Device-Induced Hemostatic Disorders in Mechanically Assisted Circulation

被引:28
|
作者
Wang, Shigang [1 ]
Griffith, Bartley P. [1 ]
Wu, Zhongjun J. [1 ,2 ]
机构
[1] Univ Maryland, Sch Med, Dept Surg, 10 South Pine St,MSTF 436, Baltimore, MD 21201 USA
[2] Univ Maryland, A James Clark Sch Engn, Fischell Dept Bioengn, College Pk, MD 20742 USA
基金
美国国家卫生研究院;
关键词
cardiopulmonary bypass; extracorporeal membrane oxygenation; ventricular assist device; mechanical shear stress; hemostatic disorder; VON-WILLEBRAND-FACTOR; GLYCOPROTEIN IB-ALPHA; EXTRACORPOREAL MEMBRANE-OXYGENATION; NEWTONIAN BLOOD-FLOW; PLATELET-DERIVED MICROPARTICLES; DISSIPATIVE PARTICLE DYNAMICS; TOTAL ARTIFICIAL-HEART; ARTERY-BYPASS SURGERY; LONG-TERM SUPPORT; HIGH-SHEAR STRESS;
D O I
10.1177/1076029620982374
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Mechanically assisted circulation (MAC) sustains the blood circulation in the body of a patients undergoing cardiac surgery with cardiopulmonary bypass (CPB) or on ventricular assistance with a ventricular assist device (VAD) or on extracorporeal membrane oxygenation (ECMO) with a pump-oxygenator system. While MAC provides short-term (days to weeks) support and long-term (months to years) for the heart and/or lungs, the blood is inevitably exposed to non-physiological shear stress (NPSS) due to mechanical pumping action and in contact with artificial surfaces. NPSS is well known to cause blood damage and functional alterations of blood cells. In this review, we discussed shear-induced platelet adhesion, platelet aggregation, platelet receptor shedding, and platelet apoptosis, shear-induced acquired von Willebrand syndrome (AVWS), shear-induced hemolysis and microparticle formation during MAC. These alterations are associated with perioperative bleeding and thrombotic events, morbidity and mortality, and quality of life in MCS patients. Understanding the mechanism of shear-induce hemostatic disorders will help us develop low-shear-stress devices and select more effective treatments for better clinical outcomes.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] SLOW RELEASE APROTININ DELIVERY FOR CONTROL OF INTRAUTERINE DEVICE-INDUCED HEMORRHAGE
    TUTTLE, ME
    BAKER, RW
    LAUFE, LE
    JOURNAL OF MEMBRANE SCIENCE, 1980, 7 (03) : 351 - 358
  • [32] Cardiac implantable electronic device-induced tricuspid regurgitation: Implications and management
    Crossley, George H.
    Kowal, Jamie
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2024, 35 (06) : 1217 - 1218
  • [33] Left ventricular assist device-induced molecular changes in the failing myocardium
    Soppa, Gopal K. R.
    Barton, Paul J. R.
    Terracciano, Cesare M. N.
    Yacoub, Magdi H.
    CURRENT OPINION IN CARDIOLOGY, 2008, 23 (03) : 206 - 218
  • [34] Device-Induced Platelet Dysfunction in Patients after Left Ventricular Assist Device Implantation
    Klaeske, K.
    Dieterlen, M.
    Eifert, S.
    Sieg, F.
    Scholz, U.
    Garbade, J.
    Jawad, K.
    Borger, M. A.
    Meyer, A. L.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2021, 40 (04): : S180 - S180
  • [35] Medical device-induced thrombosis: what causes it and how can we prevent it?
    Jaffer, I. H.
    Fredenburgh, J. C.
    Hirsh, J.
    Weitz, J. I.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2015, 13 : S72 - S81
  • [36] Evaluating the biological benefits brought on skin by device-induced mechano-stimulation
    Vedula, S.
    Bourokba, N.
    Caberlotto, E.
    Bornschlogl, T.
    Ghibaudo, M.
    Lynch, B.
    Nouveau, S.
    Querleux, B.
    Ruiz, L.
    Phong, M.
    JOURNAL OF INVESTIGATIVE DERMATOLOGY, 2017, 137 (05) : S56 - S56
  • [37] Investigation of Doppler Indices in Copper Intrauterine Device-induced Heavy Menstrual Bleeding
    Ozbay, Koray
    Sanlikan, Fatih
    CLINICAL AND EXPERIMENTAL OBSTETRICS & GYNECOLOGY, 2022, 49 (06):
  • [38] Blood pressure in a mechanically assisted circulation: "Heat, spirit, perfection"
    Joly, Joanna M.
    Pamboukian, Salpy V.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (04): : 406 - 407
  • [39] Residual Heart Failure on Mechanically Assisted Circulation A Call to Action
    Cogswell, Rebecca
    Rogers, Joseph G.
    JACC-HEART FAILURE, 2022, 10 (07) : 482 - 484
  • [40] Balancing Bleeding and Clotting: The Known Unknowns of Mechanically Assisted Circulation
    Rogers, Joseph G.
    CIRCULATION-HEART FAILURE, 2017, 10 (05)