SODIUM DODECYL SULFATE COATED γ-ALUMINA SUPPORT MODIFIED BY A NEW SCHIFF BASE FOR SOLID PHASE EXTRACTION AND FLAME-AAS DETERMINATION OF LEAD AND COPPER IONS

被引:6
|
作者
Shokrollahi, Ardeshir [1 ]
Montazerozohori, Morteza [1 ]
Mehrpour, Tahere [1 ]
Tavallali, Hossein [2 ]
Khafri, Batool Zare [2 ]
Montaseri, Zohreh [2 ]
机构
[1] Univ Yasuj, Dept Chem, Yasuj 7591874831, Iran
[2] Payame Noor Univ, Dept Chem, Shiraz 71365944, Iran
来源
QUIMICA NOVA | 2013年 / 36卷 / 09期
关键词
solid phase extraction; Pb2+ and Cu2+ ions; BHAH; ENVIRONMENTAL-SAMPLES; TRACE-METALS; ONLINE PRECONCENTRATION; EQUILIBRIUM-CONSTANTS; SILICA-GEL; ICP-AES; SORBENT; SEPARATION; ACID; PB;
D O I
10.1590/S0100-40422013000900013
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A simple and fast approach for solid phase extraction is herein described, and used to determine trace amounts of Pb2+ and Cu2+ metal ions. The solid phase support is sodium dodecyl sulfate (SDS)-coated gamma-alumina modified with bis(2-hydroxy acetophenone)-1,6-hexanediimine (BHAH) ligand. The adsorbed ions were stripped from the solid phase by 6 mL of 4 M nitric acid as eluent. The eluting solution was analyzed by flame atomic absorption spectrometry (FAAS). The sorption recovery of metal ions was investigated with regard to the effects of pH, amount of ligand, gamma-alumina and surfactant and the amount and type of eluent. Complexation of BHAH with Pb2+ or Cu2+ ions was examined via spectrophotometry using the HypSpec program. The detection limit for Cu2+ was 7.9 mu g L-1 with a relative standard deviation of 1.67%, while that for Pb2+ was 6.4 mu g L-1 with a relative standard deviation of 1.64%. A preconcentration factor of 100 was achieved for these ions. The method was successfully applied to determine analyte concentrations in samples of liver, parsley, cabbage, and water.
引用
收藏
页码:1354 / 1359
页数:6