An efficiency comparison of four heat and moisture exchangers used in the laryngectomized patient

被引:24
|
作者
Grolman, W
Blom, ED
Branson, RD
Schouwenburg, PF
Hamaker, RC
机构
[1] METHODIST HOSP,HEAD & NECK SURG ASSOCIATES,INDIANAPOLIS,IN
[2] UNIV CINCINNATI,MED CTR,DIV TRAUMA CRIT CARE,CINCINNATI,OH 45267
来源
LARYNGOSCOPE | 1997年 / 107卷 / 06期
关键词
D O I
10.1097/00005537-199706000-00017
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Bypassing the upper airway places the burden of humidification on the lower airway. For this reason passive heat and moisture exchangers (HMEs) are used in the laryngectomized patient in an attempt to minimize the effect of lost upper airway function. We measured efficiency and airflow resistance and calculated the costs of four HMEs used in the laryngectomized patient. The HMEs were measured according a modified International Standards Organization (ISO) 9360 standard. The airflow resistance was measured at flow rates of 15, 30, and 60 L/min. The measurements were repeated three times. Costs were calculated with two realistic scenarios. The study found that there are significant differences in moisture output and airflow resistance between the HMEs tested. There are major daily cost differences between these devices. This study shows that filter material and size influence the HME's moisture output efficiency and airflow resistance considerably. The construction differences and filter and housing type have great influence on the HME's daily costs. We believe that knowledge of the efficiency in combination with the average daily costs of the HMEs allows the clinician to make a balanced choice of which filter to use.
引用
收藏
页码:814 / 820
页数:7
相关论文
共 50 条
  • [21] Effects of heat and moisture exchangers on tracheal mucociliary clearance in laryngectomized patients: a multi-center case–control study
    C. van den Boer
    S. H. Muller
    V. van der Noort
    R. A. Valdés Olmos
    A. Minni
    C. Parrilla
    F. J. M. Hilgers
    M. W. M. van den Brekel
    S. van der Baan
    European Archives of Oto-Rhino-Laryngology, 2015, 272 : 3439 - 3450
  • [22] HEAT AND MOISTURE EXCHANGERS - REPLY
    TILLING, SE
    HAYES, B
    BRITISH JOURNAL OF ANAESTHESIA, 1988, 61 (03) : 368 - 369
  • [23] Heat-moisture exchangers
    Seakins, P
    INTENSIVE CARE MEDICINE, 1996, 22 (07) : 714 - 714
  • [24] PRESSURE-FLOW CHARACTERISTICS OF COMMONLY USED HEAT-MOISTURE EXCHANGERS
    PLOYSONGSANG, Y
    BRANSON, R
    RASHKIN, MC
    HURST, JM
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (03): : 675 - 678
  • [25] PRESSURE-FLOW CHARACTERISTICS OF COMMONLY USED HEAT-MOISTURE EXCHANGERS
    PLOYSONGSANG, Y
    BRANSON, R
    RASHKIN, MC
    HURST, JM
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1986, 133 (04): : A124 - A124
  • [26] Effects of heat and moisture exchangers on tracheal mucociliary clearance in laryngectomized patients: a multi-center case-control study
    van den Boer, C.
    Muller, S. H.
    van der Noort, V.
    Olmos, R. A. Valdes
    Minni, A.
    Parrilla, C.
    Hilgers, F. J. M.
    van den Brekel, M. W. M.
    van der Baan, S.
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2015, 272 (11) : 3439 - 3450
  • [27] Comparison of humidification performances of heat and moisture exchangers (HME) on a bench test
    Taillé, S
    Lefrançois, F
    Lellouche, F
    Deye, N
    Maggiore, SM
    Jouvet, P
    Ricard, JD
    Fumagalli, B
    Brochard, I
    INTENSIVE CARE MEDICINE, 2001, 27 : S211 - S211
  • [28] Chronic adherence to heat and moisture exchanger use in laryngectomized patients
    Pedemonte-Sarrias, Gabriel
    Villatoro-Sologaistoa, Juan Carlos
    Ale-Inostroza, Paulina
    Lopez-Vilas, Montserrat
    Leon-Vintro, Xavier
    Quer-Agusti, Miquel
    ACTA OTORRINOLARINGOLOGICA ESPANOLA, 2013, 64 (04): : 247 - 252
  • [29] HEAT AND MOISTURE EXCHANGERS AND BREATHING FILTERS
    HEDLEY, RM
    ALLTGRAHAM, J
    BRITISH JOURNAL OF ANAESTHESIA, 1994, 73 (02) : 227 - 236