Validation of the Pangao PG-800B26 upper arm blood pressure monitor in the general population according to the European Society of Hypertension and the British Hypertension Society protocols

被引:2
|
作者
Chen, Liang [1 ]
Li, Jiyu [2 ]
Wen, Jing [3 ]
Guo, Changfeng [4 ]
Zhang, Jingying [3 ]
Yu, Zhen [1 ]
机构
[1] Tongji Univ, Peoples Hosp 10, Clin Skills Training Ctr, Room 304,Bldg 6,301 Yanchangzhong Rd, Shanghai 200072, Peoples R China
[2] Tongji Univ, Peoples Hosp 10, Dept Gen Surg, Shanghai, Peoples R China
[3] Tongji Univ, Peoples Hosp 10, Dept Cardiol, Shanghai, Peoples R China
[4] Tongji Univ, Peoples Hosp 10, Dept Emergency, Shanghai, Peoples R China
关键词
blood pressure measurement; blood pressure monitor; validation study;
D O I
10.1097/MBP.0000000000000289
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective This study aimed to evaluate the accuracy of the automated oscillometric upper arm blood pressure monitor Pangao PG-800B26 for home blood pressure monitoring according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010 and the British Hypertension Society (BHS) protocols. Participants and methods Systolic and diastolic blood pressures were measured sequentially in 33 and 85 adults, respectively, on the basis of the ESH-IP and BHS protocols using a mercury sphygmomanometer (two observers) and the device (one supervisor). The procedures and analysis methods of the protocols were followed precisely. Results The device fulfilled the criteria of the ESH-IP, with device-observer differences of 1.01 +/- 5.16 and -0.58 +/- 4.17 mmHg for systolic and diastolic blood pressure, respectively. Furthermore, the A/A grade of the BHS protocol was also achieved for overall grading and for the three pressure levels, with average differences of 0.85 +/- 6.35 and -0.15 +/- 5.65 mmHg for systolic and diastolic blood pressure, respectively, which also fulfilled the requirements of the Association for the Advancement of Medical Instrumentation. Conclusion The Pangao PG-800B26 fulfilled the criteria of the ESH-IP 2010 and achieved the A/A grade of the BHS protocol, and hence can be recommended for home use in adults. Copyright (C) 2018 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:41 / 44
页数:4
相关论文
共 50 条
  • [1] Validation of the Pangao PG-800B11 blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols
    Bing, Sen
    Zhang, Cunhai
    Wang, Liping
    Li, Linyi
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2014, 19 (06) : 366 - 369
  • [2] Validation of the Pangao PG-800A36 automatic wrist blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols
    Zhao, Hairong
    Qiao, Weichang
    Zhang, Rui
    Cui, Peng
    Hou, Fanglin
    Zhang, Wenli
    BLOOD PRESSURE MONITORING, 2018, 23 (01) : 37 - 40
  • [3] Validation of the Pangao PG-800A11 wrist device assessed according to the European Society of Hypertension and the British Hypertension Society protocols
    Xie, Peigen
    Wang, Yanling
    Xu, Xiaoying
    Huang, Fei
    Pan, Jingru
    BLOOD PRESSURE MONITORING, 2015, 20 (02) : 108 - 111
  • [4] Validation of the SAA-102 home blood pressure monitor according to the protocols of the European Society of Hypertension, the Association for the Advancement of Medical Instrumentation and the British Society of Hypertension
    Zaetta, Vania
    Daniele, Longo
    Perkovic, Davor
    Prattico, Francesco
    Barisa, Marlena
    Perfetti, Paola
    Gabrieli, Alberto
    Buonocore, Francesco
    Winnicki, Mikolaj
    BLOOD PRESSURE MONITORING, 2007, 12 (06) : 363 - 368
  • [5] Validation of the fully automated Pangao PG-800B68 upper-arm device according to the European Society of Hypertension International Protocol revision 2010
    Yao, Rui
    Du, You-You
    Li, Ya-Peng
    Zhang, Yan-Zhou
    Chen, Qing-Hua
    Zhao, Luo-Sha
    Li, Ling
    BLOOD PRESSURE MONITORING, 2015, 20 (04) : 221 - 224
  • [6] Validation of the SAW-102 wrist home blood pressure monitor according to the protocols of the British Hypertension Society, the Association for the Advancement of Medical Instrumentation, and the European Society of Hypertension
    Zaetta, Vania
    Longo, Daniele
    Perkovic, Davor
    Perfetti, Paola
    Gabrieli, Alberto
    Prattico, Francesco
    Winnicki, Mikolaj
    BLOOD PRESSURE MONITORING, 2009, 14 (01) : 32 - 36
  • [7] Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols
    Liu, Ze-Yu
    Zhang, Qing-Han
    Ye, Xiao-Lei
    Liu, Da-Peng
    Cheng, Kang
    Zhang, Chun-Hai
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2017, 22 (02) : 101 - 104
  • [8] Validation of the Rossmax AC1000f upper-arm blood pressure monitor in adults according to the British Hypertension Society Protocol
    Huang, Jian-Feng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2020, 25 (04) : 231 - 235
  • [9] Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002
    Lueders, Stephan
    Krueger, Ralf
    Zemmrich, Claudia
    Forstner, Klaus
    Sturm, Claus-Dieter
    Bramlage, Peter
    BLOOD PRESSURE MONITORING, 2012, 17 (06) : 248 - 252
  • [10] Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol
    Zeng, Wei-Fang
    Kang, Yuan-Yuan
    Liu, Ming
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2013, 18 (03) : 177 - 181