Proximal femoral nail antirotation against dynamic hip screw for unstable trochanteric fractures; a prospective randomized comparison

被引:39
|
作者
Zehir, Sinan [1 ]
Zehir, R. [2 ]
Zehir, Sultan [3 ]
Azboy, I. [4 ]
Haykir, N. [5 ]
机构
[1] Hitit Univ, Res & Training Hosp, Orthoped & Traumatol Clin, TR-19100 Merkez, Corum, Turkey
[2] Carsamba State Hosp, Samsun, Turkey
[3] Corum Publ Hlth Ctr, Corum, Turkey
[4] Dicle Univ, Res & Training Hosp, Orthoped & Traumatol Clin, Diyarbakir, Turkey
[5] Kartal Res & Training Hosp, Istanbul, Turkey
关键词
Hip fracture; Proximal femoral nail; Hip screw; Osteoporosis; FIXATION; CUTOUT;
D O I
10.1007/s00068-014-0463-y
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We sought to determine whether intramedullary fixation with proximal femoral nail antirotation produces comparable outcomes to dynamic hip screw in the treatment of unstable trochanteric fractures. Patients were randomly allocated to receive proximal femoral nail antirotation (Group 1, n = 96, mean age; 77.22 +/- A 6.82 years) or dynamic hip screw (Group 2, n = 102, mean age; 76.86 +/- A 6.74 years). Outcome measures were time of operation and fluoroscopy, amount of blood loss and occurrence of surgery-related complications. Tip-apex distance and femoral neck shortening were also evaluated. Patients were evaluated at the sixth month to assess the recovery of walking ability. Survival information was obtained from a civil registry. Operative and fluoroscopy times were significantly shorter and blood loss was significantly lower in Group 1 than those in Group 2. Complication rates, mean tip-apex indices and recovery of walking ability were similar between groups, whereas independent walking was more common in Group 1 than in Group 2. Until the sixth month, screw cutout occurred in eight (7.8 %) and seven (7.3 %) patients in Group 1 and Group 2, respectively (p = 0.88). Three-year survival rate was 61.6 +/- A 9.4 vs 57.3 +/- A 9.7 % in Group 1 and Group 2, respectively (p = 0.50). Proximal femoral nail antirotation technique offers better recovery than dynamic hip screw, whereas both techniques possess the same risk of postoperative complications.
引用
收藏
页码:393 / 400
页数:8
相关论文
共 50 条
  • [31] Comparison of unstable inter-trochanteric femur fracture treated with Dynamic Hip Screw and Proximal Femur Nail
    Saleem, Muhammad
    Ahmed, Masroor
    Kumar, Mukesh
    Sahar, Khurram
    Hussain, Ghulam
    Bux, Muhammad
    RAWAL MEDICAL JOURNAL, 2020, 45 (03): : 648 - 651
  • [32] Treatment of Trochanteric Fractures with the PFNA (Proximal Femoral Nail Antirotation) Nail System Report of Early Results
    Takigami, Iori
    Matsumoto, Kazu
    Ohara, Akira
    Yamanaka, Kazuteru
    Naganawa, Toshitaka
    Ohashi, Minoru
    Date, Kazuto
    Shimizu, Katsuji
    BULLETIN OF THE HOSPITAL FOR JOINT DISEASES, 2008, 66 (04): : 276 - 279
  • [33] Biomechanical comparison of dynamic hip screw and Gamma nail for the treatment of unstable trochanteric fractures: a finite element study
    Huang, Xiaowei
    Yu, Baoqing
    Gu, Yong
    Li, Zexiang
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (05): : 7867 - 7874
  • [34] Is proximal femoral nail antirotation superior to gamma nail and dynamic hip screw in treatment of intertrochanteric fractures? a pairwise and network meta-analysis
    Yin, Si
    Dai, Xing
    Zhang, Dangfeng
    Du, Heng
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2016, 9 (12): : 22993 - 23007
  • [35] Evaluation of dynamic hip screw plate v/s proximal femoral nail for unstable inter-trochanteric fracture femur
    Pundkar, A. G.
    Modi, N. S.
    Baitule, R. W.
    Pundkar, G. N.
    JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2016, 4 (03): : 283 - 287
  • [36] TROCHANTERIC FRACTURES OF THE FEMUR - A RANDOMIZED PROSPECTIVE TRIAL COMPARING THE JEWETT NAIL-PLATE WITH THE DYNAMIC HIP SCREW
    ESSER, MP
    KASSAB, JY
    JONES, DHA
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (04): : 557 - 560
  • [37] JEWETT NAIL-PLATE OR AO DYNAMIC HIP SCREW FOR TROCHANTERIC FRACTURES - A RANDOMIZED PROSPECTIVE CONTROLLED TRIAL
    BANNISTER, GC
    GIBSON, AGF
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1983, 65 (02): : 218 - 218
  • [38] Proximal Femoral Nail Antirotation versus Gamma3 nail for intramedullary nailing of unstable trochanteric fractures. A randomised comparative study
    Vaquero, J.
    Munoz, J.
    Prat, S.
    Ramirez, C.
    Aguado, H. J.
    Moreno, E.
    Perez, M. D.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2012, 43 : S47 - S54
  • [39] The Proximal Femoral Nail Antirotation (PFNA) in the Treatment of Proximal Femoral Fractures
    Kristek, Dalibor
    Lovric, Ivan
    Kristek, Jozo
    Biljan, Marta
    Kristek, Gordana
    Sakic, Kata
    COLLEGIUM ANTROPOLOGICUM, 2010, 34 (03) : 937 - 940
  • [40] Comparison of cutout resistance of dynamic condylar screw and proximal femoral nail in reverse oblique trochanteric fractures: A biomechanical study
    Cheema, Gursimrat Singh
    Rastogi, Amit
    Singh, Vakil
    Goel, Satish Chandra
    Mishra, Diwakar
    Arora, Sumit
    INDIAN JOURNAL OF ORTHOPAEDICS, 2012, 46 (03) : 259 - 265