Ventricular tachyarrhythmia treatment and prevention by subthreshold stimulation with stretchable epicardial multichannel electrode array

被引:28
|
作者
Sunwoo, Sung-Hyuk [1 ,2 ,3 ,4 ]
Cha, Myung-Jin [5 ,6 ]
Han, Sang Ihn [1 ,2 ,3 ]
Kang, Hyejeong [1 ,6 ]
Cho, Ye Seul [6 ]
Yeom, Da-Hae [6 ]
Park, Chan Soon [6 ]
Park, Na Kyeong [7 ]
Choi, Seong Woo [8 ]
Kim, Sung Joon [7 ]
Cha, Gi Doo [1 ,2 ,3 ]
Jung, Dongjun [1 ,2 ,3 ]
Choi, Suji [1 ,2 ,3 ]
Oh, Seil [6 ,9 ]
Nam, Gi-Byoung [4 ]
Hyeon, Taeghwan [1 ,2 ,3 ]
Kim, Dae-Hyeong [1 ,2 ,3 ,10 ]
Lee, Seung-Pyo [1 ,6 ,9 ]
机构
[1] Inst Basic Sci IBS, Ctr Nanoparticle Res, Seoul 08826, South Korea
[2] Seoul Natl Univ, Sch Chem & Biol Engn, Seoul 08826, South Korea
[3] Seoul Natl Univ, Inst Chem Proc, Seoul 08826, South Korea
[4] Seoul Natl Univ, Inst Radiat Med, Med Res Ctr, Seoul 03080, South Korea
[5] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Cardiol & Internal Med, Seoul 05505, South Korea
[6] Seoul Natl Univ Hosp, Dept Internal Med, Div Cardiol, Seoul 03080, South Korea
[7] Seoul Natl Univ, Coll Med, Dept Physiol, Seoul 03080, South Korea
[8] Dongguk Univ, Coll Med, Dept Physiol, Gyeongju 38066, South Korea
[9] Seoul Natl Univ, Coll Med, Dept Internal Med, Seoul 03080, South Korea
[10] Seoul Natl Univ, Dept Mat Sci & Engn, Seoul 08826, South Korea
关键词
IMPEDANCE SPECTROSCOPY; LEAD PLACEMENT; TACHYCARDIA; TERMINATION; PACEMAKER; DEFIBRILLATORS; PERFORATION; REDUCTION; THERAPIES; PEDOTPSS;
D O I
10.1126/sciadv.adf6856
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The implantable cardioverter-defibrillator (ICD) is an effective method to prevent sudden cardiac death in high -risk patients. However, the transvenous lead is incompatible with large-area electrophysiological mapping and cannot accommodate selective multichannel precision stimulations. Moreover, it involves high-energy shocks, resulting in pain, myocardial damage, and recurrences of ventricular tachyarrhythmia (VTA). We present a method for VTA treatment based on subthreshold electrical stimulations using a stretchable epicardial multi-channel electrode array, which does not disturb the normal contraction or electrical propagation of the ventri-cle. In rabbit models with myocardial infarction, the infarction was detected by mapping intracardiac electrograms with the stretchable epicardial multichannel electrode array. Then, VTAs could be terminated by sequential electrical stimuli from the epicardial multichannel electrode array beginning with low-energy sub-threshold stimulations. Last, we used these subthreshold stimulations to prevent the occurrence of additional VTAs. The proposed protocol using the stretchable epicardial multichannel electrode array provides opportu-nities toward the development of innovative methods for painless ICD therapy.
引用
收藏
页数:12
相关论文
共 15 条
  • [1] A simultaneous multichannel monophasic action potential electrode array for epicardial repolarization mapping
    Sahakian, AV
    Peterson, MSL
    Hamer, M
    Votapka, T
    Ji, T
    Swiryn, S
    COMPUTERS IN CARDIOLOGY 1998, VOL 25, 1998, 25 : 125 - 128
  • [2] A multichannel monophasic action potential electrode array for simultaneous epicardial and endocardial repolarization mapping
    Shkurovich, S
    Sahakian, AV
    Ji, T
    Votapka, TV
    Swiryn, S
    COMPUTERS IN CARDIOLOGY 1999, VOL 26, 1999, 26 : 281 - 284
  • [3] A simultaneous multichannel monophasic action potential electrode array for in vivo epicardial repolarization mapping
    Sahakian, AV
    Peterson, MSL
    Shkurovich, S
    Hamer, M
    Votapka, T
    Ji, TY
    Swiryn, S
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (03) : 345 - 353
  • [4] Preservation of ventricular capture by anodal stimulation after dislodgment of an epicardial electrode
    Römers, H
    Bracke, FA
    Meijer, A
    Van Gelder, BM
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (01): : 99 - 101
  • [5] Epicardial but not endocardial premature stimulation initiates ventricular tachyarrhythmia in canine in vitro model of long QT syndrome
    Ueda, N
    Zipes, DP
    Wu, JS
    HEART RHYTHM, 2004, 1 (06) : 684 - 694
  • [6] Transretinal electrical stimulation by an intrascleral multichannel electrode array in rabbit eyes
    Nakauchi, K
    Fujikado, T
    Kanda, H
    Morimoto, T
    Choi, JS
    Ikuno, Y
    Sakaguchi, H
    Kamei, M
    Ohji, M
    Yagi, T
    Nishimura, S
    Sawai, H
    Fukuda, Y
    Tano, Y
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2005, 243 (02) : 169 - 174
  • [7] Transretinal electrical stimulation by an intrascleral multichannel electrode array in rabbit eyes
    Kazuaki Nakauchi
    Takashi Fujikado
    Hiroyuki Kanda
    Takeshi Morimoto
    Jun S. Choi
    Yasushi Ikuno
    Hirokazu Sakaguchi
    Motohiro Kamei
    Masahito Ohji
    Tohru Yagi
    Shigeru Nishimura
    Hajime Sawai
    Yutaka Fukuda
    Yasuo Tano
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2005, 243 : 169 - 174
  • [8] PREVENTION OF INDUCTION OF VENTRICULAR-TACHYCARDIA BY SUBTHRESHOLD STIMULATION AT ITS SITE OF ORIGIN
    SHENASA, M
    MEHRA, R
    CIRCULATION, 1990, 82 (04) : 182 - 182
  • [9] Stretchable surface electromyography electrode array patch for tendon location and muscle injury prevention
    Yang, Shuaijian
    Cheng, Jinhao
    Shang, Jin
    Hang, Chen
    Qi, Jie
    Zhong, Leni
    Rao, Qingyan
    He, Lei
    Liu, Chenqi
    Ding, Li
    Zhang, Mingming
    Chakrabarty, Samit
    Jiang, Xingyu
    NATURE COMMUNICATIONS, 2023, 14 (01)
  • [10] Stretchable surface electromyography electrode array patch for tendon location and muscle injury prevention
    Shuaijian Yang
    Jinhao Cheng
    Jin Shang
    Chen Hang
    Jie Qi
    Leni Zhong
    Qingyan Rao
    Lei He
    Chenqi Liu
    Li Ding
    Mingming Zhang
    Samit Chakrabarty
    Xingyu Jiang
    Nature Communications, 14