Bead-based polymerase chain reaction on a microchip

被引:8
|
作者
Hilton, John P. [1 ]
Nguyen, ThaiHuu [1 ]
Barbu, Mihaela [2 ]
Pei, Renjun [2 ]
Stojanovic, Milan [2 ]
Lin, Qiao [1 ]
机构
[1] Columbia Univ, Dept Mech Engn, New York, NY 10027 USA
[2] Columbia Univ, Dept Med, Div Clin Pharmacol & Expt Therapeut, New York, NY 10032 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
Polymerase chain reaction; Bead; DNA capture; DNA purification; Microfluidics; PHASE DNA AMPLIFICATION; BORDETELLA-PERTUSSIS; MAGNETIC PARTICLES; PCR; CHIP; STREPTAVIDIN; MOLECULES; PATHOGENS; DIAGNOSIS; EMULSION;
D O I
10.1007/s10404-012-0993-8
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
We present a bead-based approach to microfluidic polymerase chain reaction (PCR), enabling fluorescent detection and sample conditioning in a single microchamber. Bead-based PCR, while not extensively investigated in microchip format, has been used in a variety of bioanalytical applications in recent years. We leverage the ability of bead-based PCR to accumulate fluorescent labels following DNA amplification to explore a novel DNA detection scheme on a microchip. The microchip uses an integrated microheater and temperature sensor for rapid control of thermal cycling temperatures, while the sample is held in a microchamber fabricated from (poly)dimethylsiloxane and coated with Parylene. The effects of key bead-based PCR parameters, including annealing temperature and concentration of microbeads in the reaction mixture, are studied to achieve optimized device sensitivity and detection time. The device is capable of detecting a synthetically prepared section of the Bordetella pertussis genome in as few as 10 temperature cycles with times as short as 15 min. We then demonstrate the use of the procedure in an integrated device; capturing, amplifying, detecting, and purifying template DNA in a single microfluidic chamber. These results show that this method is an effective method of DNA detection which is easily integrated in a microfluidic device to perform additional steps such as sample pre-conditioning.
引用
收藏
页码:749 / 760
页数:12
相关论文
共 50 条
  • [31] Ultrasonic enhancement of bead-based bioaffinity assays
    Wiklund, M.
    Hertz, H. M.
    LAB ON A CHIP, 2006, 6 (10) : 1279 - 1292
  • [32] Analyzing Cell- and Bead-Based Assays
    Bungard, Claire
    Zhang, Xiao
    Robertson, Alasdair
    GENETIC ENGINEERING & BIOTECHNOLOGY NEWS, 2011, 31 (09): : 38 - 39
  • [33] Bead-based HTS applications in drug discovery
    Meza, MB
    DRUG DISCOVERY TODAY, 2000, : 38 - 41
  • [34] A Chip for Detecting Tuberculosis Drug Resistance Based on Polymerase Chain Reaction (PCR)-Magnetic Bead Molecule Platform
    Lyu, Jingtong
    Wu, Wenjie
    Cheng, Peng
    Liu, Xun
    Luo, Fei
    Zhang, Zehua
    Tang, Kanglai
    Xu, Jianzhong
    FRONTIERS IN MICROBIOLOGY, 2018, 9
  • [35] Microfluidic, bead-based assay: Theory and experiments
    Thompson, Jason A.
    Bau, Haim H.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (02): : 228 - 236
  • [36] A novel bead-based fluorescence immunoassay for aldosterone
    Sun, Min
    Liu, Chao
    JOURNAL OF BIOMEDICAL RESEARCH, 2011, 25 (03): : 213 - 219
  • [37] Limit of detection for a bead-based diffraction biosensor
    Lim, Youbong
    Nolte, David D.
    Arif, Khalid
    Savran, Cagri
    NANOSCALE IMAGING, SENSING, AND ACTUATION FOR BIOMEDICAL APPLICATIONS VII, 2010, 7574
  • [38] Bead-based Enzymatic Assay On-a-chip
    Karnik, Sonali J.
    Cahoon, Stacey
    Bhushan, Abhinav
    FASEB JOURNAL, 2017, 31
  • [39] A functionally dynamic microchamber with rapid mixing and reaction capabilities for magnetic bead-based immunoassay
    Do, J
    Ahn, CH
    Micro Total Analysis Systems 2004, Vol 2, 2005, (297): : 416 - 418
  • [40] Holographically encoded microparticles for bead-based assays
    Birtwell, Sam W.
    Banu, Shahanara
    Zheludev, Nikolay I.
    Morgan, Hywel