The electromagnetic environment assessment of nuclear magnetic resonance device in hospital

被引:0
|
作者
Fu, Lu [1 ]
Yan, Zhaowen [1 ]
Shi, Guochang [1 ]
机构
[1] Beihang Univ, Dept Elect Informat Engn, Beijing, Peoples R China
关键词
electromagnetic environment; test; NMR device;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The purpose of this paper is to assess electromagnetic environment of nuclear magnetic resonance (NMR) device in hospital, in considering of the proposed construction of subway lines which would pass the hospital. A method is proposed to quantitatively analyze and evaluate electromagnetic interference of NMR device. It is derived through the analysis and evaluation: when subway line passes through the hospital, the field strength surrounding the hospital can reach 30dBuT to 40dBuT at several frequencies, which is greater than 20dBuT-the threshold of the NMR device. It can ensure the NMR equipment working normally, after applying an electromagnetic shielding measure which makes sure that the electromagnetic shielding effectiveness reaches 80dB.
引用
收藏
页码:1099 / 1102
页数:4
相关论文
共 50 条
  • [1] Design and implementation of an electromagnetic noise test system in nuclear magnetic resonance sounding environment
    College of Instrumentation and Electrical Engineering, Jilin University, Changchun
    130061, China
    Jilin Daxue Xuebao (Gongxueban), 6 (2034-2042):
  • [2] Medical high-technology assessment and implementation in a community hospital: Nuclear magnetic resonance
    Mantil, J.
    Willett, R.
    Sawyer, W.
    Biomedical Instrumentation and Technology, 1991, 25 (04): : 289 - 296
  • [3] Design and optimization of the magnetic circuit of a mobile nuclear magnetic resonance device for magnetic resonance imaging
    Popella, H
    Henneberger, G
    COMPEL-THE INTERNATIONAL JOURNAL FOR COMPUTATION AND MATHEMATICS IN ELECTRICAL AND ELECTRONIC ENGINEERING, 2001, 20 (01) : 269 - 278
  • [4] DEVICE FOR THERMOSTATTING SPECIMENS IN NUCLEAR MAGNETIC RESONANCE EXPERIMENTS
    ERMAKOV, VI
    ZAGORETS, PA
    GRUNAU, AP
    ZHURNAL FIZICHESKOI KHIMII, 1964, 38 (04): : 1030 - 1031
  • [5] Electromagnetic Evaluation of HTS RF Coils for Nuclear Magnetic Resonance
    Yamada, T.
    Saito, A.
    Oikawa, S.
    Koshita, K.
    Takahashi, M.
    Maeda, H.
    Ohshima, S.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2015, 25 (03)
  • [6] Occupational exposure to electromagnetic fields in magnetic resonance environment: basic aspects and review of exposure assessment approaches
    Hartwig, Valentina
    Romeo, Stefania
    Zeni, Olga
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2018, 56 (04) : 531 - 545
  • [7] PROGRAMMING DEVICE OF A SPECTROMETER FOR DOUBLE NUCLEAR-QUADRUPOLE RESONANCE-NUCLEAR MAGNETIC RESONANCE-RESONANCE
    ANFEROV, VN
    MIKHALKOV, VM
    MOLCHANOV, SV
    FEDOTOV, VV
    IZVESTIYA AKADEMII NAUK SSSR SERIYA FIZICHESKAYA, 1981, 45 (03): : 565 - 572
  • [8] Occupational exposure to electromagnetic fields in magnetic resonance environment: basic aspects and review of exposure assessment approaches
    Valentina Hartwig
    Stefania Romeo
    Olga Zeni
    Medical & Biological Engineering & Computing, 2018, 56 : 531 - 545
  • [9] The electromagnetic environment of Magnetic Resonance Imaging systems. Occupational exposure assessment reveals RF harmonics
    Gourzoulidis, G.
    Karabetsos, E.
    Skamnakis, N.
    Kappas, C.
    Theodorou, K.
    Tsougos, I.
    Maris, T. G.
    INTERNATIONAL CONFERENCE ON BIO-MEDICAL INSTRUMENTATION AND RELATED ENGINEERING AND PHYSICAL SCIENCES (BIOMEP 2015), 2015, 637
  • [10] CONTROL DEVICE FOR PULSE APPARATUS OF NUCLEAR MAGNETIC-RESONANCE
    BABKIN, AF
    GOLOD, OS
    MAKARENK.VL
    STRELTSO.VS
    PRIBORY I TEKHNIKA EKSPERIMENTA, 1973, (03): : 132 - 134