Microscopic analysis of opacification in Ioflex® hydrophilic acrylic intraocular lenses

被引:5
|
作者
Ventura, Bruna Vieira [1 ]
Ventura, Marcelo [1 ,2 ]
Lira, Wagner [1 ]
Ventura, Camila Vieira [1 ]
Santhiago, Marcony Rodrigues [3 ]
Werner, Liliana [4 ]
机构
[1] Altino Ventura Fdn, BR-50070020 Recife, PE, Brazil
[2] Hosp Olhos Pernambuco HOPE, Recife, PE, Brazil
[3] Cleveland Clin, Cole Eye Inst, Dept Refract Surg, Cleveland, OH 44106 USA
[4] Univ Utah, John A Moran Eye Ctr, Intermt Ocular Res Ctr, Salt Lake City, UT USA
关键词
Lenses; intraocular; Lens implantation; Reoperation; Device removal; Microscopy; Case reports; POSTOPERATIVE OPACIFICATION; CALCIFICATION; COMPLICATIONS;
D O I
10.1590/S0034-72802012000300002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Five cases of intraocular lens (IOL) opacification in patients implanted with the Mediphacos Ioflex (R) IOL are described. Clinical data in each case was obtained from the patient's medical record. The five explanted IOLs underwent gross and light microscopic analysis. Selected lenses were processed for further evaluation and multiple sagittal cuts were stained with the alizarin red and von Kossa methods. Light microscope analysis confirmed the presence of deposits on and within the lens, which stained positive for calcium. The reason why calcification has occurred in these cases remains unclear, but surgeons should be aware of this potential late postoperative complication.
引用
收藏
页码:149 / 154
页数:6
相关论文
共 50 条
  • [41] Hydrophilic acrylic intraocular lens optic opacification in a diabetic patient
    Dae Il Park
    Sung Woo Ha
    Seong Bae Park
    Helen Lew
    Japanese Journal of Ophthalmology, 2011, 55 : 595 - 599
  • [42] Hydrophilic acrylic intraocular lens optic opacification in a diabetic patient
    Park, Dae Il
    Ha, Sung Woo
    Park, Seong Bae
    Lew, Helen
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2011, 55 (06) : 595 - 599
  • [43] Long-term efficiency of cataract surgery with hydrophilic acrylic Ioflex intraocular lens
    Jorge, Priscilla A.
    Koch, Camila Ribeiro
    Jorge, Delano
    Kara-Junior, Newton
    CLINICS, 2017, 72 (09) : 543 - 546
  • [44] Comparison of posterior capsule opacification rates between hydrophilic and hydrophobic single-viece acrylic intraocular lenses
    Heatley, CJ
    Spalton, DJ
    Kumar, A
    Jose, R
    Boyce, J
    Bender, LE
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04): : 718 - 724
  • [45] Exchange technique for opacified hydrophilic acrylic intraocular lenses
    Voros, GM
    Strong, NP
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2005, 15 (04) : 465 - 467
  • [46] Three cases of hydrophilic acrylic Intraocular lenses exchange
    Nakanome, S
    Watanabe, H
    Tanaka, K
    Kobayakawa, S
    Tochikubo, T
    SOE 2003: XIV CONGRESS OF THE EUROPEAN SOCIETY OF OPHTHALMOLOGY, 2003, : 263 - 267
  • [47] Microscopic analysis of an opacified OFT CRYL® hydrophilic acrylic intraocular lens
    Ventura, Bruna Vieira
    MacLean, Kyle Douglas
    Lira, Wagner
    de Oliveira, Daniele Mendes
    Ventura, Camila Vieira
    Werner, Liliana
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2016, 79 (04) : 255 - 257
  • [48] Opacification of Intraocular Lenses
    Khoramnia, Ramin
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2021, 238 (07) : 831 - 842
  • [49] Scanning electron microscopic analysis of foldable acrylic and hydrogel intraocular lenses
    Kohnen, T
    Magdowski, G
    Koch, DD
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 : 1342 - 1350
  • [50] Opacification of hydrophilic acrylic intraocular lens attributable to calcification: Investigation on mechanism
    Gartaganis, Sotirios P.
    Kanellopoulou, Dimitra G.
    Mela, Ephigenia K.
    Panteli, Vassiliki S.
    Koutsoukos, Petros G.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 146 (03) : 395 - 403