Development of design methods for a centrifugal blood pump with a fluid dynamic approach: Results in hemolysis tests

被引:31
|
作者
Masuzawa, T
Tsukiya, T
Endo, S
Tatsumi, E
Taenaka, Y
Takano, H
Yamane, T
Nishida, M
Asztalos, B
Miyazoe, Y
Ito, K
Sawairi, T
Konishi, Y
机构
[1] Ibaraki Univ, Dept Mech Engn, Hitachi, Ibaraki 3168511, Japan
[2] Natl Cardiovasc Ctr, Res Inst, Dept Artificial Organs, Osaka, Japan
[3] Mech Engn Lab, Biomimet Div, Tsukuba, Ibaraki, Japan
[4] Nikkiso Co, Shizuoka Plant, Ctr Res & Dev, Shizuoka, Japan
关键词
rotary blood pump; centrifugal pump; blood trauma; hemolysis test; flow visualization; computational fluid dynamics;
D O I
10.1046/j.1525-1594.1999.06417.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to examine the relationship between local flow conditions and the hemolysis level by integrating hemolysis tests, flow visualization, and computational fluid dynamics to establish practical design criteria for centrifugal blood pumps with lower levels of hemolysis. The Nikkiso centrifugal blood pump was used as a standard model, and pumps with different values of 3 geometrical parameters were tested. The studied parameters were the radial gap between the outer edge of the impeller vane and the casing wall, the position of the outlet port, and the discharge angle of the impeller vane. The effect of a narrow radial gap on hemolysis was consistent with no evidence that the outlet port position or the vane discharge angle affected blood trauma in so far as the Nikkiso centrifugal blood pump was concerned. The radial gap should be considered as a design parameter of a centrifugal blood pump to reduce blood trauma.
引用
收藏
页码:757 / 761
页数:5
相关论文
共 50 条
  • [31] Assessment of the Efficiency of a Single-Use Blood Pump in Hemolysis Tests
    Semenikhin, P., V
    Skvortsov, A. E.
    Filipenko, K., V
    Remizov, M., V
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2022, 56 (04): : 251 - 255
  • [32] Assessment of the Efficiency of a Single-Use Blood Pump in Hemolysis Tests
    P. V. Semenikhin
    A. E. Skvortsov
    K. V. Filipenko
    M. V. Remizov
    Biomedical Engineering, 2022, 56 : 251 - 255
  • [33] Optimal Design of the Hydrodynamic Multi-Arc Bearing in a Centrifugal Blood Pump for the Improvement of Bearing Stiffness and Hemolysis Level
    Yasui, Kazuya
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Kawaguchi, Yasuo
    Yamane, Takashi
    ARTIFICIAL ORGANS, 2013, 37 (09) : 768 - 777
  • [34] Computational fluid dynamics analysis to establish the design process of a centrifugal blood pump: Second report
    Miyazoe, Y
    Sawairi, T
    Ito, K
    Konishi, Y
    Yamane, T
    Nishida, M
    Asztalos, B
    Masuzawa, T
    Tsukiya, T
    Endo, S
    Taenaka, Y
    ARTIFICIAL ORGANS, 1999, 23 (08) : 762 - 768
  • [35] Computational fluid dynamics analysis to establish the design process of a centrifugal blood pump: Second report
    Nikkiso Co., Ltd., Shizuoka, Japan
    不详
    不详
    不详
    不详
    Artif. Organs, 8 (762-768):
  • [36] Relationship between pump speed design and hemolysis in an axial flow blood pump
    Texas Heart Institute, Houston, TX, United States
    不详
    不详
    不详
    ARTIF. ORGANS, 5 (564-567):
  • [37] Improvement of hemolysis performance in a hydrodynamically levitated centrifugal blood pump by optimizing a shroud size
    Kosaka, Ryo
    Sakota, Daisuke
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (02) : 157 - 163
  • [38] Relationship between pump speed design and hemolysis in an axial flow blood pump
    Anai, H
    Wakisaka, Y
    Nakatani, T
    Taenaka, Y
    Takano, H
    Hadama, T
    ARTIFICIAL ORGANS, 1996, 20 (06) : 564 - 567
  • [39] Hemolysis estimation in turbulent flow for the FDA critical path initiative centrifugal blood pump
    Avci, Mesude
    Heck, Margaret
    O'Rear, Edgar A.
    Papavassiliou, Dimitrios V.
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2021, 20 (05) : 1709 - 1722
  • [40] Influence of radial clearance and rotor motion to hemolysis in a journal bearing of a centrifugal blood pump
    Kataoka, Hiroyuki
    Kimura, Yuichi
    Fujita, Hajime
    Takatani, Setsuo
    ARTIFICIAL ORGANS, 2006, 30 (11) : 841 - 854