Speech recognition as a function of the number of electrodes used in the SPEAK cochlear implant speech processor

被引:286
|
作者
Fishman, KE [1 ]
Shannon, RV [1 ]
Slattery, WH [1 ]
机构
[1] HOUSE EAR RES INST, LOS ANGELES, CA 90057 USA
来源
关键词
cochlear implant; speech recognition; electrical stimulation; electrode number;
D O I
10.1044/jslhr.4005.1201
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Speech recognition was measured in listeners with the Nucleus-22 SPEAK speech processing strategy as a function of the number of electrodes. Speech stimuli were analyzed into 20 frequency bands and processed according to the usual SPEAK processing strategy. In the normal clinical processor each electrode is assigned to represent the output of one Filter. To create reduced-electrode processors the output of several adjacent filters were directed to a single electrode, resulting in processors with 1, 2, 4, 7, 10, and 20 electrodes. The overall spectral bandwidth was preserved, but the number of active electrodes was progressively reduced. After a 2-day period of adjustment to each processor, speech recognition performance was measured on medial consonants, vowels, monosyllabic words, and sentences. Performance with a single electrode processor was poor in all listeners, and average performance increased dramatically on all test materials as the number of electrodes was increased from 1 to 4. No differences in average performance were observed on any test in the 7-, 10-, and 20-electrode conditions. On sentence and consonant tests there was no difference between average performance with the 4-electrode and 20-electrode processors. This pattern of results suggests that cochlear implant listeners are not able to make Full use of the spectral information on all 20 electrodes. Further research is necessary to understand the reasons For this limitation and to understand how to increase the amount of spectral information in speech received by implanted listeners.
引用
收藏
页码:1201 / 1215
页数:15
相关论文
共 50 条
  • [1] Speech recognition as a function of the number of channels for pediatric cochlear implant recipients
    Gifford, Rene H.
    Sunderhaus, Linsey W.
    Holder, Jourdan T.
    Berg, Katelyn A.
    Dawant, Benoit M.
    Noble, Jack H.
    Perkins, Elizabeth
    Camarata, Stephen
    JASA EXPRESS LETTERS, 2022, 2 (09):
  • [2] Speech recognition with the MPEAK and SPEAK speech-coding strategies of the nucleus cochlear implant
    Holden, LK
    Skinner, MW
    Holden, TA
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1997, 116 (02) : 163 - 167
  • [3] Association of Speech Processor Technology and Speech Recognition Outcomes in Adult Cochlear Implant Users
    Dixon, Peter R.
    Shipp, David
    Smilsky, Kari
    Lin, Vincent Y.
    Le, Trung
    Chen, Joseph M.
    OTOLOGY & NEUROTOLOGY, 2019, 40 (05) : 595 - 601
  • [4] Cochlear implant speech recognition with speech maskers
    Stickney, GS
    Zeng, FG
    Litovsky, R
    Assmann, P
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2004, 116 (02): : 1081 - 1091
  • [5] Cochlear implant speech recognition with speech maskers
    Stickney, G.S. (stickney@uci.edu), 1600, Acoustical Society of America (116):
  • [6] Effects of the number of active electrodes on tone and speech perception by nucleus 22 cochlear implant users with SPEAK strategy
    Hsu, CJ
    Horng, MJ
    Fu, QJ
    UPDATES IN COCHLEAR IMPLANTATION, 2000, 57 : 257 - 259
  • [7] Evoked cortical activity and speech recognition as a function of the number of simulated cochlear implant channels
    Friesen, L. M.
    Tremblay, K. L.
    Rohila, N.
    Wright, R. A.
    Shannon, R. V.
    Baskent, D.
    Rubinstein, J. T.
    CLINICAL NEUROPHYSIOLOGY, 2009, 120 (04) : 776 - 782
  • [8] THE EFFECTS OF VARYING THE NUMBER OF COCHLEAR IMPLANT ELECTRODES ON SPEECH-PERCEPTION
    HOLMES, AE
    KEMKER, FJ
    MERWIN, GE
    AMERICAN JOURNAL OF OTOLOGY, 1987, 8 (03): : 240 - 246
  • [9] Improving speech comprehension using a new cochlear implant speech processor
    Mueller-Deile, J.
    Kortmann, T.
    Hoppe, U.
    Hessel, H.
    Morsnowski, A.
    HNO, 2009, 57 (06) : 567 - 574
  • [10] Folded Architecture for Digital Gammatone Filter Used in Speech Processor of Cochlear Implant
    Karuppuswamy, Rajalakshmi
    Arumugam, Kandaswamy
    Priya, Swathi M.
    ETRI JOURNAL, 2013, 35 (04) : 697 - 705