The wear pattern in metal-on-metal hip prostheses

被引:48
|
作者
Anissian, HL [1 ]
Stark, A
Good, V
Dahlstrand, H
Clarke, IC
机构
[1] Karolinska Hosp, Dept Orthopaed Surg, S-17176 Stockholm, Sweden
[2] Loma Linda Univ, Med Ctr, Dept Orthopaed Surg, Loma Linda, CA 92350 USA
来源
关键词
implants; hip-joint simulator; wear; run-in phase; steady-state phase; metal-on-metal prosthesis;
D O I
10.1002/jbm.1068
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accumulated data suggest that there is a strong correlation between polyethylene wear and osteolysis, which ultimately leads to prosthetic loosening. Second-generation metal-on-metal prostheses have been introduced, with an eye toward resolving this wear-induced osteolysis problem. The metal particles and ions are biologically active and can affect the cell homeostasis. Thus, defining the wear pattern and ratio of a given metal-on-metal prosthesis system is desirable. An early high-wear or run-in phase followed by a low-wear phase or steady state has been suggested for metal-on-metal hip prostheses. The aim of this study was to define the wear pattern of metal-on-metal bearings. The prosthesis systems were tested in a joint simulator. An early accelerated wear phase transformed to a slower wear phase after 700,000 cycles. The run-in and steady-state wear rates for combined head and liner averaged 2.22 mm(3)/Mc and 1.0 mm(3)/ million cycles, respectively. The metal-on-metal prosthesis featured a biphasic wear trend with the wear ratio in run in being more than twofold higher than the steady-state phase. (C) 2001 John Wiley & Sons, Inc.
引用
收藏
页码:673 / 678
页数:6
相关论文
共 50 条
  • [1] A Lexicon for Wear of Metal-on-Metal Hip Prostheses
    McKellop, Harry A.
    Hart, Alister
    Park, Sang-Hyun
    Hothi, Harry
    Campbell, Pat
    Skinner, John A.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2014, 32 (09) : 1221 - 1233
  • [2] Wear of surface engineered metal-on-metal hip prostheses
    J. Fisher
    X. Q. Hu
    T. D. Stewart
    S. Williams
    J. L. Tipper
    E. Ingham
    M. H. Stone
    C. Davies
    P. Hatto
    J. Bolton
    M. Riley
    C. Hardaker
    G. H. Isaac
    G. Berry
    Journal of Materials Science: Materials in Medicine, 2004, 15 : 225 - 235
  • [3] Wear of surface engineered metal-on-metal hip prostheses
    Fisher, J
    Hu, XQ
    Stewart, TD
    Williams, S
    Tipper, JL
    Ingham, E
    Stone, MH
    Davies, C
    Hatto, P
    Bolton, J
    Riley, M
    Hardaker, C
    Isaac, GH
    Berry, G
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (03) : 225 - 235
  • [4] The wear of metal-on-metal total hip prostheses measured in a hip simulator
    Scholes, SC
    Green, SM
    Unsworth, A
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H6) : 523 - 530
  • [5] Influence of simulator kinematics on the wear of metal-on-metal hip prostheses
    Firkins, P.J.
    Tipper, J.L.
    Ingham, E.
    Stone, M.H.
    Farrar, R.
    Fisher, J.
    Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2001, 215 (01) : 119 - 122
  • [6] A study of the wear of explanted metal-on-metal resurfacing hip prostheses
    Joyce, Thomas J.
    Langton, David J.
    Nargol, Antoni V. F.
    TRIBOLOGY INTERNATIONAL, 2011, 44 (05) : 517 - 522
  • [7] Influence of simulator kinematics on the wear of metal-on-metal hip prostheses
    Firkins, PJ
    Tipper, JL
    Ingham, E
    Stone, MH
    Farrar, R
    Fisher, J
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H1) : 119 - 121
  • [8] Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses
    Lord, J. K.
    Langton, D. J.
    Nargol, A. V. F.
    Joyce, T. J.
    WEAR, 2011, 272 (01) : 79 - 87
  • [9] Metal-on-metal pairs for hip prostheses
    Plitz, W
    Veihelmann, A
    Pellengahr, C
    ORTHOPADE, 2003, 32 (01): : 17 - 22
  • [10] Imaging of Metal-on-Metal Hip Prostheses
    Squire, Matthew W.