Hybrid ultra-low-field MRI and magnetoencephalography system based on a commercial whole-head neuromagnetometer

被引:83
|
作者
Vesanen, Panu T. [1 ]
Nieminen, Jaakko O. [1 ]
Zevenhoven, Koos C. J. [1 ]
Dabek, Juhani [1 ]
Parkkonen, Lauri T. [1 ,2 ]
Zhdanov, Andrey V. [1 ,3 ]
Luomahaara, Juho [4 ,5 ]
Hassel, Juha [4 ]
Penttila, Jari [5 ]
Simola, Juha [2 ]
Ahonen, Antti I. [2 ]
Makela, Jyrki P. [3 ]
Ilmoniemi, Risto J. [1 ]
机构
[1] Aalto Univ, Sch Sci, Dept Biomed Engn & Computat Sci, Espoo, Finland
[2] Elekta Oy, Helsinki, Finland
[3] Univ Helsinki, Cent Hosp, HUSLAB, BioMag Lab, Helsinki, Finland
[4] VTT Tech Res Ctr Finland, Espoo, Finland
[5] Aivon Oy, Espoo, Finland
基金
芬兰科学院;
关键词
ultra-low-field MRI; magnetoencephalography; MEG-MRI; superconducting quantum interference device; MICROTESLA MRI; MU-T; BRAIN; CONTRAST; NOISE;
D O I
10.1002/mrm.24413
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Ultra-low-field MRI uses microtesla fields for signal encoding and sensitive superconducting quantum interference devices for signal detection. Similarly, modern magnetoencephalography (MEG) systems use arrays comprising hundreds of superconducting quantum interference device channels to measure the magnetic field generated by neuronal activity. In this article, hybrid MEG-MRI instrumentation based on a commercial whole-head MEG device is described. The combination of ultra-low-field MRI and MEG in a single device is expected to significantly reduce coregistration errors between the two modalities, to simplify MEG analysis, and to improve MEG localization accuracy. The sensor solutions, MRI coils (including a superconducting polarizing coil), an optimized pulse sequence, and a reconstruction method suitable for hybrid MEG-MRI measurements are described. The performance of the device is demonstrated by presenting ultra-low-field-MR images and MEG recordings that are compared with data obtained with a 3T scanner and a commercial MEG device. Magn Reson Med, 2013. (c) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1795 / 1804
页数:10
相关论文
共 50 条
  • [1] Noise amplification in parallel whole-head ultra-low-field magnetic resonance imaging using 306 detectors
    Lin, Fa-Hsuan
    Vesanen, Panu T.
    Nieminen, Jaakko O.
    Hsu, Yi-Cheng
    Zevenhoven, Koos C. J.
    Dabek, Juhani
    Parkkonen, Lauri T.
    Zhdanov, Andrey
    Ilmoniemi, Risto J.
    MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (02) : 595 - 600
  • [2] A whole-head magnetoencephalography system with compact axial gradiometer structure
    Lee, Y. H.
    Yu, K. K.
    Kwon, H.
    Kim, J. M.
    Kim, K.
    Park, Y. K.
    Yang, H. C.
    Chen, K. L.
    Yang, S. Y.
    Horng, H. E.
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2009, 22 (04):
  • [3] The stability of source localization in a whole-head magnetoencephalography system demonstrated by auditory evoked field measurements
    Chen, Kuen-Lin
    Yang, Hong-Chang
    Tsai, Sung-Ying
    Liu, Yu-Wei
    Liao, Shu-Hsien
    Horng, Herng-Er
    Lee, Yong-Ho
    Kwon, Hyukchan
    JOURNAL OF APPLIED PHYSICS, 2011, 110 (07)
  • [4] Electromagnet design for ultra-low-field MRI
    Shen, Sheng
    Wu, Jiamin
    Guo, Pan
    Wang, Hongyi
    Chen, Fangge
    Meng, Fanqin
    Xu, Zheng
    INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2020, 63 (02) : 267 - 278
  • [5] Gradient Coil Design and Optimization for an Ultra-Low-Field MRI System
    Shen, Sheng
    Koonjoo, Neha
    Kong, Xiaohan
    Rosen, Matthew S.
    Xu, Zheng
    APPLIED MAGNETIC RESONANCE, 2022, 53 (06) : 895 - 914
  • [6] Gradient Coil Design and Optimization for an Ultra-Low-Field MRI System
    Sheng Shen
    Neha Koonjoo
    Xiaohan Kong
    Matthew S. Rosen
    Zheng Xu
    Applied Magnetic Resonance, 2022, 53 : 895 - 914
  • [7] Rotary Scanning Acquisition in Ultra-Low-Field MRI
    Hsu, Yi-Cheng
    Zevenhoven, Koos C. J.
    Chu, Ying-Hua
    Dabek, Juhani
    Ilmoniemi, Risto J.
    Lin, Fa-Hsuan
    MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (06) : 2255 - 2264
  • [8] Ultra-low-field MRI for the detection of liquid explosives
    Espy, M.
    Flynn, M.
    Gomez, J.
    Hanson, C.
    Kraus, R.
    Magnelind, P.
    Maskaly, K.
    Matlashov, A.
    Newman, S.
    Owens, T.
    Peters, M.
    Sandin, H.
    Savukov, I.
    Schultz, L.
    Urbaitis, A.
    Volegov, P.
    Zotev, V.
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2010, 23 (03):
  • [9] Development of Compact Ultra-Low-Field MRI System Using an Induction Coil
    Oyama, Daisuke
    Adachi, Yoshiaki
    Higuchi, Masanori
    Tsuyuguchi, Naohiro
    Uehara, Gen
    IEEE TRANSACTIONS ON MAGNETICS, 2017, 53 (11)
  • [10] BabyMEG: A whole-head pediatric magnetoencephalography system for human brain development research
    Okada, Yoshio
    Hamalainen, Matti
    Pratt, Kevin
    Mascarenas, Anthony
    Miller, Paul
    Han, Menglai
    Robles, Jose
    Cavallini, Anders
    Power, Bill
    Sieng, Kosal
    Sun, Limin
    Lew, Seok
    Doshi, Chiran
    Ahtam, Banu
    Dinh, Christoph
    Esch, Lorenz
    Grant, Ellen
    Nummenmaa, Aapo
    Paulson, Douglas
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (09):