Automatic pubic arch localization in 3D ultrasound images using pre-operative image information for prostate biopsy and brachytherapy

被引:0
|
作者
Shao, Wei [1 ]
Wu, Ruoyun [2 ]
Thng, Choon Hua [3 ]
Ling, Keck Voon [1 ]
Ho, Henry Sun Sien [4 ]
Cheng, Christopher Wai Sam [4 ]
Ng, Wan Sing [5 ]
机构
[1] Nanyang Technol Univ, Sch Elect & Elect Engn, Singapore 639798, Singapore
[2] Tan Tock Seng Hosp, Clin Res Unit, Singapore, Singapore
[3] Natl Canc Ctr, Dept Oncol Imaging, Singapore, Singapore
[4] Singapore Gen Hosp, Dept Urol, Singapore, Singapore
[5] Nanyang Technol Univ, Sch Mech & Aerosp Engn, Singapore 639798, Singapore
关键词
Image registration; Transrectal ultrasound; Pubic arch interference; Brathytherapy;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Transrectal ultrasound (TRUS) is widely used as the primary imaging modality for transperineal prostate biopsy and treatment Along the passage of the biopsy needle or seed implantation, the pubic arch is a potential barrier and required to be identified before the planning. This paper presented a method that can automatically locate the pubic arch in three-dimensional transrectal ultrasound images, utilizing the extracted pubic arch information from pre-operative MRI or CT images of the same patient. Based on the true surface description of the pubic arch segmented from the pre-operative images, a surface-to-volume image registration technique was applied to automatically search for the best fit between the pubic surface and its TRUS image volume. The self-registration test showed that our method has good accuracy and consistency. Similar consistency was also found in the experiments using three sets of patient data.
引用
收藏
页码:82 / 84
页数:3
相关论文
共 50 条
  • [11] The Accuracy of 3D Rectal Ultrasound for Pre-Operative Rectal Cancer Staging
    Arya, Aman V.
    Romagnoli, Cesare
    Man, Lu
    Hussain, Nadeem
    Fenster, Aaron
    Yan, Brian
    GASTROENTEROLOGY, 2012, 142 (05) : S342 - S342
  • [12] Integration of intra-operative 3D ultrasound with pre-operative MRI for neurosurgical guidance
    Gobbi, DG
    Comeau, RM
    Lee, BKH
    Peters, TM
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1738 - 1740
  • [13] Automatic 3D segmentation of intravascular ultrasound images using region and contour information
    Cardinal, MHR
    Meunier, J
    Soulez, G
    Maurice, RL
    Thérasse, T
    Cloutier, G
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2005, PT 1, 2005, 3749 : 319 - 326
  • [14] MANUAL AND AUTOMATIC 3D SEGMENTATION OF THE PUBIC PART OF THE LEVATOR ANI MUSCLE IN ULTRASOUND IMAGES
    Van den Noort, F.
    Grob, A. T.
    Van Stralen, M.
    Slump, C. H.
    Van der Vaart, C.
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2017, 28 : S149 - S150
  • [15] Ultrasound and 3D MR Image Registration of the Prostate for Brachytherapy Surgical Navigation
    Zhang, Shihui
    Jiang, Shan
    Yang, Zhiyong
    Liu, Ranlu
    MEDICINE, 2015, 94 (40)
  • [16] Automated localization of implanted seeds in 3D TRUS images used for prostate brachytherapy
    Wei, Zhouping
    Gardi, Lori
    Downey, Donal B.
    Fenster, Aaron
    MEDICAL PHYSICS, 2006, 33 (07) : 2404 - 2417
  • [17] Segmentation of radioactive seed in 3D ultrasound images for intraoperative LDR prostate brachytherapy
    Wei, Z.
    Fenster, A.
    MEDICAL PHYSICS, 2006, 33 (06) : 2229 - 2230
  • [18] Automatic segmentation of 3D prostate MR images with iterative localization refinement
    Zhou, Wenhui
    Tao, Xing
    Wei, Zhan
    Lin, Lili
    DIGITAL SIGNAL PROCESSING, 2020, 98
  • [19] Rigid 3D Registration of Pre-operative Information for Semi-Autonomous Surgery
    Piccinelli, Nicola
    Roberti, Andrea
    Tagliabue, Eleonora
    Setti, Francesco
    Kronreif, Gernot
    Muradore, Riccardo
    Fiorini, Paolo
    2020 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2020, : 139 - 145
  • [20] Automatic 3D seed location and orientation detection in CT image for prostate brachytherapy
    Nguyen, Huu-Giao
    Fouard, Celine
    Meneu, Francois
    Giraud, Jean-Yves
    Troccaz, Jocelyne
    2014 IEEE 11th International Symposium on Biomedical Imaging (ISBI), 2014, : 1320 - 1323