Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study

被引:3
|
作者
Kim, Jihun
Keum, Ki Chang
Lee, Ho
Hong, Chae-Seon
Park, Kwangwoo
Kim, Jin Sung
机构
[1] Department of Radiation Oncology, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemoon-gu, Seoul
基金
新加坡国家研究基金会;
关键词
Image quality; In-treatment; Cone-beam computed tomography; Stereotactic body radiation therapy; Lung; Flattening filter free; MEGAVOLTAGE SCATTER CORRECTION; BODY RADIATION-THERAPY; MODULATED ARC THERAPY; COMPUTED-TOMOGRAPHY; TUMOR MOTION; VERIFICATION; 3D;
D O I
10.1186/s13014-020-01668-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Rotational beam delivery enables concurrent acquisition of cone-beam CT (CBCT), thereby facilitating further geometric verification of patient setup during radiation treatment. However, it is challenging to acquire CBCT during stereotactic body radiation therapy (SBRT) using flattening-filter free X-ray beams, in which a high radiation dose is delivered. This study presents quantitative evaluation results of the image quality in four-dimensional (4D) in-treatment CBCT acquired during SBRT delivery. Methods The impact of megavoltage (MV) scatter and acquisition parameters on the image quality was evaluated using Catphan 503 and XSight lung tracking phantoms. The in-treatment CBCT images of the phantoms were acquired while delivering 16 SBRT plans. The uniformity, contrast, and contrast-to-noise ratio (CNR) of the in-treatment CBCT images were calculated and compared to those of CBCT images acquired without SBRT delivery. Furthermore, the localizing accuracy of the moving target in the XSight lung phantom was evaluated for 10 respiratory phases. Results The CNR of the 3D-reconstucted Catphan CBCT images was reduced from 6.3 to 2.6 due to the effect of MV treatment scatter. Both for the Catphan and XSight phantoms, the CBCT image quality was affected by the tube current and monitor units (MUs) of the treatment plan. The lung target in the XSight tracking phantom was most visible for extreme phases; the mean CNRs of the lung target in the in-treatment CBCT images (with 40 mA tube current) across the SBRT plans were 3.2 for the end-of-exhalation phase and 3.0 for the end-of-inhalation phase. The lung target was localized with sub-millimeter accuracy for the extreme respiratory phases. Conclusions Full-arc acquisition with an increased tube current (e.g. 40 mA) is recommended to compensate for degradation in the CBCT image quality due to unflattened MV beam scatter. Acquiring in-treatment CBCT with a high-MU treatment beam is also suggested to improve the resulting CBCT image quality.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Towards Quality Assurance of 4D Radiotherapy Using a Deformable Lung Phantom
    Serban, M.
    Alexander, A.
    Seuntjens, J.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [22] 4D-CBCT in lung SBRT: characterization and validation using 6D motion platform and antropomorphic phantom
    Boschetti, A.
    Fusella, M.
    Fiandra, C.
    Giglioli, F. R.
    Ragona, R.
    Ricardi, U.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S607 - S607
  • [23] 4D-CBCT frameless image guidance in lung SBRT: Comparison between 4D automatic fusion algorithm and manual fusion
    Gomez-Tejedor Alonso, S.
    Diaz Silvera, C.
    Rincon Perez, M.
    Penedo Cobos, J. M.
    Garcia Castejon, M. A.
    Luna Tirado, J.
    Perez Casas, A. M.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S815 - S815
  • [24] Feasibility Study of Intrafractional 4D CBCT for a Target Monitoring During VMAT-Based SBRT
    Kim, J.
    Lee, E.
    Park, K.
    Kim, J.
    Keum, K.
    Lee, H.
    MEDICAL PHYSICS, 2018, 45 (06) : E383 - E383
  • [25] Fast 3D CBCT imaging for Lung SBRT: Is image quality preserved ?
    De Rijcke, B.
    Van Geeteruyen, R.
    De Rijcke, E.
    Lievens, Y.
    Bogaert, E.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S85 - S86
  • [26] Fundamental Study on Actual Delivered Dose Calculation Technique for Lung SBRT Via CBCT Image Acquired during Phase-Gated VMAT Delivery Using a Dynamic Thorax Phantom
    Iramina, H.
    Nakamura, M.
    Sasaki, M.
    Mizowaki, T.
    MEDICAL PHYSICS, 2024, 51 (10) : 7961 - 7961
  • [27] Validation of 4D CT Image Reconstruction Using a 3D Lung Phantom
    Zheng, Y.
    Girbino, M.
    Oshinsky, R.
    Vroege, L.
    Jesseph, F.
    Lee, S.
    Yuan, J.
    Machtay, M.
    Sohn, J.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [28] Enhancement of 4D CBCT Image Quality Using An Adaptive Prior Image Constrained Compressed Sensing
    Lee, H.
    Yoon, J.
    Lee, E.
    Cho, S.
    Park, K.
    Choi, W.
    Keum, K.
    MEDICAL PHYSICS, 2015, 42 (06) : 3639 - 3639
  • [29] Development of the 4D Phantom Using the 3D Gel Dosimetry for Liver SBRT: A Preliminary Study
    Noh, S.
    Lee, M.
    Yoon, K.
    Kwak, J.
    Cho, B.
    Kim, S.
    Kim, H.
    Jeong, C.
    MEDICAL PHYSICS, 2018, 45 (06) : E326 - E327
  • [30] 4D cone beam CT-based dose assessment for SBRT lung cancer treatment
    Cai, Weixing
    Dhou, Salam
    Cifter, Fulya
    Myronakis, Marios
    Hurwitz, Martina H.
    Williams, Christopher L.
    Berbeco, Ross I.
    Seco, Joao
    Lewis, John H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 554 - 568