Real-Time Temperature Mapping in Ultra-Low Freezers as a Standard Quality Assessment

被引:8
|
作者
Powell, Samantha [1 ]
Molinolo, Alfredo [1 ]
Masmila, Edgar [1 ]
Kaushal, Sharmeela [1 ]
机构
[1] UC San Diego Moores Canc Ctr, Biorepository & Tissue Technol Shared Resources, RM 3331,3855 Hlth Sci Dr, La Jolla, CA 92093 USA
关键词
storage; biorepository; CAP accreditation; real-time temperature; ultra-low freezer;
D O I
10.1089/bio.2018.0108
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
Adequate preservation of biospecimens has been proven to be critical to obtain reliable and reproducible results in genomics, transcriptomics, proteomics, and many other assays. Most biological assays can be performed on specimens preserved in -80 degrees C ultra-low freezers, but their quality can be influenced by temperature variability within storage units. Thus, regulatory standards such as those from the College of American Pathologists (CAP), the federal Clinical Laboratory Improvement Amendments, and standards from the Food and Drug Administration require temperature mapping, a standard quality assessment for accreditation when using ultra-low freezers for long-term biospecimen storage. The current mapping methods, providing annual/periodic data, may not be adequate indicators of temperature stability within the different zones of the freezers. In addition, they frequently require manual handling of biospecimens periodically, as they require freezers to be emptied or rearranged temporarily for the installation of temperature probes, risking the integrity of biospecimen quality. In this article, we describe a novel monitoring methodology based on real-time temperature reading of multiple zones by permanently installing thermocouples. An online cloud-based application records temperature variations within 1 minute intervals, and its 24/7 alert system triggers text alarm messages to a predefined set of users when temperature values increase above preset defaults. This provides an opportunity to take remedial action and to obtain a better-quality assessment. Our results indicate that real-time temperature monitoring at multiple zones of a freezer with a 1 minute resolution is a stable and sustainable methodology and, most importantly, lowers the risk of compromising the quality of the biospecimen. The design and use of the real-time monitoring system for ultra-low freezers is one of the acceptable methods by CAP to ensure the stability of biospecimen quality during long-term storage.
引用
收藏
页码:139 / 142
页数:4
相关论文
共 50 条
  • [21] Ultra-low power UWB for real time biomedical wireless sensing
    Lee, C. Y., Circuits and Systems Society, IEEE CASS; Science Council of Japan; The Inst. of Electronics, Inf. and Communication Engineers, IEICE; The Institute of Electrical and Electronics Engineers, Inc., IEEE (Institute of Electrical and Electronics Engineers Inc.):
  • [22] Real-time MR temperature mapping of ablation procedures
    Daniell, AL
    Sinha, S
    Sinha, US
    RADIOLOGY, 2000, 217 : 349 - 349
  • [23] RETICULA: Real-Time Code Quality Assessment
    Frunzio, Luigi
    Lin, Bin
    Lanza, Michele
    Bavota, Gabriele
    2018 25TH IEEE INTERNATIONAL CONFERENCE ON SOFTWARE ANALYSIS, EVOLUTION AND REENGINEERING (SANER 2018), 2018, : 542 - 546
  • [24] Real-time skin quality assessment system
    Krajna, Aleksandra
    Kocejko, Tomasz
    2024 16TH INTERNATIONAL CONFERENCE ON HUMAN SYSTEM INTERACTION, HSI 2024, 2024,
  • [25] Towards Real-time Image Quality Assessment
    Geary, Bobby
    Grecos, Christos
    REAL-TIME IMAGE AND VIDEO PROCESSING 2011, 2011, 7871
  • [26] Real-time Video Quality Assessment Platform
    Papp, Istvan
    Lukic, Nemanja
    Marceta, Zoran
    Teslic, Nikola
    2009 IEEE INTERNATIONAL CONFERENCE ON CONSUMER ELECTRONICS, 2009, : 215 - +
  • [27] Real-Time Ultra-Low Power ECG Anomaly Detection Using an Event-Driven Neuromorphic Processor
    Bauer, Felix Christian
    Muir, Dylan Richard
    Indiveri, Giacomo
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2019, 13 (06) : 1575 - 1582
  • [28] Ultra-low density geochemical mapping in Zimbabwe
    Zhao, Gengxin
    He, Fuqing
    Dai, Xuefu
    Zhang, Surong
    Yu, Rengan
    JOURNAL OF GEOCHEMICAL EXPLORATION, 2014, 144 : 552 - 571
  • [29] Development of a programmable standard of ultra-low capacitance values
    Khan, M. S.
    Seron, O.
    Thuillier, G.
    Thevenot, O.
    Gournay, P.
    Piquemal, F.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (05):
  • [30] Real-Time Motor Unit Tracking From sEMG Signals With Adaptive ICA on a Parallel Ultra-Low Power Processor
    Orlandi, Mattia
    Rapa, Pierangelo Maria
    Zanghieri, Marcello
    Frey, Sebastian
    Kartsch, Victor
    Benini, Luca
    Benatti, Simone
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2024, 18 (04) : 771 - 782