摘要
Purpose: Magnetic particle imaging (MPI) applies oscillating magnetic fields to determine the distribution of magnetic nanoparticles in vivo. Using a receive coil, the change of the particle magnetization can be detected. However, the signal induced by the nanoparticles is superimposed by the direct feedthrough interference of the sinusoidal excitation field, which couples into the receive coils. As the latter is several magnitudes higher, the extraction of the particle signal from the excitation signal is a challenging task. Methods: One way to remove the interfering signal is to suppress the excitation signal by means of a band-stop filter. However, this technique removes parts of the desired particle signal, which are essential for direct reconstruction of the particle concentration. A way to recover the entire particle signal is to cancel out the excitation signal by coupling a matching cancellation signal into the receive chain. However, the suppression rates that can be achieved by signal cancellation are not as high as with the filtering method, which limits the sensitivity of this method. In order to unite the advantages of both methods, in this work the authors propose to combine the filtering and the cancellation technique. All methods were compared by measuring 10 mu l Resovist, in the same field generator only switching the signal processing parts. Results: The reconstructed time signals of the three methods, show the advantage of the proposed combination of filtering and cancellation. The method preserves the fundamental frequency and is able to detect the tracer signal at its full bandwidth even for low concentrations. Conclusions: By recovering the full particle signal the SNR can be improved and errors in the x-space reconstruction are prevented. The authors show that the combined method provides this full particle signal and makes it possible to improve image quality. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4794482]
引用
收藏
页数:9
相关论文
共 50 条
  • [31] EQUIVALENCE OF DIGITAL AND ANALOG SIGNAL PROCESSING
    STEIGLIT.K
    INFORMATION AND CONTROL, 1965, 8 (05): : 455 - &
  • [32] Memristive Systems For Analog Signal Processing
    Biolek, Dalibor
    Biolkova, Viera
    Kolka, Zdenek
    2014 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2014, : 2588 - 2591
  • [33] MICROWAVE PHOTONICS FOR ANALOG SIGNAL PROCESSING
    Yi, Xiaoke
    Minasian, Robert
    2015 14TH INTERNATIONAL CONFERENCE ON OPTICAL COMMUNICATIONS AND NETWORKS (ICOCN), 2015,
  • [34] Equivalent circuits of receive antennas in signal processing arrays
    Kildal, PS
    MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 1999, 21 (04) : 244 - 246
  • [35] Analog and Digital Signal Processing for Pressure Source Imaging at 190 MeV Proton Beam
    Venicelli, Elia A.
    Riva, Michele
    Zannoni, Mario
    Baschirotto, Andrea
    De Matteis, Marcello
    8TH INTERNATIONAL CONFERENCE ON ACOUSTIC AND RADIO EEV NEUTRINO DETECTION ACTIVITIES (ARENA 2018), 2019, 216
  • [36] Experimental and simulation studies on the behavior of signal harmonics in magnetic particle imaging
    Murase K.
    Konishi T.
    Takeuchi Y.
    Takata H.
    Saito S.
    Radiological Physics and Technology, 2013, 6 (2) : 399 - 414
  • [37] INVESTIGATION OF THE MAGNETIC PARTICLE IMAGING SIGNAL'S DEPENDENCY ON FERROFLUID CONCENTRATION
    Gehrcke, J. -P.
    Rueckert, M. A.
    Kampf, T.
    Kullmann, W. H.
    Jakob, P. M.
    Behr, V. C.
    MAGNETIC NANOPARTICLES: PARTICLE SCIENCE, IMAGING TECHNOLOGY, AND CLINICAL APPLICATIONS, 2010, : 73 - 78
  • [38] DICTIONARY-BASED BACKGROUND SIGNAL ESTIMATION FOR MAGNETIC PARTICLE IMAGING
    Knopp, Tobias
    Grosser, Mirco
    Graeser, Matthias
    Gerkman, Timo
    Moeddel, Martin
    2021 IEEE 18TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2021, : 1540 - 1543
  • [39] A CMOS Analog Cell and its applications in analog signal processing
    Abuelma'Atti, MT
    Alabsi, MA
    INTERNATIONAL JOURNAL OF ELECTRONICS, 2006, 93 (04) : 251 - 267
  • [40] Research on reconfigurable analog array for multifunctional analog signal processing
    Wang, Youren
    Zhu, Mingtao
    Cui, Jiang
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2010, 31 (06): : 1269 - 1275