Combined detection of C-reactive protein and PBMC quantification from whole blood in an integrated lab-on-a-disc microfluidic platform

被引:18
|
作者
Uddin, Rokon [1 ]
Donolato, Marco [2 ]
Hwu, En-Te [1 ]
Hansen, Mikkel Fougt [1 ]
Boisen, Anja [1 ]
机构
[1] Tech Univ Denmark, Dept Micro & Nanotechnol, DTU Nanotech, Bldg 345B, DK-2800 Lyngby, Denmark
[2] BluSense Diagnost, Fruebjergvej 3, DK-2100 Copenhagen O, Denmark
基金
新加坡国家研究基金会; 欧洲研究理事会;
关键词
CRP; PBMC; Centrifugal microfluidics; Magnetic beads; Optomagnetic readout; Optical imaging; CELL COUNT; AGGLUTINATION ASSAY; CRP DETECTION; RISK; SENSITIVITY; IMMUNOASSAY; INFECTIONS; MORTALITY;
D O I
10.1016/j.snb.2018.07.015
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There is an increasing need for portable and low-cost diagnostic devices for detecting inflammatory/infectious diseases in a rapid and user-friendly fashion. Here, we present a lab-on-a-disc solution, which performs automated sample pre-treatment and combinedly detects small molecules and counts cells in a whole blood sample with a volume of 8.75 mu L with a sample to answer time of 14 min. It is used to detect two common inflammation/ infection biomarkers, C-reactive protein (CRP) and peripheral blood mononuclear cell (PBMC) count. The whole blood sample was separated into plasma and PBMC fractions using density gradient centrifugation and centrifugo- pneumatic valving. On-disc CRP detection was performed in the extracted plasma using a CRP-antibodyfunctionalized magnetic nanobead (MNB)-based agglutination assay and a Blu-ray-based optomagnetic detection unit. On-disc PBMC scanning and quantification was performed using an optical imaging unit. Both detection units were integrated on the centrifugal platform and the entire study was automated in order to ensure reliability of the assay and user-friendliness of the method. We measured the CRP level of subjects with different CRP levels and obtained approximately 73% PBMC extraction efficiency compared to hospital results. The concurrent/ combined detection of these two common biomarkers in an automated microfluidic platform with integrated detection units and with a low sample-to-answer time is a significant step forward towards a low-cost, out-of-lab, and portable tool to detect multiple biomarkers of significantly different nature (molecules and cells).
引用
收藏
页码:634 / 642
页数:9
相关论文
共 50 条
  • [21] Microfluidic integrated capacitive biosensor for C-reactive protein label-free and real-time detection
    Liu, Danyang
    Zhou, Lin
    Huang, Lihong
    Zuo, Zhaorui
    Ho, Vincent
    Jin, Laidi
    Lu, Yun
    Chen, Xianfeng
    Zhao, Jianlong
    Qian, Dahong
    Liu, Huiying
    Mao, Hongju
    ANALYST, 2021, 146 (17) : 5380 - 5388
  • [22] Lab-on-CD microfluidic platform for rapid separation and mixing of plasma from whole blood
    Ju-Nan Kuo
    Bo-Shiun Li
    Biomedical Microdevices, 2014, 16 : 549 - 558
  • [23] Lab-on-CD microfluidic platform for rapid separation and mixing of plasma from whole blood
    Kuo, Ju-Nan
    Li, Bo-Shiun
    BIOMEDICAL MICRODEVICES, 2014, 16 (04) : 549 - 558
  • [24] A diagnostic platform for rapid, simultaneous quantification of procalcitonin and C-reactive protein in human serum
    Cao, Xiangkun Elvis
    Ongagna-Yhombi, Serge Y.
    Wang, Ruisheng
    Ren, Yue
    Srinivasan, Balaji
    Hayden, Joshua A.
    Zhao, Zhen
    Erickson, David
    Mehta, Saurabh
    EBIOMEDICINE, 2022, 76
  • [25] Label-free impedance detection of cancer cells from whole blood on an integrated centrifugal microfluidic platform
    Nwankire, Charles E.
    Venkatanarayanan, Anita
    Glennon, Thomas
    Keyes, Tia E.
    Forster, Robert J.
    Ducree, Jens
    BIOSENSORS & BIOELECTRONICS, 2015, 68 : 382 - 389
  • [26] A low-cost electrochemical biosensor platform for C-reactive protein detection
    Guillem, Pedro
    Bustos, Rosa-Helena
    Garzon, Vivian
    Munoz, Andres
    Juez, Graciela
    SENSING AND BIO-SENSING RESEARCH, 2021, 31
  • [27] A Magnetic-bead Based Microfluidic System for Automatic C-reactive protein Detection
    Lee, Wen-Bin
    Lin, Hsin-I
    Shiesh, Shu-Chu
    Lee, Gwo-Bin
    MNHMT2009, VOL 1, 2010, : 257 - 260
  • [28] Integrated Microfluidic Device for Accurate Extracellular Vesicle Quantification and Protein Markers Analysis Directly from Human Whole Blood
    Zhou, Sisi
    Hu, Tao
    Zhang, Fen
    Tang, Dezhi
    Li, Dake
    Cao, Jian
    Wei, Wei
    Wu, Yafeng
    Liu, Songqin
    ANALYTICAL CHEMISTRY, 2020, 92 (01) : 1574 - 1581
  • [29] Dual-Aptamer Assay for C-reactive Protein Detection by Using Field-effect Transistors on an Integrated Microfluidic System
    Kao, Wei-Chieh
    Chu, Chia-Ho
    Chang, Wen-Hsin
    Wang, Yu-Lin
    Lee, Gwo-Bin
    2016 IEEE 11TH ANNUAL INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS (NEMS), 2016,
  • [30] DETECTION OF COMBINED PROCALCITONIN AND C-REACTIVE PROTEIN APPLIED IN THE DIAGNOSIS OF BACTERIAL INFECTIONS
    Li, Y. L.
    Zhai, L. C.
    Ji, J. H.
    Liu, L. Y.
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2017, 31 (01): : 177 - 181