Evaluation of the calibration procedure of active personal dosemeters for interventional radiology

被引:8
|
作者
Bordy, J. -M. [1 ]
Daures, J. [1 ]
Clairand, I. [2 ]
Denoziere, M. [1 ]
Donadille, L. [2 ]
d'Errico, F. [3 ,4 ]
Gouriou, J. [1 ]
Itie, C. [2 ]
Struelens, L. [5 ]
机构
[1] CEA Saclay, CEA LIST, Lab Natl Henri Becquerel, F-91191 Gif Sur Yvette, France
[2] IRSN, F-92224 Fontenay Aux Roses, France
[3] Univ Pisa, Fac Ingn, I-56126 Pisa, Italy
[4] Yale Univ, New Haven, CT 06520 USA
[5] Belgian Nucl Res Ctr SCK CEN, B-2400 Mol, Belgium
关键词
D O I
10.1093/rpd/ncn225
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
An overview of the use of active personal dosemeters (APD) in interventional radiology is presented. It is based on the work done by the working package 7 of the CONRAD coordinated action supported by the EC within the frame of the 6th FP. This study was done in collaboration with the working package 4 of CONRAD to deal with the calculations required for studying the new calibration facility. The main requirements of the standard for the APD and the difficulties caused by the use of pulsed radiations are presented through the results of an intercomparison organised in a realistic calibration facility similar to the workplace situation in interventional radiology. The main characteristics of this facility are presented.
引用
收藏
页码:87 / 92
页数:6
相关论文
共 50 条
  • [41] Systematic study on the influence of inductive chargers on active personal dosemeters
    Ketelhut, Steffen
    Zutz, Hayo
    Hupe, Oliver
    JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (04)
  • [42] Evaluation of CdZnTe spectrometer performance in measuring energy spectra during interventional radiology procedure
    Lin Chai
    Lian Chen
    Cui-Ping Yang
    Dong-Dong Zhou
    Meng-Meng Yang
    Wei-Wei Qu
    Gao-Long Zhang
    Da-Qian Hei
    Shou-Ping Xu
    Xin-Jian Chen
    NuclearScienceandTechniques, 2019, 30 (09) : 91 - 96
  • [43] Active personal dosemeters for individual monitoring and other new developments
    Bolognese-Milsztajn, T
    Ginjaume, M
    Luszik-Bhadra, M
    Vanhavere, F
    Wahl, W
    Weeks, A
    RADIATION PROTECTION DOSIMETRY, 2004, 112 (01) : 141 - 168
  • [44] CHARACTERIZATION OF A THIN SILICON SENSOR FOR ACTIVE NEUTRON PERSONAL DOSEMETERS
    Takada, M.
    Nunomiya, T.
    Nakamura, T.
    Matsumoto, T.
    Masuda, A.
    RADIATION PROTECTION DOSIMETRY, 2016, 170 (1-4) : 213 - 217
  • [45] Efficiency of personal dosimetry methods in vascular interventional radiology
    Bacchim Neto, Fernando Antonio
    Fattori Alves, Allan Felipe
    Mascarenhas, Yvone Maria
    Giacomini, Guilherme
    Pelegrino Bastos Maues, Nadine Helena
    Nicolucci, Patrcia
    Macedo de Freitas, Carlos Clayton
    Alvarez, Matheus
    de Pina, Diana Rodrigues
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 37 : 58 - 67
  • [46] Overview of active personal dosemeters for individual monitoring in the European union
    Ginjaume, M.
    Bolognese-Milsztajn, T.
    Luszik-Bhadra, M.
    Vanhavere, F.
    Wahl, W.
    Weeks, A.
    RADIATION PROTECTION DOSIMETRY, 2007, 125 (1-4) : 261 - 266
  • [47] Intervention or procedure? What's at stake for interventional radiology
    Joffre, F.
    JOURNAL DE RADIOLOGIE, 2008, 89 (03): : 287 - 287
  • [48] Skin entrance radiation dose in an interventional radiology procedure
    Chu, Robert Y. L.
    Thomas, George
    Maqbool, Feroz
    HEALTH PHYSICS, 2006, 91 (01): : 41 - 46
  • [49] Skin entrance radiation dose in an interventional radiology procedure
    Chu, R
    Thomas, G
    Maqbool, F
    MEDICAL PHYSICS, 2005, 32 (06) : 1908 - 1908