EFFECT OF TUBING LENGTH PUMP ON FLOW DERIVED INDEX DURING ROTARY BLOOD PUMP SUPPORT

被引:0
|
作者
Khienwad, Thananya [1 ]
Deepankaew, Ronnachit [1 ]
Wannawat, Praemai [1 ]
Naiyanetr, Phornphop [1 ]
机构
[1] Mahidol Univ, Fac Engn, Dept Biomed Engn, 25-25 Phuttamonthon 4 Rd, Salaya 73170, Nakhon Pathom, Thailand
关键词
Rotary blood pump; Mock circulatory loop; Hemodynamic parameter; Pump flow; Pump speed; Tubing length pump;
D O I
10.2316/P.2017.852-034
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The variable of connecting tube length (CTL) in rotary blood pumps (RBPs) is normally existed for the external circulatory support and thus eventually incurs the different outcome as compared to the internal circulatory support. This study aims to determine the effect of the various CTL on the measured parameters from RBP. To serve this purpose, a derived indexes from RBP under mock circulatory loop (MCL) were examined. I-Q which is a slope of the linear regression between the maximal time derivative of pump flow, dQ/dt(max), and its peak to peak, Q(P2P), was determined to evaluate the effect of 3 different CTLs; 130, 50 and 10 cm (length of inlet tube), and 130, 50 and 50 cm (length of outlet tube). Moreover, the parameters dH/dt(max), which are the maximal time derivative of pressure head (H); AOP-LVP, and its peak to peak were derived. The slope of a linear regression between dQ/dt(max) and dH/dt(max) decreased during reduce CTL in contrast to I-Q index. Thus, this study proved that I-Q index was not significant influenced by a hydraulic resistance (including CTL) and therefore can be used to monitor the cardiac function in case of both external and internal RBP support.
引用
收藏
页码:270 / 276
页数:7
相关论文
共 50 条
  • [31] CFD MODELLING OF ROTARY BLOOD PUMP
    Szwast, M.
    Moskal, A.
    Piatkiewicz, W.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (08): : 702 - 702
  • [32] Unloading effect of a rotary blood pump assessed by mathematical modeling
    Vandenberghe, S
    Segers, P
    Meyns, B
    Verdonck, P
    ARTIFICIAL ORGANS, 2003, 27 (12) : 1094 - 1101
  • [33] Investigation of Blood Hemolysis Study in Rotary Blood Pump between Continuous Flow and Pulsatile Flow
    Rungsirikunnan, Chitsanupong
    Mondee, Kamonwan
    Gesprasert, Goragoch
    Naiyanetr, Phornphop
    2018 11TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2018), 2018,
  • [34] Estimating Flow Rate Using the Motor Torque in a Rotary Blood Pump
    Hijikata, Wataru
    Rao, Jun
    Abe, Shodai
    Takatani, Setsuo
    Shinshi, Tadahiko
    SENSORS AND MATERIALS, 2015, 27 (04) : 297 - 308
  • [35] Hemodynamics during Rotary Blood Pump Support with Speed Synchronization in Heart Failure Condition: A Modelling Study
    Htet, Zwe Lin
    Aye, Thin Pa Pa
    Singhavilai, Thamvarit
    Naiyanetr, Phornphop
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 3307 - 3310
  • [36] SAFETY AWARE PUMP-CONTROL FOR A ROTARY ECMO BLOOD PUMP
    Stollenwerk, A.
    Gathmann, F.
    Arens, J.
    Bensberg, R.
    Walter, M.
    Kopp, R.
    Kowalewski, S.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (08): : 617 - 617
  • [37] Pulsatile control of a rotary blood pump - Therapy optimisation with extracorporeal lung support
    Kuehn, Jan
    Stollenwerk, Andre
    Kowalewski, Stefan
    Brendle, Christian
    Walter, Marian
    Leonhardt, Steffen
    Wardeh, Markus
    Rossaint, Rolf
    Kopp, Ruedger
    ATP EDITION, 2015, (10): : 52 - 59
  • [38] Evaluation of Left Ventricular Relaxation in Rotary Blood Pump Recipients Using the Pump Flow Waveform: A Simulation Study
    Moscato, Francesco
    Granegger, Marcus
    Naiyanetr, Phornphop
    Wieselthaler, Georg
    Schima, Heinrich
    ARTIFICIAL ORGANS, 2012, 36 (05) : 470 - 478
  • [39] Baroreflex Sensitivity of an Arterial Wall During Rotary Blood Pump Assistance
    Yambe, Tomoyuki
    Imachi, Kou
    Shiraishi, Yasuyuki
    Yamaguchi, Tasuku
    Shibata, Mune-ichi
    Kameyama, Takeshi
    Yoshizawa, Makoto
    Sugita, Norihiro
    ARTIFICIAL ORGANS, 2009, 33 (09) : 767 - 769
  • [40] EFFECT OF INFLOW CANNULA ON ROTARY BLOOD PUMP SPEED PULSING STRATEGY
    Wong, K. C.
    Timms, D. L.
    Steinseifer, U.
    Schmitz-Rode, T.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (08): : 613 - 613