STATISTICAL REVIEW OF THE INSULATION CAPACITY OF THE GEIGER-MUELLER COUNTER

被引:17
|
作者
Obrenovic, Marija D. [1 ]
Janicijevic, Aco J. [2 ]
Arbutina, Dalibor S. [3 ]
机构
[1] Univ Belgrade, Fac Elect Engn, Belgrade, Serbia
[2] Univ Belgrade, Fac Technol & Met, Belgrade, Serbia
[3] Univ Belgrade, Fac Mech Engn, Belgrade, Serbia
来源
关键词
Geiger Mueller counter; spontaneous breakdown; gas filtering experiment; BREAKDOWN;
D O I
10.2298/NTRP180913009O
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
This work considers the manifestation of spontaneous breakdowns of Geiger-Mueller counters. This is an experimental type of work. The reasons leading to the spontaneous breakdowns of Geiger-Mueller counters have been analysed under controlled laboratory conditions. The experiments were carried out under the "constant voltage". The tested Geiger-Mueller chambers were commercial chambers of the radial electric field and homogenous electric field. The experimental-statistical methods were used in order to choose the chambers with identical features (with 0.1 % of statistical reliability).The results of experiments showed that the spontaneous breakdown of the Geiger-Mueller counter happens even in the conditions of shielding. They also showed that those breakdowns have cumulative features. The reason for that is a positive feedback between the spontaneous breakdowns. The explanation is confirmed and quantified by the working gas filtering experiments.