cell count results with seven in-clinic and two commercial laboratory hematology analyzers
    Becker, Martina
    Moritz, Andreas
    Giger, Urs
    VETERINARY CLINICAL PATHOLOGY, 2008, 37 (04) : 373 - 384
  • [30] EVALUATION OF CANINE AND FELINE HEMOGRAMS USING THE IDEXX PROCYTE DX ANALYZER COMPARED TO MANUAL DIFFERENTIAL
    Owen, J. L.
    Wamsley, H. L.
    Conrad, B. L.
    Harvey, J. W.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2010, 24 (03) : 755 - 755