Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood

被引:18
|
作者
Yilmaz, Oznur [1 ]
Karapinar, Tolga [1 ]
机构
[1] Firat Univ, Dept Internal Med, Fac Vet Med, Elazig, Turkey
关键词
calcium; cattle; cow; point-of-care; sensitivity; PORTABLE CLINICAL ANALYZER; GAS PARTIAL PRESSURES; SUBCLINICAL HYPOCALCEMIA; ELECTROLYTE CONCENTRATIONS; NEUTROPHIL FUNCTION; DAIRY; PREVALENCE; PREDICTION; PLASMA; DEVICE;
D O I
10.1111/vcp.12705
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa(2+)) concentration in bovine blood is limited. Objective The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa(2+) in the blood of cattle with various disorders. Methods In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa(2+) concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa(2+) concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve. Results The Passing-Bablok regression equation that fit the iCa(2+) concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] x X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa(2+) concentration cutoff values of <1.00 and <1.18 mmol/L, respectively. Conclusions iCa(2+) concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity.
引用
收藏
页码:31 / 35
页数:5
相关论文
共 50 条
  • [21] Precision Evaluation of the i-STAT® Glucose Test in the i-STAT G Cartridge using the i-STAT 1 System for Venous, Arterial and Capillary Whole Blood Specimens in Clinical Settings
    Thibodeau, L.
    Aguirre, M. Gonzalez
    Hahn, S.
    Zhang, H.
    CLINICAL CHEMISTRY, 2024, 70 : I129 - I129
  • [22] Evaluation of the i-STAT Alinity point-of-care analyzer in pediatric patient populations
    Di Meo, Ashley
    Lee, Elsie
    Quach, Kim
    Nemec, Rose
    Zappitelli, Michael
    Licht, Christoph
    Adeli, Khosrow
    CLINICAL BIOCHEMISTRY, 2022, 109 : 104 - 104
  • [23] EVALUATION OF A NOVEL POINT-OF-CARE SYSTEM, THE I-STAT PORTABLE CLINICAL ANALYZER
    ERICKSON, KA
    WILDING, P
    CLINICAL CHEMISTRY, 1993, 39 (02) : 283 - 287
  • [24] Evaluation of i-STAT creatinine assay
    Gault, MH
    Seymour, ME
    Howell, WE
    NEPHRON, 2001, 88 (02): : 178 - 182
  • [25] Evaluation of i-STAT portable clinical analyzer in a neonatal and pediatric intensive care unit
    Murthy, JN
    Hicks, JM
    Soldin, SJ
    CLINICAL BIOCHEMISTRY, 1997, 30 (05) : 385 - 389
  • [26] Reliability of the i-STAT for the determination of blood electrolyte (K+, Na+, and CI-) concentrations in cattle
    Yildirim, E.
    Karapinar, T.
    Hayirli, A.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2015, 29 (01) : 388 - 394
  • [27] EVALUATION OF THE STABILITY OF IONIZED MAGNESIUM AND IONIZED CALCIUM ON THE "STAT PROFILE PRIME PLUS ANALYZER" OF NOVA BIOMEDICAL
    Gilson, M.
    Watar, F.
    Kovacs, S.
    Gadisseur, R.
    Cavalier, E.
    OSTEOPOROSIS INTERNATIONAL, 2019, 30 : S533 - S533
  • [28] Blood gases in the dog: Comparison of results obtained with the I-STAT portable analyzer and a reference method
    Malet, C
    Verwaerde, P
    Lagente, M
    de La Farge, F
    Braun, JP
    REVUE DE MEDECINE VETERINAIRE, 2000, 151 (07) : 733 - 733
  • [29] Evaluation of the stability of ionized magnesium and ionized calcium stability on the "stat profile prime plus analyzer" of nova biomedical
    Gilson, M.
    Ladang, A.
    Kovacs, S.
    Watar, F.
    Musso, G.
    Gadisseur, R.
    Le Goff, C.
    Cavalier, E.
    CLINICA CHIMICA ACTA, 2019, 493 : S776 - S776
  • [30] Accuracy of the i-STAT™ bedside blood gas analyser
    Sediame, S
    Zerah-Lancner, F
    d'Ortho, MP
    Adnot, S
    Harf, A
    EUROPEAN RESPIRATORY JOURNAL, 1999, 14 (01) : 214 - 217