Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes

被引:80
|
作者
Bonora, B. [1 ]
Maran, A. [1 ]
Ciciliot, S. [1 ]
Avogaro, A. [1 ]
Fadini, G. P. [1 ]
机构
[1] Univ Padua, Div Metab Dis, Dept Med, Via Giustiniani 2, I-35128 Padua, Italy
来源
关键词
Sensors; Continuous glucose monitoring; Hypoglycaemia; Variability; Calibration; BLOOD-GLUCOSE; VARIABILITY; INSTABILITY; ADOLESCENTS; FREQUENCY; ACCURACY; CHILDREN; INDEX;
D O I
10.1007/s40618-016-0495-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Continuous glucose monitoring (CGM) is being increasingly used in clinical practice. The flash glucose monitoring (FGM) and CGM are different systems of interstitial glucose recording. We aimed to determine the agreement between the factory-calibrated FGM FreeStyle Libre (FSL) and the gold-standard CGM Dexcom G4 Platinum (DG4P). We analyzed data from n = 8 outpatients with type 1 diabetes, who wore the FSL and DG4P for up to 14 days during their habitual life. We aligned FSL and DG4P recordings to obtain paired glucose measures. We calculated correlation coefficients, mean absolute relative difference (MARD), percentages in Clarke error grid areas, time spent in hyperglycaemia, target glycaemia, or hypoglycaemia, as well as glucose variability with both sensors. Comparison with self-monitoring of blood glucose (SMBG) was also performed. Patients varied in terms of age, diabetes duration, and HbA1c (from 5.9 to 9.6 %). In the pooled analysis of 10,020 paired values, there was a good correlation between FSL and DG4P (r (2) = 0.76; MARD = 18.1 +/- 14.8 %) with wide variability among patients. The MARD was significantly higher during days 11-14 than in days 1-10, and during hypoglycaemia (19 %), than in normoglycaemia (16 %) or hyperglycaemia (13 %). Average glucose profiles and MARD versus SMBG were similar between the two sensors. Time spent in normo-, hyper-, or hypoglycaemia, and indexes of glucose variability was similarly estimated by the two sensors. In outpatients with type 1 diabetes, we found good agreement between the FSL and DG4P. No significant difference was detected in the estimation of clinical diagnostic parameters.
引用
收藏
页码:1391 / 1399
页数:9
相关论文
共 50 条
  • [1] Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes
    B. Bonora
    A. Maran
    S. Ciciliot
    A. Avogaro
    G. P. Fadini
    Journal of Endocrinological Investigation, 2016, 39 : 1391 - 1399
  • [2] Accuracy and Reliability of Continuous Glucose Monitoring Systems: A Head-to-Head Comparison
    Luijf, Yoeri M.
    Mader, Julia K.
    Doll, Werner
    Pieber, Thomas
    Farret, Anne
    Place, Jerome
    Renard, Eric
    Bruttomesso, Daniela
    Filippi, Alessio
    Avogaro, Angelo
    Arnolds, Sabine
    Benesch, Carsten
    Heinemann, Lutz
    DeVries, J. Hans
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (08) : 721 - 726
  • [3] A head-to-head comparison of personal and professional continuous glucose monitoring systems in people with type 1 diabetes: Hypoglycaemia remains the weak spot
    Moser, Othmar
    Pandis, Marlene
    Aberer, Felix
    Kojzar, Harald
    Hochfellner, Daniel
    Elsayed, Hesham
    Motschnig, Melanie
    Augustin, Thomas
    Kreuzer, Philipp
    Pieber, Thomas R.
    Sourij, Harald
    Mader, Julia K.
    DIABETES OBESITY & METABOLISM, 2019, 21 (04): : 1043 - 1048
  • [4] Head-to-head comparison of two continuous glucose monitoring systems on a cardio-surgical ICU
    M. A. Punke
    C. Decker
    M. Petzoldt
    D. A. Reuter
    K. H. Wodack
    H. Reichenspurner
    M. Kubik
    S. Kluge
    Journal of Clinical Monitoring and Computing, 2019, 33 : 895 - 901
  • [5] Head-to-head comparison of two continuous glucose monitoring systems on a cardio-surgical ICU
    Punke, M. A.
    Decker, C.
    Petzoldt, M.
    Reuter, D. A.
    Wodack, K. H.
    Reichenspurner, H.
    Kubik, M.
    Kluge, S.
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2019, 33 (05) : 895 - 901
  • [6] THE I-HART CGM STUDY: EIGHT-WEEK HEAD-ON-HEAD COMPARISON OF GLYCAEMIC VARIABILITY BETWEEN FLASH AND CONTINUOUS GLUCOSE MONITORING SYSTEMS IN ADULTS WITH TYPE 1 DIABETES
    Avari, P.
    Jugnee, N.
    Reddy, M.
    Oliver, N.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 : A13 - A14
  • [7] Continuous Glucose Monitoring and Recreational Scuba Diving in Type 1 Diabetes: Head-to-Head Comparison Between Free Style Libre 3 and Dexcom G7 Performance
    Gamarra, Elena
    Careddu, Giovanni
    Fazi, Andrea
    Turra, Valentina
    Morelli, Ambra
    Camponovo, Chiara
    Trimboli, Pierpaolo
    DIABETES TECHNOLOGY & THERAPEUTICS, 2024, 26 (11) : 829 - 841
  • [8] Even Head-to-Head Comparison Studies of Continuous Glucose Monitoring System Performance Can be Biased
    Pleus, Stefan
    Eichenlaub, Manuel
    Waldenmaier, Delia
    Freckmann, Guido
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2024,
  • [9] Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions
    Kropff, J.
    Bruttomesso, D.
    Doll, W.
    Farret, A.
    Galasso, S.
    Luijf, Y. M.
    Mader, J. K.
    Place, J.
    Boscari, F.
    Pieber, T. R.
    Renard, E.
    DeVries, J. H.
    DIABETES OBESITY & METABOLISM, 2015, 17 (04): : 343 - 349
  • [10] A head-to-head comparison between Guardian Connect and FreeStyle Libre systems and an evaluation of user acceptability of sensors in patients with type 1 diabetes
    Yeoh, Ester
    Png, Doanna
    Khoo, Jonathan
    Chee, Ying Jie
    Sharda, Puja
    Low, Serena
    Lim, Su Chi
    Subramaniam, Tavintharan
    DIABETES-METABOLISM RESEARCH AND REVIEWS, 2022, 38 (07)