Very Long-Term Outcomes of the Carpentier-Edwards Perimount Aortic Valve in Patients Aged 60 or Younger

被引:94
|
作者
Bourguignon, Thierry
El Khoury, Rym
Candolfi, Pascal
Loardi, Claudia
Mirza, Alain
Boulanger-Lothion, Julie
Bouquiaux-Stablo-Duncan, Anne-Lorraine
Espitalier, Fabien
Marchand, Michel
Aupart, Michel
机构
[1] Tours Univ Hosp, Dept Cardiac Surg, Tours, France
[2] Edwards Lifesci, Dept Biostat, Nyon, Switzerland
来源
ANNALS OF THORACIC SURGERY | 2015年 / 100卷 / 03期
关键词
BIOPROSTHETIC VALVE; ROSS PROCEDURE; HEART-VALVE; PERICARDIAL BIOPROSTHESIS; REPLACEMENT; POSITION; REOPERATION; IMPLANTATION; GUIDELINES; EXPERIENCE;
D O I
10.1016/j.athoracsur.2015.03.105
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Aortic valve replacement using a bioprosthesis remains controversial for patients younger than 60 years because of missing data on long-term outcomes in this age group. Methods. From 1984 to 2008, 383 Carpentier-Edwards Perimount pericardial aortic bioprostheses were implanted in 373 patients 60 years or younger (mean age, 51.0 +/- 9.2 years; 19% female). Multiple valve replacements were excluded from our cohort. Baseline clinical, perioperative, and follow-up data were recorded prospectively. The mean follow-up was 8.6 +/- 5.9 years, for a total of 3,299 valve-years. Follow-up was complete for 95.3% of patients included. Results. Operative mortality rate was 1.3%. Eighty-five late deaths occurred, for a linearized rate of 2.6%/valveyear. Actuarial survival rates averaged 78.1% +/- 2.6%, 65.6% +/- 3.5%, and 46.8% +/- 6.0% after 10, 15, and 20 years of follow-up, respectively. Mortality rate associated with reoperation was 2.3%. Actuarial freedom from reoperation rates attributable to structural valve deterioration at 10, 15, and 20 years were, respectively, 88.3% +/- 2.4%, 70.8% +/- 4.1%, and 38.1% +/- 5.6%. Competing risk analysis demonstrated an actual risk of explantation secondary to structural valve deterioration at 20 years of 41.6% +/- 4.1%. Expected valve durability was 17.6 years for this age group. Conclusions. In selected patients 60 years or younger undergoing aortic valve replacement with the Carpentier-Edwards Perimount bioprosthesis, the expected valve durability was 17.6 years. Reoperation for structural valve deterioration was associated with a low risk of mortality. (C) 2015 by The Society of Thoracic Surgeons
引用
收藏
页码:853 / 859
页数:7
相关论文
共 50 条
  • [11] Late hemodynamic and clinical outcomes of aortic valve replacement with the Carpentier-Edwards Perimount pericardial bioprosthesis
    Dellgren, G
    David, TE
    Raanani, E
    Armstrong, S
    Ivanov, J
    Rakowski, H
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 124 (01): : 146 - 154
  • [12] Long-term durability and patient functional status of the Carpentier-Edwards® Perimount® pericardial bioprosthesis in the aortic position
    Frater, RWM
    Furlong, P
    Cosgrove, DM
    Okies, JE
    Colburn, LQ
    Katz, AS
    Lowe, NL
    Ryba, EA
    JOURNAL OF HEART VALVE DISEASE, 1998, 7 (01): : 48 - 53
  • [13] 20 years' durability of Carpentier-Edwards Perimount stented pericardial aortic valve
    Bergoend, E.
    Aupart, M. R.
    Mirza, A.
    Meurisse, Y. A.
    Sirinelli, A. L.
    Neville, P. H.
    Marchand, M. A.
    AORTIC ROOT SURGERY: THE BIOLOGICAL SOLUTION, 2010, : 441 - +
  • [14] Carpentier-Edwards Pericardial Middle-aged Patients Aortic Valve in
    Sakamoto, Yoshimasa
    Hashimoto, Kazuhiro
    Okuyama, Hiroshi
    Ishii, Shinichi
    Inoue, Takahiro
    Kinouchi, Katsushi
    Abe, Takayuki
    GENERAL THORACIC AND CARDIOVASCULAR SURGERY, 2005, 53 (09) : 465 - 469
  • [15] Long-term outcomes of the Carpentier-Edwards pericardial valve prosthesis in the aortic position: Effect of patient age
    Biglioli, P
    Spampinato, N
    Cannata, A
    Musumeci, A
    Parolari, A
    Gagliardi, C
    Alamanni, F
    JOURNAL OF HEART VALVE DISEASE, 2004, 13 : S49 - S51
  • [16] Long-term survival after Carpentier-Edwards Perimount aortic valve replacement in Western Denmark: a multi-centre observational study
    Krasniqi, Lytfi
    Kronby, Mads P.
    Riber, Lars P. S.
    JOURNAL OF CARDIOTHORACIC SURGERY, 2021, 16 (01)
  • [17] Long-term survival after Carpentier-Edwards Perimount aortic valve replacement in Western Denmark: a multi-centre observational study
    Lytfi Krasniqi
    Mads P. Kronby
    Lars P. S. Riber
    Journal of Cardiothoracic Surgery, 16
  • [18] Long-term Clinical and Hemodynamic Performance of the Medtronic Hancock II versus the Carpentier-Edwards Perimount Aortic Bioprostheses
    Chan, Vincent
    Kulik, Alexander
    Tran, Anthony
    Lam, B-Khanh
    Hendry, Paul
    Rubens, Fraser
    Mesana, Thierry
    Ruel, Marc
    CIRCULATION, 2009, 120 (18) : S849 - S849
  • [19] Thromboembolic events after aortic valve replacement in elderly patients with a Carpentier-Edwards perimount pericardial bioprosthesis
    Mistiaen, W
    Van Cauwelaert, P
    Muylaert, P
    Sys, SU
    Harrisson, F
    Bortier, H
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2004, 127 (04): : 1166 - 1170
  • [20] Short-term hemodynamic performance of the mitral Carpentier-Edwards PERIMOUNT pericardial valve
    Firstenberg, MS
    Morehead, AJ
    Thomas, JD
    Smedira, NG
    Cosgrove, DM
    Marchand, MA
    ANNALS OF THORACIC SURGERY, 2001, 71 (05): : S285 - S288