Preservation of ventricular capture by anodal stimulation after dislodgment of an epicardial electrode

被引:2
|
作者
Römers, H [1 ]
Bracke, FA [1 ]
Meijer, A [1 ]
Van Gelder, BM [1 ]
机构
[1] Catharina Hosp, Dept Cardiol, NL-5623 EJ Eindhoven, Netherlands
来源
关键词
anodal capture; epicardial electrode; electrode dislodgment; cardiac resynchronization therapy;
D O I
10.1111/j.1540-8159.2006.00288.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 58-year-old man had a bipolar epicardial left ventricular (LV) implanted for cardiac resynchronization therapy after a failed transvenous approach. The system was programmed in an LV tip-right ventricular coil configuration, but during follow-up loss of LV capture with recurrence of symptoms occurred. Changing to an LV tip-LV ring configuration restored LV pacing through anodal capture. Loss of cathodal capture was caused by dislodgment of the cathodal electrode due to a broken fixation suture. Anodal capture was used to reinstall resynchronization therapy, which resulted in clinical improvement. There were no adverse effects from anodal stimulation in a follow-up period of 6 months.
引用
收藏
页码:99 / 101
页数:3
相关论文
共 50 条
  • [1] Ventricular capture by anodal pacemaker stimulation
    Occhetta, E
    Bortnik, M
    Marino, P
    EUROPACE, 2006, 8 (05): : 385 - 387
  • [2] Anodal stimulation in pediatric patients with permanent epicardial ventricular pacemakers
    Boscamp, Nicholas S.
    Liberman, Leonardo
    Silver, Eric S.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (09): : 1593 - 1598
  • [3] Anodal right ventricular capture during left ventricular stimulation in CRT-implantable cardioverter defibrillators
    Thibault, B
    Roy, D
    Guerra, PG
    Macle, L
    Dubuc, M
    Gagné, P
    Greiss, I
    Novak, P
    Furlani, A
    Talajic, M
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (07): : 613 - 619
  • [4] POSTOPERATIVE PROGRAMMED VENTRICULAR STIMULATION - RELIABILITY OF EPICARDIAL STIMULATION AFTER SURGERY FOR VENTRICULAR ARRHYTHMIAS
    PAGE, P
    ARCINIEGAS, JG
    PLUMB, VJ
    ZIMMERN, SH
    HENTHORN, RW
    KARP, RB
    WALDO, AL
    UNION MEDICALE DU CANADA, 1983, 112 (11): : 1041 - +
  • [5] Anodal Capture, Cathodal Capture, and Left Ventricular Cardiac Excitation
    Mounsey, J. Paul
    Knisley, Stephen B.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (06) : 650 - 652
  • [6] Left ventricular pseudoaneurysm after epicardial patch electrode placement
    Maisel, WH
    Lukanich, JM
    Bueno, R
    Friedman, PL
    Reimold, SC
    ANNALS OF THORACIC SURGERY, 1998, 66 (06): : 2102 - 2104
  • [7] Right ventricular anodal capture in biventricular stimulation for heart failure: A look at multiple lead models
    Steinhaus, D
    Suleman, A
    Vlach, K
    Germanson, N
    Hebert, K
    McVenes, R
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (05) : 107A - 107A
  • [8] Mechanism of anodal stimulation in cardiac ventricular myocytes
    Ranjan, R
    Chiamvimonvat, N
    Tomaselli, GF
    Marban, E
    BIOPHYSICAL JOURNAL, 1997, 72 (02) : TUAM8 - TUAM8
  • [9] Ventricular tachyarrhythmia treatment and prevention by subthreshold stimulation with stretchable epicardial multichannel electrode array
    Sunwoo, Sung-Hyuk
    Cha, Myung-Jin
    Han, Sang Ihn
    Kang, Hyejeong
    Cho, Ye Seul
    Yeom, Da-Hae
    Park, Chan Soon
    Park, Na Kyeong
    Choi, Seong Woo
    Kim, Sung Joon
    Cha, Gi Doo
    Jung, Dongjun
    Choi, Suji
    Oh, Seil
    Nam, Gi-Byoung
    Hyeon, Taeghwan
    Kim, Dae-Hyeong
    Lee, Seung-Pyo
    SCIENCE ADVANCES, 2023, 9 (13)
  • [10] ELECTROPHYSIOLOGICAL RESPONSES TO ANODAL AND CATHODAL STIMULATION WITH EXTRACELLULAR ELECTRODE
    SEKIYA, S
    TSUTSUMI, T
    YAZAKI, Y
    OHTA, A
    HARUMI, K
    JOURNAL OF MOLECULAR AND CELLULAR CARDIOLOGY, 1985, 17 : 73 - 73