IAEA-SSDL bilateral comparisons for diagnostic level air kerma measurement standards

被引:1
|
作者
Csete, Istvan [1 ]
Toroi, Paula [1 ]
Steuer, Andreas [2 ]
Hourdakis, Costas [3 ]
Gabris, Frantisek [4 ]
Jozela, Sibusiso [5 ]
Kosunen, Antti [6 ]
Cardoso, Joao [7 ]
Sochor, Vladimir [8 ]
Persson, Linda [9 ]
Cindro, Denis Glavic [10 ]
Arib, Mehenna [11 ]
Smekhov, Mark [12 ]
机构
[1] IAEA, POB 100, A-1400 Vienna, Austria
[2] BEV, Forschungszentrum Seibersdof, Dosimetrielab, Vienna, Austria
[3] Greek Atom Energy Commiss IRCL GAEC EIM, POB 60092, Athens 15310, Greece
[4] IBA Dosimetry GmbH, Dosimetry Lab, Bahnhofstr 5, D-90592 Schwarzenbruck, Germany
[5] NMISA, Private Bag X34, ZA-0040 Pretoria, South Africa
[6] Radiat & Nucl Safety Author STUK, Laippatie 4, FI-00881 Helsinki, Finland
[7] Inst Super Tecn, LPRS, LMRI, Estr Nacl 10, P-2695066 Bobadela Lrs, Portugal
[8] CMI, Radiova 1, CZ-10200 Prague, Czech Republic
[9] Swedish Radiat Safety Author SSM, Solna Strandvag 96, SE-17116 Stockholm, Sweden
[10] Jozef Stefan Inst, Jamova 39, SI-1000 Ljubljana, Slovenia
[11] KFSHRC, POB 3354, Riyadh 11211, Saudi Arabia
[12] Sheba Med Ctr, Minist Hlth, Radiat Control Dept, Aharon Katsir Ave 7, Ramat Gan, Israel
关键词
Diagnostic X-ray radiation qualities; Ionization chamber; Calibration; Comparison;
D O I
10.1016/j.ejmp.2018.02.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Comparisons of national standard of air kerma for conventional and mammographic diagnostic X-ray radiation qualities were conducted by the IAEA. Eleven secondary standards dosimetry laboratories provided calibration data for Exradin A3 and Radcal RC6M transfer ionization chambers circulated. Each comparison result expressed as the ratio of the participant and IAEA calibration coefficient were within the acceptance limit of +/- 2.5%. From the 67 results of 11 participants and 10 available beam qualities, the comparison result was within its standard uncertainty in 63 cases, and within the expanded (k=2) uncertainty in four cases. Detailed calibration uncertainty budgets from participant laboratories are presented. The relative standard calibration uncertainty of each participant was in the range of 0.5-1.3%. These results indicate that the calibration related uncertainty component is reasonable low for a clinical measurement. In addition to the calibration coefficient, other corrections should be applied for clinical measurement to achieve the recommended accuracy.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 20 条
  • [1] Comparison of the air kerma standards of the IAEA and the BIPM in mammography x-rays
    Kessler, C.
    Burns, D.
    Czap, L.
    Csete, I.
    Gomola, I.
    METROLOGIA, 2013, 50
  • [2] Comparison of air kerma standards for medium-energy x radiation between the MKEH and the IAEA
    Csete, Istvan
    Czap, Ladislav
    Gomola, Igor
    METROLOGIA, 2012, 49
  • [3] International comparisons of air kerma standards in Cs-137 gamma radiation
    Leitner, A
    Witzani, J
    Boutillon, M
    AllisyRoberts, P
    Delaunay, F
    Leroy, E
    Lamperti, P
    Strachotinsky, C
    Csete, I
    METROLOGIA, 1997, 34 (02) : 169 - 175
  • [4] COOMET regional comparison of national measurement standards of air kerma for 137Cs γ radiation at protection level
    Bueermann, L.
    Oborin, A. V.
    Milevsky, V. S.
    Salas, G. Walwyn
    Sukhishvili, S.
    Ginga, I.
    Ivanov, R.
    Gudelis, A.
    Gomola, I.
    METROLOGIA, 2014, 51
  • [5] COOMET regional comparison of national measurement standards of air kerma for 137Cs γ radiation at protection level
    Büermann, L.
    Oborin, A.V.
    Milevsky, V.S.
    Salas, G. Walwyn
    Sukhishvili, S.
    Ginga, I.
    Ivanov, R.
    Gudelis, A.
    Gomola, I.
    Metrologia, 2014, 51 (1A)
  • [6] Comparisons of air kerma and absorbed dose to water standards in Co-60 radiation beams for radiotherapy
    Kottler, Christian
    Steurer, Andreas
    Delaunay, Frank
    de Pooter, Jacco
    Mulas, Cristina Garcia
    Mihailescu, Liviu-Cristian
    METROLOGIA, 2020, 57 (1A)
  • [7] Bilateral comparison COOMET.RI(I)-S2 of the air-kerma standards of the IAEA and the VNIIM, Russian Federation in medium energy x-rays
    Czap, L.
    Gomola, I.
    Oborin, A. V.
    Villevalde, A. Y.
    METROLOGIA, 2014, 51
  • [8] Comparisons of the standards for air kerma of the GUM and the BIPM for 60Co and 137Cs gamma radiation
    Allisy-Roberts, P. J.
    Kessler, C.
    Burns, D. T.
    Derlacinski, M.
    Kokocinski, J.
    METROLOGIA, 2011, 48
  • [9] Comparison of the air kerma standards for 137Cs and 60Co gamma-ray beams between the IAEA and the NIST
    Minniti, Ronaldo
    Czap, Ladislav
    METROLOGIA, 2011, 48
  • [10] Comparison of air kerma measurements between the PTB and the IAEA for x-radiation qualities used in general diagnostic radiology and mammography
    Csete, Istvan
    Bueermann, Ludwig
    Gomola, Igor
    Girzikowsky, Reinhard
    METROLOGIA, 2013, 50