Physical Model-Based Metal Artifact Reduction (MAR) Scheme for a 3D Cone-Beam CT Extremity Imaging System

被引:1
|
作者
Yang, D. [1 ]
Senn, R. A. [1 ]
Packard, N. [1 ]
Richard, S. [1 ]
Yorkston, J. [1 ]
机构
[1] Carestream Hlth Inc, Res & Innovat Labs, Rochester, NY 14615 USA
关键词
CBCT imaging system; CBCT metal artifacts reduction; image reconstruction; COMPUTED-TOMOGRAPHY; ALGORITHM;
D O I
10.1117/12.2006207
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In Cone Beam CT Imaging, metallic and other dense objects, such as implantable orthopedic appliances, surgical clips and staples, and dental fillings, are often acquired as part of the image dataset. These high-density, high atomic mass objects attenuate X-rays in the diagnostic energy range much more strongly than soft tissue or bony structures, resulting in photon starvation at the detector. In addition, signal behind the metal objects suffer from increased quantum noise, scattered radiation, and beam hardening. All of these effects combine to create nonlinearities which are further amplified by the reconstruction algorithm, such as conventional filtered back-projection (FBP), producing strong artifacts in the form of streaking. They reduce image quality by masking soft tissue structures, not only in the immediate vicinity of the dense object, but also throughout the entire image volume. A novel, physical-model-based, metal-artifact reduction scheme (MAR) is proposed to mitigate the metal-induced artifacts. The metal objects are segmented in the projection domain, and a physical model based method is adopted to fill in the segmented area. The FDK1 reconstruction algorithm is then used for the final reconstruction.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Moving metal artifact reduction in cone-beam CT scans with implanted cylindrical gold markers
    Toftegaard, Jakob
    Fledelius, Walther
    Seghers, Dieter
    Huber, Michael
    Brehm, Marcus
    Worm, Esben S.
    Elstrom, Ulrik V.
    Poulsen, Per R.
    MEDICAL PHYSICS, 2014, 41 (12)
  • [22] A 3D imaging system for dental imaging based on digital tomosynthesis and cone beam CT
    Badea, C
    Kolitsi, Z
    Pallikarakis, N
    MEDICON 2001: PROCEEDINGS OF THE INTERNATIONAL FEDERATION FOR MEDICAL & BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2001, : 739 - 741
  • [23] INDEPENDENT COMPONENT ANALYSIS BASED RING ARTIFACT REDUCTION IN CONE-BEAM CT IMAGES
    Chen, Yen-Wei
    Duan, Guifen
    2009 16TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, VOLS 1-6, 2009, : 4189 - 4192
  • [24] Cone-Beam Angle Dependency of 3D Models Computed from Cone-Beam CT Images
    Cho, Myung Hye
    Hegazy, Mohamed A. A.
    Cho, Min Hyoung
    Lee, Soo Yeol
    SENSORS, 2022, 22 (03)
  • [25] Known-Component Metal Artifact Reduction (KC-MAR) for Intraoperative Cone-Beam CT in Spine Surgery: A Clinical Pilot Study
    Zhang, X.
    Uneri, A.
    Doerr, S.
    Stayman, J.
    Zygourakis, C.
    Lo, S.
    Theodore, N.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E131 - E131
  • [26] 3D cone-beam CT reconstruction for circular trajectories
    Grass, M
    Köhler, T
    Proksa, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02): : 329 - 347
  • [27] Imaging Study of Pseudo-CT Synthesized From Cone-Beam CT Based on 3D CycleGAN in Radiotherapy
    Sun, Hongfei
    Fan, Rongbo
    Li, Chunying
    Lu, Zhengda
    Xie, Kai
    Ni, Xinye
    Yang, Jianhua
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [28] A nonconvex model-based combined geometric calibration scheme for micro cone-beam CT with irregular trajectories
    Li, Guang
    Chen, Xue
    You, Chenyu
    Huang, Xinhai
    Deng, Zhenhao
    Luo, Shouhua
    MEDICAL PHYSICS, 2023, 50 (05) : 2759 - 2774
  • [29] Comparison of State-Of-The-Art Interpolation-Based Metal Artifact Reduction (MAR) Algorithms for Cone-Beam Computed Tomography (CBCT)
    Seghers, D.
    Berkus, T.
    Oelhafen, M.
    Munro, P.
    Star-Lack, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3628 - 3628
  • [30] 3D ROI imaging for cone-beam computed tomography
    Wiegert, J
    Bertram, M
    Wulff, J
    Schäfer, D
    Weese, R
    Netsch, T
    Schombergc, H
    Rose, G
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 7 - 12