Development of a highly sensitive, second antibody format chemiluminescence enzyme immunoassay for the determination of 17β-estradiol in wastewater

被引:72
|
作者
Zhao, LX
Lin, JM [1 ]
Li, ZJ
Ying, XT
机构
[1] Chinese Acad Sci, State Key Lab Environm Chem & Ecotoxicol, Ecoenvironm Sci Res Ctr, Beijing 100085, Peoples R China
[2] Beijing Acad Sci Technol, Kemei Dongya Inst Biotech, Beijing 100012, Peoples R China
基金
中国国家自然科学基金;
关键词
17; beta-estradiol; fluorescein-iso-thiocyanate (FITC)-anti-FITC system; chemiluminescence immunoassay; wastewater;
D O I
10.1016/j.aca.2005.11.034
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the present work, a highly sensitive chemiluminescence enzyme immunoassay (CLEIA) method for the determination of 17 beta-estradiol (E2) in the wastewater was developed, which used the fluorescein-iso-thiocyanate (FITC)-anti-FITC system as solid phase and the biotin-streptavidin system as signal amplification system, with the highly sensitive 4-methoxy-4-(3-phosphatephenyl)-spiro-(1,2-dioxetane-3,2'-adamantane) (AMPPD)-alkaline phosphatase (ALP) system as chemiluminescence detection system. The specific polyclonal antibody was produced against a conjugate of estradiol-bovine serum albumin (BSA), biotinoyl-diaminodioxaoctane-estradiol conjugate was synthesized. The specificity of anti-E2 antibody was also studied. The results showed that anti-E2 antibody has no cross-reactions for estriol, ethinyl E2, estrone commonly coexisting with E2 in the wastewater. Samples were prepared with solid-phase extraction using C-18 cartridges for the removal of matrix effects. The linear range was 2.5-1600 pg ml(-1) and detection limit was 1.5 pg ml(-1). As our known of knowledge, the method was more sensitive for the determination of E2 in water compared with other immunoassay methods. Recoveries were in the range of 80-110% in spiked tap water and wastewater. This method has been successfully used for the detection of E2 in wastewater, compared with the commercially radioimmunoassay (RIA) kit, the correlation was good. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:290 / 295
页数:6
相关论文
共 50 条
  • [1] Development of a Highly Sensitive Chemiluminescence Enzyme Immunoassay for the Determination of Bisphenol A
    Liu, Lirui
    Liu, Liqin
    Chen, Xueqing
    Shi, Guoqing
    ADVANCES IN ENVIRONMENTAL TECHNOLOGIES, PTS 1-6, 2013, 726-731 : 1283 - 1286
  • [2] Development and validation of a simple, sensitive, second antibody format enzyme immunoassay for LH determination in plasma
    Prakash, BS
    Paul, V
    Anandlaxmi, N
    JOURNAL OF IMMUNOLOGICAL METHODS, 2002, 270 (02) : 281 - 290
  • [3] A secondary antibody format chemiluminescence immunoassay for the determination of estradiol in human serum
    Xin, Tian-Bing
    Chen, Hui
    Lin, Zhen
    Liang, Shu-Xuan
    Lin, Jin-Ming
    TALANTA, 2010, 82 (04) : 1472 - 1477
  • [4] Highly sensitive second-antibody enzyme immunoassay for determination of estradiol-17β concentration in blood plasma of the mithun (Bos frontalis)
    Mondal, Mohan
    Prakash, Bukkaraya Samudram
    ZOOLOGICAL SCIENCE, 2007, 24 (04) : 408 - 413
  • [5] Development and validation of a simple, sensitive, second antibody format enzyme immunoassay (EIA) for LH determination in mithun (Bos frontalis) plasma
    Mondal, M
    Dhali, A
    Prakash, B
    Rajkhowa, C
    Prakash, BS
    JOURNAL OF IMMUNOASSAY & IMMUNOCHEMISTRY, 2005, 26 (02): : 157 - 167
  • [6] DETERMINATION OF SALBUTAMOL, CLENBUTEROL, AND BROMBUTEROL IN URINE BY A HIGHLY SENSITIVE CHEMILUMINESCENCE ENZYME IMMUNOASSAY
    Wu, Yuping
    Xu, Fei
    Jiang, Haiyang
    Tao, Xiaoqi
    Zhu, Kui
    Liu, Wenli
    Cui, Yifang
    Huang, Xiaoyong
    Ding, Shuangyang
    ANALYTICAL LETTERS, 2014, 47 (16) : 2761 - 2773
  • [7] Development of a highly sensitive and selective microplate chemiluminescence enzyme immunoassay for the determination of free thyroxine in human serum
    Wang, Xu
    Chen, Hui
    Lin, Jin-Ming
    Ying, Xitang
    INTERNATIONAL JOURNAL OF BIOLOGICAL SCIENCES, 2007, 3 (05): : 274 - 280
  • [8] Development of a highly sensitive chemiluminescence enzyme immunoassay using enhanced luminol as substrate
    Tao, Xiaoqi
    Wang, Wenjun
    Wang, Zhanhui
    Cao, Xingyuan
    Zhu, Jinghui
    Niu, Lanlan
    Wu, Xiaoping
    Jiang, Haiyang
    Shen, Jianzhong
    LUMINESCENCE, 2014, 29 (04) : 301 - 306
  • [9] Detection of 17β-estradiol in river and human urine by highly sensitive chemiluminescent enzyme immunoassay
    Wei Hui-Bin
    Lin Jin-Ming
    Wu Dan-Ning
    Zhao Li-Xia
    Li Zhen-Jia
    Ying Xi-Tang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2007, 35 (03) : 320 - 324
  • [10] A HIGHLY SENSITIVE MICROTITRE PLATE ENZYME-IMMUNOASSAY FOR ESTRADIOL-17-BETA
    MAUREL, MC
    LABROUSSE, H
    TERQUI, M
    AVRAMEAS, S
    JOURNAL OF IMMUNOLOGICAL METHODS, 1987, 102 (02) : 165 - 172