Rapid gradient-echo imaging

被引:79
|
作者
Hargreaves, Brian A. [1 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
基金
美国国家卫生研究院;
关键词
gradient echo; steady state; spoiling; balanced SSFP; STATE FREE PRECESSION; NUCLEAR MAGNETIC-RESONANCE; STEADY-STATE; TRANSVERSE MAGNETIZATION; CLINICAL-APPLICATIONS; SENSORY STIMULATION; PULSE SEQUENCES; BALANCED SSFP; FLIP ANGLES; IN-PHASE;
D O I
10.1002/jmri.23742
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Gradient-echo sequences are widely used in magnetic resonance imaging (MRI) for numerous applications ranging from angiography to perfusion to functional MRI. Compared with spin-echo techniques, the very short repetition times of gradient-echo methods enable very rapid 2D and 3D imaging, but also lead to complicated steady states. Signal and contrast behavior can be described graphically and mathematically, and depends strongly on the type of spoiling: fully balanced (no spoiling), gradient spoiling, or radiofrequency (RF)-spoiling. These spoiling options trade off between high signal and pure T1 contrast, while the flip angle also affects image contrast in all cases, both of which can be demonstrated theoretically and in image examples. As with spin-echo sequences, magnetization preparation can be added to gradient-echo sequences to alter image contrast. Gradient-echo sequences are widely used for numerous applications such as 3D perfusion imaging, functional MRI, cardiac imaging, and MR angiography. J. Magn. Reson. Imaging 2012; 36:13001313. (C) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1300 / 1313
页数:14
相关论文
共 50 条
  • [31] RHEUMATOID-ARTHRITIS LESIONS OF THE WRIST EXAMINED BY RAPID GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING
    MESKE, S
    FRIEDBURG, H
    HENNIG, J
    REINBOLD, W
    STAPPERT, K
    SCHUMICHEN, C
    SCANDINAVIAN JOURNAL OF RHEUMATOLOGY, 1990, 19 (03) : 235 - 238
  • [32] GRADIENT-ECHO MR IMAGING OF THE LUMBAR SPINE - COMPARISON WITH SPIN-ECHO TECHNIQUE
    WATANABE, AT
    TEITELBAUM, GP
    LUFKIN, RB
    TSURUDA, JS
    JINKINS, JR
    BRADLEY, WG
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (03) : 410 - 414
  • [33] Vessel architecture imaging using multiband gradient-echo/spin-echo EPI
    Zhang, Ke
    Yun, Seong Dae
    Triphan, Simon M. F.
    Sturm, Volker J.
    Buschle, Lukas R.
    Hahn, Artur
    Heiland, Sabine
    Bendszus, Martin
    Schlemmer, Heinz-Peter
    Shah, N. Jon
    Ziener, Christian H.
    Kurz, Felix T.
    PLOS ONE, 2019, 14 (08):
  • [34] GRASE (GRADIENT-ECHO AND SPIN-ECHO) IMAGING - A NOVEL FAST MRI TECHNIQUE
    OSHIO, K
    FEINBERG, DA
    MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) : 344 - 349
  • [35] GRASE (GRADIENT-ECHO AND SPIN-ECHO) MR IMAGING - A NEW FAST CLINICAL IMAGING TECHNIQUE
    FEINBERG, DA
    OSHIO, K
    RADIOLOGY, 1991, 181 (02) : 597 - 602
  • [36] OPTIMIZATION OF GRADIENT-ECHO IMAGING PARAMETERS FOR INTRACAVAL FILTERS AND TRAPPED THROMBOEMBOLI
    TEITELBAUM, GP
    ORTEGA, HV
    VINITSKI, S
    CLARK, RA
    WATANABE, AT
    MATSUMOTO, AH
    RIFKIN, MD
    BARTH, KH
    RADIOLOGY, 1990, 174 (03) : 1013 - 1019
  • [37] MAGNETIZATION-PREPARED CARDIAC IMAGING USING GRADIENT-ECHO ACQUISITION
    LIU, Y
    RIEDERER, SJ
    EHMAN, RL
    MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) : 271 - 275
  • [38] Theory and application of static field inhomogeneity effects in gradient-echo imaging
    Reichenbach, JR
    Venkatesan, R
    Yablonskiy, DA
    Thompson, MR
    Lai, S
    Haacke, EM
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (02): : 266 - 279
  • [39] Combined spin- and gradient-echo perfusion-weighted imaging
    Schmiedeskamp, Heiko
    Straka, Matus
    Newbould, Rexford D.
    Zaharchuk, Greg
    Andre, Jalal B.
    Olivot, Jean-Marc
    Moseley, Michael E.
    Albers, Gregory W.
    Bammer, Roland
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (01) : 30 - 40
  • [40] Importance of clot structure in gradient-echo magnetic resonance imaging of hematoma
    Taber, KH
    Hayman, LA
    Herrick, RC
    Kirkpatrick, JB
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1996, 6 (06): : 878 - 883