Device-Induced Hemostatic Disorders in Mechanically Assisted Circulation

被引:28
|
作者
Wang, Shigang [1 ]
Griffith, Bartley P. [1 ]
Wu, Zhongjun J. [1 ,2 ]
机构
[1] Univ Maryland, Sch Med, Dept Surg, 10 South Pine St,MSTF 436, Baltimore, MD 21201 USA
[2] Univ Maryland, A James Clark Sch Engn, Fischell Dept Bioengn, College Pk, MD 20742 USA
基金
美国国家卫生研究院;
关键词
cardiopulmonary bypass; extracorporeal membrane oxygenation; ventricular assist device; mechanical shear stress; hemostatic disorder; VON-WILLEBRAND-FACTOR; GLYCOPROTEIN IB-ALPHA; EXTRACORPOREAL MEMBRANE-OXYGENATION; NEWTONIAN BLOOD-FLOW; PLATELET-DERIVED MICROPARTICLES; DISSIPATIVE PARTICLE DYNAMICS; TOTAL ARTIFICIAL-HEART; ARTERY-BYPASS SURGERY; LONG-TERM SUPPORT; HIGH-SHEAR STRESS;
D O I
10.1177/1076029620982374
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Mechanically assisted circulation (MAC) sustains the blood circulation in the body of a patients undergoing cardiac surgery with cardiopulmonary bypass (CPB) or on ventricular assistance with a ventricular assist device (VAD) or on extracorporeal membrane oxygenation (ECMO) with a pump-oxygenator system. While MAC provides short-term (days to weeks) support and long-term (months to years) for the heart and/or lungs, the blood is inevitably exposed to non-physiological shear stress (NPSS) due to mechanical pumping action and in contact with artificial surfaces. NPSS is well known to cause blood damage and functional alterations of blood cells. In this review, we discussed shear-induced platelet adhesion, platelet aggregation, platelet receptor shedding, and platelet apoptosis, shear-induced acquired von Willebrand syndrome (AVWS), shear-induced hemolysis and microparticle formation during MAC. These alterations are associated with perioperative bleeding and thrombotic events, morbidity and mortality, and quality of life in MCS patients. Understanding the mechanism of shear-induce hemostatic disorders will help us develop low-shear-stress devices and select more effective treatments for better clinical outcomes.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] The impact of shear stress on device-induced platelet hemostatic dysfunction relevant to thrombosis and bleeding in mechanically assisted circulation
    Chen, Zengsheng
    Zhang, Jiafeng
    Li, Tieluo
    Tran, Douglas
    Griffith, Bartley P.
    Wu, Zhongjun J.
    ARTIFICIAL ORGANS, 2020, 44 (05) : E201 - E213
  • [2] Device-induced platelet dysfunction in mechanically assisted circulation increases the risks of thrombosis and bleeding
    Chen, Zengsheng
    Zhang, Jiafeng
    Kareem, Kafayat
    Tran, Douglas
    Conway, Robert G.
    Arias, Katherin
    Griffith, Bartley P.
    Wu, Zhongjun J.
    ARTIFICIAL ORGANS, 2019, 43 (08) : 745 - 755
  • [3] Linking Computational Fluid Dynamics Modeling to Device-Induced Platelet Defects in Mechanically Assisted Circulation
    Zhang, Jiafeng
    Han, Dong
    Chen, Zengsheng
    Wang, Shigang
    Sun, Wenji
    Griffith, Bartley P.
    Wu, Zhongjun J.
    ASAIO JOURNAL, 2024, 70 (12) : 1085 - 1093
  • [4] The impact of shear stress on device-induced platelet hemostatic dysfunction relevant to thrombosis and bleeding in mechanically assisted circulation (vol 44, E201, 2020)
    Chen, Z.
    Zhang, J.
    Li, T.
    Tran, D.
    Griffith, B. P.
    Wu, Z. J.
    ARTIFICIAL ORGANS, 2020, 44 (08) : 907 - 908
  • [5] TASER DEVICE-INDUCED RHABDOMYOLYSIS IS UNLIKELY
    Ho, Jeffrey D.
    Dawes, Donald M.
    JOURNAL OF EMERGENCY MEDICINE, 2011, 40 (01): : 68 - 69
  • [6] In vitro study on device-induced damage to blood cellular components and degradation of von Willebrand factor in a CentriMag pump-assisted circulation
    Wang, Shigang
    Sun, Wenji
    Han, Dong
    Clark, Kiersten P.
    Griffith, Bartley P.
    Wu, Zhongjun J.
    ARTIFICIAL ORGANS, 2024, 48 (09) : 988 - 996
  • [7] TASER DEVICE-INDUCED RHABDOMYOLYSIS IS UNLIKELY REPLY
    Sanford, Janyce M.
    JOURNAL OF EMERGENCY MEDICINE, 2011, 40 (01): : 69 - 69
  • [8] Computational model of device-induced thrombosis and thromboembolism
    Goodman, PD
    Barlow, ET
    Crapo, PM
    Mohammad, SF
    Solen, KA
    ANNALS OF BIOMEDICAL ENGINEERING, 2005, 33 (06) : 780 - 797
  • [9] Mathematical and computational modeling of device-induced thrombosis
    Manning, Keefe B.
    Nicoud, Franck
    Shea, Susan M.
    CURRENT OPINION IN BIOMEDICAL ENGINEERING, 2021, 20
  • [10] Hemocompatibility testing according to ISO 10993-4: Discrimination between pyrogen- and device-induced hemostatic activation
    Stang, Katharina
    Krajewski, Stefanie
    Neumann, Bernd
    Kurz, Julia
    Post, Marcell
    Stoppelkamp, Sandra
    Fennrich, Stefan
    Avci-Adali, Meltem
    Armbruster, Doris
    Schlensak, Christian
    Burgener, Iwan Anton
    Wendel, Hans Peter
    Walker, Tobias
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2014, 42 : 422 - 428