Peripheral dose measurements for 6 and 18 MV photon beams on a linear accelerator with multileaf collimator

被引:21
|
作者
Mazonakis, Michalis [1 ]
Zacharopoulou, Fotini [1 ]
Varveris, Haralambos [2 ]
Damilakis, John [3 ]
机构
[1] Univ Hosp Iraklion, Dept Med Phys, Iraklion 71110, Greece
[2] Univ Crete, Dept Radiotherapy, Fac Med, GR-71409 Iraklion, Greece
[3] Univ Crete, Dept Med Phys, Fac Med, GR-71409 Iraklion, Greece
关键词
multileaf collimator; peripheral dose; risk estimation;
D O I
10.1118/1.2977533
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Peripheral dose (PD) to critical structures outside treatment volume is of clinical importance. The aim of the current study was to estimate PD on a linear accelerator equipped with multileaf collimator (MLC). Dose measurements were carried out using an ionization chamber embedded in a water phantom for 6 and 18 MV photon beams. PD values were acquired for field sizes from 5x5 to 20x20 cm(2) in increments of 5 cm at distances up to 24 cm from the field edge. Dose data were obtained at two collimator orientations where the measurement points are shielded by MLC and jaws. The variation of PD with the source to skin distance (SSD), depth, and lateral displacement of the measurement point was evaluated. To examine the dependence of PD upon the tissue thickness at the entrance point of the beam, scattered dose was measured using thermoluminescent dosemeters placed on three anthropomorphic phantoms simulating 5- and 10-year-old children and an average adult patient. PD from 6 MV photons varied from 0.13% to 6.75% of the central-axis maximum dose depending upon the collimator orientation, extent of irradiated area, and distance from the treatment field. The corresponding dose range from 18 MV x rays was 0.09% to 5.61%. The variation of PD with depth and with lateral displacements up to 80% of the field dimension was very small. The scattered dose from both photon beams increased with the increase of SSD or tissue thickness along beam axis. The presented dosimetric data set allows the estimation of scattered dose outside the primary beam. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:4396 / 4403
页数:8
相关论文
共 50 条
  • [1] Measurements of peripheral dose for multileaf collimator based linear accelerator
    Balasubramanian, R.
    Sellakumar, P.
    Bilimagga, Ramesh S.
    Supe, Sanjay S.
    Sankar, B. N.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2006, 11 (06) : 281 - 285
  • [2] Dosimetric evaluation of 120-leaf multileaf collimator in a Varian linear accelerator with 6-MV and 18-MV photon beams
    Mohan, R.
    Jayesh, K.
    Joshi, R. C.
    Al-Idrisi, Maha
    Narayanamurthy, P.
    Das Majumdar, Saroj Kumar
    JOURNAL OF MEDICAL PHYSICS, 2008, 33 (03) : 114 - 118
  • [3] Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams
    De Vlamynck, K
    Palmans, H
    Verhaegen, F
    De Wagter, C
    De Neve, W
    Thierens, H
    MEDICAL PHYSICS, 1999, 26 (09) : 1874 - 1882
  • [4] A theoretical multileaf collimator model for fast Monte Carlo dose calculation of linac 6/10 MV photon beams
    Aboulbanine, Zakaria
    El Khayati, N.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (05):
  • [5] Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams
    Bilge, Hatice
    Cakir, Aydin
    Okutan, Murat
    Acar, Hilal
    PHYSICA MEDICA, 2009, 25 (02) : 101 - 104
  • [6] The characterization of unflattened photon beams from a 6 MV linear accelerator
    Cashmore, Jason
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (07): : 1933 - 1946
  • [7] Investigation and modelling of the surface dose from linear accelerator produced 6 and 10 MV photon beams
    Lamb, A
    Blake, S
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (05): : 1133 - 1146
  • [8] Monte Carlo simulation approaches to dose distributions for 6 MV photon beams in clinical linear accelerator
    Tartar, Ahmet
    BIOCYBERNETICS AND BIOMEDICAL ENGINEERING, 2014, 34 (02) : 90 - 100
  • [9] A comparative study of the peripheral doses from a linear accelerator with a multileaf collimator system
    Acun, Hediye
    Zubaroglu, Ali
    Kemikler, Gonul
    Bozkurt, Ahmet
    RADIATION PROTECTION DOSIMETRY, 2014, 158 (03) : 299 - 306
  • [10] Peripheral dose from a linear accelerator equipped with multileaf collimation
    Stern, RL
    MEDICAL PHYSICS, 1999, 26 (04) : 559 - 563